| Literature DB >> 27375901 |
Marie Janaillac1, Sonia Labarinas1, Riccardo E Pfister1, Oliver Karam1.
Abstract
Background. In premature infants, maintaining blood partial pressure of carbon dioxide (pCO2) value within a narrow range is important to avoid cerebral lesions. The aim of this study was to assess the accuracy of a noninvasive transcutaneous method (TcpCO2), compared to blood partial pressure of carbon dioxide (pCO2). Methods. Retrospective observational study in a tertiary neonatal intensive care unit. We analyzed the correlation between blood pCO2 and transcutaneous values and the accuracy between the trends of blood pCO2 and TcpCO2 in all consecutive premature infants born at <33 weeks' gestational age. Results. 248 infants were included (median gestational age: 29 + 5 weeks and median birth weight: 1250 g), providing 1365 pairs of TcpCO2 and blood pCO2 values. Pearson's R correlation between these values was 0.58. The mean bias was -0.93 kPa with a 95% confidence limit of agreement of -4.05 to +2.16 kPa. Correlation between the trends of TcpCO2 and blood pCO2 values was good in only 39.6%. Conclusions. In premature infants, TcpCO2 was poorly correlated to blood pCO2, with a wide limit of agreement. Furthermore, concordance between trends was equally low. We warn about clinical decision-making on TcpCO2 alone when used as continuous monitoring.Entities:
Year: 2016 PMID: 27375901 PMCID: PMC4916268 DOI: 10.1155/2016/8041967
Source DB: PubMed Journal: Crit Care Res Pract ISSN: 2090-1305
Examples of concordance classification between blood pCO2 and transcutaneous pCO2.
| Example | Blood pCO2 | TcpCO2 | Trends' difference | Trends' concordance | ||||
|---|---|---|---|---|---|---|---|---|
|
|
| Trend |
|
| Trend | |||
| 1 | 5.2 | 5.6 | +7.7% | 5.2 | 5.2 | +0% | 7.7% | Good |
| 2 | 5.2 | 5.6 | +7.7% | 5.2 | 6.4 | +23.1% | 15.4% | Moderate |
| 3 | 5.2 | 5.6 | +7.7% | 5.8 | 4.6 | −21.4% | 29.1% | Poor |
Four examples of blood pCO2 and TcpCO2 trends within six-hour periods, with difference between trends, and the qualitative assessment of the concordance.
All pCO2 values are in kPa.
T 0: measures at baseline.
T 1: measures within a six-hour interval.
Figure 1Scatter plot of the blood pCO2 and the TcpCO2 of values measured <10 mn. Scatter plot of the blood pCO2 (x-axis) and the TcpCO2 (y-axis) of values measured <10 minutes apart. Values are in kPa. n = 1365. Pearson's R = 0.58 (p < .001).
Figure 2Bland-Altman plot of the average TcpCO2 and blood pCO2 values. Bland-Altman plot of the average TcpCO2 and blood pCO2 values (x-axis) and difference between TcpCO2 and blood pCO2 values (y-axis). n = 1365. Mean bias = −0.93 kPa and 95% confidence limit of agreement = −4.05 to +2.16 kPa.
Figure 3Scatter plot of the changes in blood pCO2 and the TcpCO2 six-hour trends. Scatter plot of the changes in blood pCO2 (x-axis) and the TcpCO2 (y-axis) six-hour trends. n = 313 pairs. Pearson's R = 0.25; p < .001.