Mohammed H Alam1,2, Dominique Auger1, Laura-Ann McGill1,2, Gillian C Smith1,2, Taigang He3, Cemil Izgi1, A John Baksi1, Rick Wage1, Peter Drivas1, David N Firmin1,2, Dudley J Pennell4,5,6.
Abstract
BACKGROUND: T2* magnetic resonance of tissue iron concentration has improved the outcome of transfusion dependant anaemia patients. Clinical evaluation is performed at 1.5 T but scanners operating at 3 T are increasing in numbers. There is a paucity of data on the relative merits of iron quantification at 3 T vs 1.5 T.Entities:
Keywords: 3 T; Heart; Iron overload; Liver; Magnetic resonance; Siderosis; T2*
Mesh:
Substances:
Year: 2016 PMID: 27391316 PMCID: PMC4938967 DOI: 10.1186/s12968-016-0259-9
Source DB: PubMed Journal: J Cardiovasc Magn Reson ISSN: 1097-6647 Impact factor: 5.364
Patient characteristics
| Variables | Patients | Controls |
|
|---|---|---|---|
|
|
| ||
| Age (years) | 30 [23–53] | 35 [26–43] | 0.41 |
| Gender (Male/Female) | 55/45 | 10/10 | 0.67 |
| Diagnosis | |||
| TM | 47 | ||
| SCA | 15 | ||
| Other hemoglobinopathy | 20 | ||
| MDS | 4 | ||
| Other | 18 | ||
| 1.5 T WB Heart T2* | 30.3 [22.6–36.8] | 32.3 [28.9–36.7] | 0.20 |
| 3 T WB Heart T2* | 18.5 [11.8–24.3] | 20.5 [18.3–24.3] | 0.14 |
| 1.5 T Liver T2* | 4.2 [1.8–8.6] | 25.8 [23.1–28.0] | <0.001 |
| 3 T Liver T2* | 2.3 [0.9–4.7] | 17.3 [14.8–21.4] | <0.001 |
Data are expressed as median [interquartile range]. SCA sickle cell anemia, MDS myelodysplasia syndrome, TM thalassemia major, WB white blood
Fig. 1a Association between cardiac T2* measurements at 1.5 T and 3 T using the white- blood sequence. The red line is the line of identity, and the black line is the best fit regression line. b Association between cardiac R2* measurements at 1.5 T and 3 T using the white-blood sequence. The red line is the line of identity, and the black line is the best fit regression line
Fig. 2a Association between cardiac T2* measurements at 1.5 T and at 3 T using the black-blood sequence. The red line is the line of identity, and the black line is the best fit regression line. b Association between cardiac R2* measurements at 1.5 T and 3 T using the black-blood sequence. The red line is the line of identity, and the black line is the best fit regression line
Fig. 3a Association between liver T2* measurements at 1.5 T and at 3 T. The red line is the line of identity, and the black line is the best fit regression line. b Association between liver R2* measurements at 1.5 T and at 3 T. The red line is the line of identity, and the black line is the best fit regression line
Fig. 4Comparison of artefacts scores for cardiac iron content assessment according to the field strength and the type of sequence used for image acquisition
Coefficients of variation and intra-class correlation for intra-observer reproducibility of T2* measurements according to acquisition sequence and magnetic field strength
| Heart | Liver | |||||
|---|---|---|---|---|---|---|
| WB | BB | |||||
| CoV | ICC (95 % CI) | CoV | ICC (95 % CI) | CoV | ICC (95 % CI) | |
| 1.5 T | 3.19 % | 0.999 (0.998–1.000) | 1.54 % | 1.000 (0.999–1.000) | 1.98 % | 0.953 (0.890–0.980) |
| 3 T | 3.54 % | 0.999 (0.997–1.000) | 3.62 % | 0.999 (0.999–1.000) | 3.62 % | 0.999 (0.997–1.000) |
BB Black Blood, CI Confidence Interval, CoV Coefficient of Variation, ICC Intra-class Correlation, WB White Blood
Fig. 5Bland-Altman plots for intra-observer variability (Panel a) and inter-observer variability (Panel b) of White-Blood T2* measurements at 3 T
Fig. 6Bland-Altman plots for intra-observer variability (Panel a) and inter-observer variability (Panel b) of Black-Blood T2* measurements at 3 T
Fig. 7Bland-Altman plots for intra-observer variability (Panel a) and inter-observer variability (Panel b) of Liver T2* measurements at 3 T
Coefficients of variation and intra-class correlation for inter-observer reproducibility of T2* measurements according to acquisition sequence and magnetic field strength
| Heart | Liver | |||||
|---|---|---|---|---|---|---|
| WB | BB | |||||
| CoV | ICC (95 % CI) | CoV | ICC (95 % CI) | CoV | ICC (95 % CI) | |
| 1.5 T | 3.34 % | 0.999 (0.998–1.000) | 2.76 % | 0.999 (0.999–1.000) | 4.24 % | 1.000 (0.999–1.000) |
| 3 T | 5.50 % | 0.998 (0.990–0.998) | 4.01 % | 0.999 (0.998–1.000) | 3.96 % | 0.999 (0.996–1.000) |
BB Black Blood, CI Confidence Interval, CoV Coefficient of Variation, ICC Intra-class Correlation, WB White Blood
Coefficients of variation and intra-class correlation for inter-study reproducibility of T2* measurements according to acquisition sequence and magnetic field strength
| Cardiac | Liver | |||||
|---|---|---|---|---|---|---|
| WB | BB | |||||
| CoV | ICC (95 % CI) | CoV | ICC (95 % CI) | CoV | ICC (95 % CI) | |
| 1.5 T | 5.55 % | 0.996 (0.989–0.998) | 4.55 % | 0.994 (0.985–0.998) | 5.69 % | 0.979 (0.949–0.992) |
| 3 T | 7.68 % | 0.970 (0.926–0.988) | 6.48 % | 0.996 (0.990–0.998) | 6.31 % | 0.998 (0.993–0.999) |
BB Black Blood, CI Confidence Interval, CoV Coefficient of Variation, ICC Intra-class Correlation, WB White Blood
Reproducibility of Black Blood T2* at 1.5 T and 3 T
| Test | Geometric Mean (95 % CI) of Squared difference |
| |
|---|---|---|---|
| T2* BB 1.5 T | T2* BB 3 T | ||
| Intra-Observer | 0.04 (0.02, 0.09) | 0.07 (0.03, 0.15) | 0.26 |
| Inter-Observer | 0.04 (0.01, 0.12) | 0.07 (0.02, 0.21) | 0.18 |
| Inter study | 0.26 (0.08, 0.80) | 0.28 (0.09, 0.82) | 0.94 |