| Literature DB >> 31196120 |
Jong Min Park1,2,3,4, Hong-Gyun Wu1,2,3,5, Hak Jae Kim1,2,3,5, Chang Heon Choi6,7,8, Jung-In Kim9,10,11.
Abstract
BACKGROUND: The aim of this study was to compare the plan quality of magnetic-resonance image-based intensity modulated radiation therapy (MRI-based-IMRT) with the MRIdian Linac system to that of volumetric modulated arc therapy (VMAT) with the TrueBeam STx system for lung stereotactic ablative radiotherapy (SABR).Entities:
Keywords: MR-IGRT; MR-linac; Planning study; SABR; VMAT
Mesh:
Year: 2019 PMID: 31196120 PMCID: PMC6567463 DOI: 10.1186/s13014-019-1314-0
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Plan information
| VMAT | IMRT with MR-linac |
| |
|---|---|---|---|
| Monitor unit (MU) | 4173 ± 439 | 8373 ± 1582 |
|
| Calculated beam on time (min) | 4.8 ± 1.0 | 18.4 ± 2.2 |
|
| Number of fields | 2 ± 0 | 8 ± 1 |
|
| Number of segments | N/A | 54 ± 7 | – |
Note: VMAT volumetric modulated arc therapy, IMRT intensity modulated radiation therapy, MR-linac magnetic resonance image linear accelerator
Dose-volumetric parameters of the PTV
| DV parameter | VMAT | IMRT with MR-linac |
|
|---|---|---|---|
| D98% (Gy) | 59.1 ± 0.3 | 60.3 ± 1.1 |
|
| D95% (Gy) | 60.0 ± 0.0 | 60.7 ± 1.2 |
|
| D5% (Gy) | 64.9 ± 1.3 | 63.8 ± 1.6 |
|
| D2% (Gy) | 65.3 ± 1.4 | 64.2 ± 1.7 |
|
| Minimum dose (Gy) | 54.6 ± 1.4 | 58.7 ± 1.3 |
|
| Maximum dose (Gy) | 66.8 ± 1.3 | 63.8 ± 1.5 |
|
| Mean dose (Gy) | 63.0 ± 0.8 | 62.3 ± 1.4 |
|
| PTV volume (cm3) | 27.2 ± 23.5 (ITV + 5 mm) | 6.3 ± 7.7 (GTV + 3 mm) |
|
| Homogeneity index | 0.099 ± 0.023 | 0.062 ± 0.014 |
|
Note: DV parameter dose-volumetric parameter, VMAT volumetric modulated arc therapy, IMRT intensity modulated radiation therapy, MR-linac magnetic resonance image linear accelerator, D dose received by at least n% volume of the planning target volume, PTV planning target volume, ITV internal target volume, GTV gross tumor volume
Fig. 1Representative patient dose distributions of volumetric modulated arc therapy and magnetic resonance image-based intensity modulated radiation therapy. Dose distributions in the axial, coronal, and sagittal planes of volumetric modulated arc therapy (VMAT) of representative patients, namely, patients A (a) and B (c), are shown. Those of magnetic resonance image-based intensity modulated radiation therapy (MRI-based IMRT) using a linear accelerator with a magnetic resonance imaging system are shown for patients A (b) and B (d). The planning target volume of VMAT were generated by adding an isotropic margin of 5 mm from the internal target volume, whereas those of MRI-based IMRT were generated by adding an isotropic margin of 3 mm from the gross tumor volume
Fig. 2Representative dose-volume histograms of the target volume and organs at risk of volumetric modulated arc therapy and magnetic resonance image-based intensity modulated radiation therapy plans. Dose volume histograms of the planning target volume, both lungs, and whole body from the volumetric modulated arc therapy and magnetic resonance image-based intensity modulated radiation therapy plans of representative patients, namely, patients A (a) and B (c) are shown. Those of organs at risk of patients A (b) and B (d) are also shown
Dose-volumetric parameters of OARs and whole body
| OAR | DV parameter | VMAT | IMRT with MR-linac |
|
|---|---|---|---|---|
| Bronchi | Maximum dose (Gy) | 24.2 ± 21.7 | 20.4 ± 19.0 |
|
| V34.8Gy (cm3) | 0.5 ± 2.0 | 0.0 ± 0.1 | 0.223 | |
| Esophagus | V30Gy (cm3) | 0.0 ± 0.0 | 0.0 ± 0.0 | – |
| Heart | Maximum dose (Gy) | 20.2 ± 16.0 | 19.1 ± 13.7 | 0.087 |
| V34Gy (cm3) | 0.7 ± 2.4 | 0.0 ± 0.1 | 0.168 | |
| Ipsilateral lung | Mean dose (Gy) | 5.7 ± 2.0 | 5.2 ± 2.1 |
|
| V20Gy (cm3) | 134.8 ± 81.4 | 116.8 ± 80.9 |
| |
| V10Gy (cm3) | 275.3 ± 139.3 | 244.7 ± 140.2 |
| |
| V5Gy (cm3) | 414.7 ± 182.0 | 358.1 ± 172.9 |
| |
| D1500cc (Gy) | 0.13 ± 0.25 | 0.24 ± 0.37 |
| |
| D1000cc (Gy) | 0.61 ± 1.15 | 0.76 ± 0.94 |
| |
| Contralateral lung | Mean dose (Gy) | 0.9 ± 0.4 | 1.0 ± 0.6 | 0.210 |
| V20Gy (cm3) | 0.0 ± 0.0 | 6.6 ± 20.1 | 0.146 | |
| V10Gy (cm3) | 0.5 ± 1.6 | 23.1 ± 44.3 |
| |
| V5Gy (cm3) | 23.4 ± 37.1 | 49.8 ± 72.1 |
| |
| D1500cc (Gy) | 0.06 ± 0.10 | 0.11 ± 0.14 |
| |
| D1000cc (Gy) | 0.21 ± 0.28 | 0.28 ± 0.17 |
| |
| Rib | Maximum dose (Gy) | 47.2 ± 16.5 | 43.0 ± 12.9 |
|
| V40Gy (cm3) | 7.7 ± 13.0 | 1.8 ± 3.9 |
| |
| Skin | Maximum dose (Gy) | 27.8 ± 9.4 | 33.0 ± 4.4 |
|
| V36Gy (cm3) | 0.1 ± 0.4 | 0.1 ± 0.2 | 0.167 | |
| Spinal cord | Maximum dose (Gy) | 11.0 ± 5.1 | 14.5 ± 5.6 |
|
| V26Gy (cm3) | 0.0 ± 0.0 | 0.0 ± 0.0 | – | |
| Whole body | Mean dose (Gy) | 1.2 ± 0.5 | 0.9 ± 0.3 |
|
| V50% (cm3) | 103.5 ± 82.8 | 78.7 ± 62.9 |
|
Note: OAR organ at risk, DV parameter dose-volumetric parameter, VMAT volumetric modulated arc therapy, IMRT intensity modulated radiation therapy, MR-linac magnetic resonance image linear accelerator, V percent volume receiving n Gy, D dose received by at least n cc volume of the planning target volume, V percent volume receiving n% of the prescription dose