Jong Min Park1,2,3,4, So-Yeon Park1,2,3,4, Jin Ho Kim1,2,3, Joel Carlson5, Jung-In Kim6,7,8,9.
Abstract
BACKGROUND: To investigate the effect of multi-leaf collimators (MLCs) with leaf width of 1.25 mm on the plan quality of volumetric modulated arc therapy (VMAT) for prostate cancer.Entities:
Keywords: Leaf width; Multi-leaf collimator; Prostate cancer; Volumetric modulated arc therapy
Mesh:
Year: 2016 PMID: 27338929 PMCID: PMC4917980 DOI: 10.1186/s13014-016-0664-0
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Dose volume constraints for target volumes as well as organs at risk
| Structure | Initial dose volume constraints | Relative priority |
|---|---|---|
| Primary plan | ||
| Primary PTV | D100% ≥ 99.5 % of the prescription dose | 150 |
| D98% ≥ 100 % of the prescription dose | 150 | |
| D2% ≤ 101 % of the prescription dose | 150 | |
| Dmax ≤ 102 % of the prescription dose | 150 | |
| Rectal wall | V44Gy ≤ 20 % | 100 |
| V30Gy ≤ 50 % | 100 | |
| Bladder | V44Gy ≤ 20 % | 100 |
| V30Gy ≤ 50 % | 100 | |
| Femoral heads | V23Gy ≤ 50 % | 80 |
| Boost plan | ||
| Boost PTV | D100% ≥ 99.5 % of the prescription dose | 150 |
| D98% ≥ 100 % of the prescription dose | 150 | |
| D2% ≤ 101 % of the prescription dose | 150 | |
| Dmax ≤ 102 % of the prescription dose | 150 | |
| Rectal wall | V22Gy ≤ 20 % | 100 |
| V18Gy ≤ 50 % | 100 | |
| Bladder | V22Gy ≤ 20 % | 100 |
| V18Gy ≤ 50 % | 100 | |
| Femoral heads | V12Gy ≤ 50 % | 80 |
Abbreviations: D Dose received by at least n% volume of the target volume, V the percent volume of a structure irradiated by at least k Gy, PTV planning target volume
Fig. 1Dose profiles by the simulated and actual multi-leaf collimators (MLCs). Dose profiles of the simulation of 1.25 mm multi-leaf collimator (MLC) with 2.5 mm width MLCs by CT image shift, real 2.5 mm MLCs and the simulation of 2.5 mm MLC with 5 mm width MLCs by CT image shift are plotted with solid, dashed and dotted lines, respectively. The profile of the simulation of 2.5 mm MLC was almost identical as that of the real 2.5 mm MLCs
Average dose-volumetric parameters of primary plans
| Original plan | Shifted plan |
| |
|---|---|---|---|
| Primary target volume | |||
| D1% (Gy) | 51.99 ± 0.36 | 51.71 ± 0.31 |
|
| Mean dose (Gy) | 50.70 ± 0.22 | 50.50 ± 0.20 |
|
| D99% (Gy) | 47.87 ± 0.36 | 47.94 ± 0.32 |
|
| D95% (Gy) | 49.33 ± 0.08 | 49.30 ± 0.07 | 0.107 |
| D5% (Gy) | 51.58 ± 0.33 | 51.32 ± 0.31 |
|
| V95% (%) | 99.07 ± 0.36 | 99.15 ± 0.35 | 0.478 |
| Conformity index | 1.06 ± 0.05 | 1.04 ± 0.04 |
|
| Homogeneity index | 0.07 ± 0.01 | 0.06 ± 0.01 |
|
| Normal tissue | |||
| Gradient measure (cm) | 1.86 ± 0.18 | 2.01 ± 0.26 |
|
| Monitor Unit (MU) | |||
| MU | 532 ± 115 | 572 ± 117 | 0.294 |
Abbreviations: D Dose received by at least n% volume of the target volume, V the percent volume of a structure irradiated by at least n% of the prescription dose
Fig. 2Dose-volume histograms (DVHs) of target volumes. Dose-volume histograms (DVHs) of a primary target volume from primary plans (a), boost target volume from boost plans (b), primary target volume from sum plans (c) and boost target volume from sum plans (d) calculated from a total of 20 volumetric modulated arc therapy (VMAT) plans for prostate cancer are shown. The average DVHs of original VMAT plans are plotted with solid black lines while those of the shifted VMAT plans are plotted with dashed black lines. Every single DVHs of original VMAT plans are plotted with solid gray lines while those of the shifted VMAT plans are plotted with dashed gray lines
Average dose-volumetric parameters of boost plans
| Original plan | Shifted plan |
| |
|---|---|---|---|
| Boost target volume | |||
| D1% (Gy) | 31.43 ± 0.24 | 31.26 ± 0.31 |
|
| Mean dose (Gy) | 30.71 ± 0.17 | 30.57 ± 0.17 |
|
| D99% (Gy) | 29.18 ± 0.10 | 29.25 ± 0.09 |
|
| D95% (Gy) | 29.92 ± 0.05 | 29.89 ± 0.04 | 0.095 |
| D5% (Gy) | 31.20 ± 0.22 | 31.02 ± 0.24 |
|
| V95% (%) | 99.24 ± 0.22 | 99.38 ± 0.21 |
|
| Conformity index | 1.02 ± 0.02 | 1.01 ± 0.02 |
|
| Homogeneity index | 0.06 ± 0.01 | 0.05 ± 0.01 |
|
| Normal tissue | |||
| Gradient measure (cm) | 1.66 ± 0.26 | 1.94 ± 0.47 |
|
| Monitor Unit (MU) | |||
| MU | 470 ± 59 | 495 ± 61 | 0.219 |
Abbreviations: D Dose received by at least n% volume of the target volume, V the percent volume of a structure irradiated by at least n% of the prescription dose
Average dose-volumetric parameters of sum plans
| Original plan | Shifted plan |
| |
|---|---|---|---|
| Boost target volume | |||
| D1% (Gy) | 83.00 ± 0.39 | 82.55 ± 0.34 |
|
| Mean dose (Gy) | 81.45 ± 0.32 | 81.12 ± 0.28 |
|
| D99% (Gy) | 77.98 ± 0.44 | 78.11 ± 0.37 |
|
| D95% (Gy) | 79.76 ± 0.33 | 79.68 ± 0.29 | 0.411 |
| Primary target volume | |||
| Mean dose (Gy) | 77.33 ± 2.22 | 77.04 ± 2.25 | 0.681 |
| D99% (Gy) | 55.11 ± 4.17 | 55.01 ± 4.23 | 0.267 |
| D95% (Gy) | 59.42 ± 6.27 | 59.16 ± 6.32 | 0.896 |
| Rectal wall | |||
| V70Gy (%) | 10.56 ± 3.08 | 10.38 ± 3.08 | 0.856 |
| V47Gy (%) | 32.40 ± 10.04 | 31.21 ± 9.51 | 0.700 |
| D50% (Gy) | 32.18 ± 11.35 | 31.22 ± 11.43 | 0.791 |
| Mean dose (Gy) | 36.68 ± 7.05 | 36.06 ± 6.97 | 0.783 |
| Bladder | |||
| V70Gy (%) | 6.44 ± 4.42 | 6.37 ± 4.36 | 0.958 |
| V47Gy (%) | 14.35 ± 8.62 | 14.19 ± 8.51 | 0.954 |
| D50% (Gy) | 12.58 ± 10.11 | 12.39 ± 10.00 | 0.952 |
| Mean dose (Gy) | 20.62 ± 9.52 | 20.43 ± 9.46 | 0.949 |
| Right femoral head | |||
| D50% (Gy) | 15.09 ± 2.31 | 14.88 ± 2.39 | 0.786 |
| Maximum dose (Gy) | 27.51 ± 4.10 | 27.54 ± 4.44 | 0.981 |
| Left femoral head | |||
| D50% (Gy) | 15.01 ± 3.55 | 14.79 ± 3.51 | 0.846 |
| Maximum dose (Gy) | 26.68 ± 3.62 | 27.34 ± 4.65 | 0.618 |
| Whole body including target volume | |||
| V81Gy (cc) | 60.92 ± 14.85 | 48.96 ± 11.68 |
|
| Body – primary target volume | |||
| V81Gy (cc) | 0.12 ± 0.13 | 0.07 ± 0.08 | 0.083 |
| V72.9Gy (cc) | 19.23 ± 4.74 | 18.77 ± 4.59 | 0.074 |
| V64.8Gy (cc) | 43.53 ± 9.61 | 43.08 ± 9.65 | 0.290 |
| V50.4Gy (cc) | 117.78 ± 23.83 | 117.06 ± 24.26 | 0.222 |
| V45.36Gy (cc) | 163.60 ± 31.56 | 163.02 ± 32.28 | 0.359 |
| V40.32Gy (cc) | 226.73 ± 42.88 | 226.39 ± 44.16 | 0.707 |
| V8.1Gy (cc) | 2963.81 ± 366.10 | 2979.18 ± 366.93 | 0.072 |
Abbreviations: D Dose received by at least n% volume of the target volume, V the percent volume of a structure irradiated by at least n Gy