Reduction of aberrant aortic haemodynamics following aortic root replacement with a mechanical valved conduit.

Eric J Keller1, S Chris Malaisrie2, Jane Kruse2, Patrick M McCarthy2, James C Carr1, Michael Markl3, Alex J Barker3, Jeremy D Collins4.   

Abstract

OBJECTIVES: Previous work suggests that aortic root and valve prostheses alter blood flow patterns in the ascending aorta, creating aberrant haemodynamics compared with those of healthy volunteers. Various valve designs have been proposed to better restore physiological haemodynamics. In this study, magnetic resonance imaging (MRI) was used to non-invasively assess three-dimensional (3D) ascending aortic haemodynamics after aortic root replacement (ARR) with a mechanical valved conduit postulated to create less turbulent blood flow.
METHODS: Ten patients (40 ± 9 years) underwent transthoracic echocardiography and contrast-enhanced multidimensional four-dimensional (4D) flow MRI at 1.5 T after ARR with an On-X mechanical valved conduit. Preoperative 4D flow MRI was available in 7 patients. Ten age- and gender-matched healthy volunteers (42 ± 13 years) were also analysed to characterize physiological flow. The presence of vortex/helix formation was graded by two blinded observers. Peak transvalvular pressure gradients were computed using the simplified Bernoulli equation. Patients' postoperative pressure gradients and helicity/vorticity grades were compared with preoperative gradients and those from healthy volunteers.
RESULTS: Intra- and interobserver ratings showed good agreement (κ = 0.93, P < 0.01 and κ = 0.84, P < 0.01, respectively). Highly helical and/or vortical flow was observed in all patients preoperatively, which was significantly reduced postoperatively (P < 0.01 and <0.01, respectively), restoring similar flow patterns similar to those seen in volunteers (P = 0.56 and 0.56). Peak transvalvular pressure gradients (ΔP) were also significantly reduced [43 ± 21 vs 12 ± 7 mmHg, P < 0.05 (Echo); 48 ± 22 vs 16 ± 9 mmHg, P < 0.05 (MRI)], but remained significantly higher than those of volunteers (6 ± 1 mmHg, P < 0.01).
CONCLUSIONS: Preliminary evidence suggests that ARR with an On-X mechanical valve significantly reduces aberrant aortic haemodynamics, producing flow patterns that resemble those in healthy volunteers.
© The Author 2016. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

Entities:  

Keywords:  Aortic valve replacement; Magnetic resonance imaging; Mechanical heart valve prosthesis

Mesh:

Year:  2016        PMID: 27245620      PMCID: PMC6390273          DOI: 10.1093/icvts/ivw173

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285


  24 in total

1.  Prosthetic replacement of the aorta is a risk factor for aortic root aneurysm development.

Authors:  Natascha Simon-Kupilik; Heinz Schima; Leopold Huber; Reinhard Moidl; Gerhard Wipplinger; Udo Losert; Ernst Wolner; Paul Simon
Journal:  Ann Thorac Surg       Date:  2002-02       Impact factor: 4.330

2.  Effect of mechanical aortic valve orientation on coronary artery flow: comparison of tilting disc versus bileaflet prostheses in pigs.

Authors:  Peter Kleine; Mirella Scherer; Ulf Abdel-Rahman; Armin A Klesius; Hans Ackermann; Anton Moritz
Journal:  J Thorac Cardiovasc Surg       Date:  2002-11       Impact factor: 5.209

3.  Flow visualization analysis in a model of artery-graft anastomosis.

Authors:  T Matsumoto; T Naiki; K Hayashi
Journal:  Biomed Mater Eng       Date:  1992       Impact factor: 1.300

4.  In vitro assessment of flow patterns and turbulence intensity in prosthetic heart valves using generalized phase-contrast MRI.

Authors:  John-Peder Escobar Kvitting; Petter Dyverfeldt; Andreas Sigfridsson; Stefan Franzén; Lars Wigström; Ann F Bolger; Tino Ebbers
Journal:  J Magn Reson Imaging       Date:  2010-05       Impact factor: 4.813

Review 5.  Prosthesis-patient mismatch: definition, clinical impact, and prevention.

Authors:  P Pibarot; J G Dumesnil
Journal:  Heart       Date:  2005-10-26       Impact factor: 5.994

6.  Time-resolved three-dimensional magnetic resonance velocity mapping of aortic flow in healthy volunteers and patients after valve-sparing aortic root replacement.

Authors:  Michael Markl; Mary T Draney; D Craig Miller; Jonathan M Levin; Eric E Williamson; Norbert J Pelc; David H Liang; Robert J Herfkens
Journal:  J Thorac Cardiovasc Surg       Date:  2005-08       Impact factor: 5.209

7.  Practical value of cardiac magnetic resonance imaging for clinical quantification of aortic valve stenosis: comparison with echocardiography.

Authors:  Shelton D Caruthers; Shiow Jiuan Lin; Peggy Brown; Mary P Watkins; Todd A Williams; Katherine A Lehr; Samuel A Wickline
Journal:  Circulation       Date:  2003-10-20       Impact factor: 29.690

Review 8.  Fluid mechanics of artificial heart valves.

Authors:  Lakshmi P Dasi; Helene A Simon; Philippe Sucosky; Ajit P Yoganathan
Journal:  Clin Exp Pharmacol Physiol       Date:  2009-02       Impact factor: 2.557

Review 9.  Molecular mechanisms responsible for the atheroprotective effects of laminar shear stress.

Authors:  Shi Pan
Journal:  Antioxid Redox Signal       Date:  2009-07       Impact factor: 8.401

10.  Time-resolved 3D MR velocity mapping at 3T: improved navigator-gated assessment of vascular anatomy and blood flow.

Authors:  Michael Markl; Andreas Harloff; Thorsten A Bley; Maxim Zaitsev; Bernd Jung; Ernst Weigang; Mathias Langer; Jürgen Hennig; Alex Frydrychowicz
Journal:  J Magn Reson Imaging       Date:  2007-04       Impact factor: 4.813

View more
  7 in total

1.  Seismocardiography and 4D flow MRI reveal impact of aortic valve replacement on chest acceleration and aortic hemodynamics.

Authors:  Ethan M I Johnson; Mozziyar Etemadi; S Chris Malaisrie; Patrick M McCarthy; Michael Markl; Alex J Barker
Journal:  J Card Surg       Date:  2019-10-15       Impact factor: 1.620

2.  Automatic correction of background phase offset in 4D-flow of great vessels and of the heart in MRI using a third-order surface model.

Authors:  Damian Craiem; Ariel F Pascaner; Mariano E Casciaro; Umit Gencer; Joaquin Alcibar; Gilles Soulat; Elie Mousseaux
Journal:  MAGMA       Date:  2019-06-22       Impact factor: 2.310

Review 3.  New imaging tools in cardiovascular medicine: computational fluid dynamics and 4D flow MRI.

Authors:  Keiichi Itatani; Shohei Miyazaki; Tokoki Furusawa; Satoshi Numata; Sachiko Yamazaki; Kazuki Morimoto; Rina Makino; Hiroko Morichi; Teruyasu Nishino; Hitoshi Yaku
Journal:  Gen Thorac Cardiovasc Surg       Date:  2017-09-19

4.  Perioperative evaluation of regional aortic wall shear stress patterns in patients undergoing aortic valve and/or proximal thoracic aortic replacement.

Authors:  Emilie Bollache; Paul W M Fedak; Pim van Ooij; Ozair Rahman; S Chris Malaisrie; Patrick M McCarthy; James C Carr; Alex Powell; Jeremy D Collins; Michael Markl; Alex J Barker
Journal:  J Thorac Cardiovasc Surg       Date:  2017-11-16       Impact factor: 5.209

5.  Differential flow improvements after valve replacements in bicuspid aortic valve disease: a cardiovascular magnetic resonance assessment.

Authors:  Malenka M Bissell; Margaret Loudon; Aaron T Hess; Victoria Stoll; Elizabeth Orchard; Stefan Neubauer; Saul G Myerson
Journal:  J Cardiovasc Magn Reson       Date:  2018-02-08       Impact factor: 5.364

Review 6.  Hemodynamic Profiles Before and After Surgery in Bicuspid Aortic Valve Disease-A Systematic Review of the Literature.

Authors:  Daniel G W Cave; Hannah Panayiotou; Malenka M Bissell
Journal:  Front Cardiovasc Med       Date:  2021-03-24

7.  Long-Term Comparison Between Pulmonary Homograft Versus Bioprosthesis for Pulmonary Valve Replacement in Tetralogy of Fallot.

Authors:  Lucia Cocomello; Marco Meloni; Filippo Rapetto; Mai Baquedano; Maria Victoria Ordoñez; Giovanni Biglino; Chiara Bucciarelli-Ducci; Andrew Parry; Serban Stoica; Massimo Caputo
Journal:  J Am Heart Assoc       Date:  2019-12-16       Impact factor: 5.501

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.