Literature DB >> 11845859

Prosthetic replacement of the aorta is a risk factor for aortic root aneurysm development.

Natascha Simon-Kupilik1, Heinz Schima, Leopold Huber, Reinhard Moidl, Gerhard Wipplinger, Udo Losert, Ernst Wolner, Paul Simon.   

Abstract

BACKGROUND: Noncompliant prostheses are used in aortic replacement. We hypothesized that this leads to increased distension and wall stress in the aortic root because of the loss of ventriculo arterial coupling.
METHODS: Pressure relations in the aortic root caused by changes of aortic elasticity simulating prosthetic aortic replacement were tested in a computer model. We then developed an in vitro model using porcine aortas and performed in vivo validation.
RESULTS: Findings in vitro and in vivo confirmed the predicted changes of the computer model. Pressure amplitude increased significantly by 17% after prosthetic replacement (p < 0.01). Pressure-time differential (Dp/dt) and dicrotic notch pressure amplitude both increased significantly. Echocardiography demonstrated systolic aortic root distension with percentage area change increasing in vitro from 28.2%+/-9.7% to 35.9%+/-10% (p < 0.05) and in vivo from 13.3%+/-3.1% to 24.3%+/-3.1% (p < 0.0001). Aortic root wall stress increased markedly.
CONCLUSIONS: Replacement of the aorta with vascular prostheses causes important negative alterations of hemodynamics and increases in wall stress.

Entities:  

Mesh:

Year:  2002        PMID: 11845859     DOI: 10.1016/s0003-4975(01)03423-3

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  1 in total

1.  Reduction of aberrant aortic haemodynamics following aortic root replacement with a mechanical valved conduit.

Authors:  Eric J Keller; S Chris Malaisrie; Jane Kruse; Patrick M McCarthy; James C Carr; Michael Markl; Alex J Barker; Jeremy D Collins
Journal:  Interact Cardiovasc Thorac Surg       Date:  2016-05-30
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.