In vitro assessment of flow patterns and turbulence intensity in prosthetic heart valves using generalized phase-contrast MRI.

John-Peder Escobar Kvitting1, Petter Dyverfeldt, Andreas Sigfridsson, Stefan Franzén, Lars Wigström, Ann F Bolger, Tino Ebbers.   

Abstract

PURPOSE: To assess in vitro the three-dimensional mean velocity field and the extent and degree of turbulence intensity (TI) in different prosthetic heart valves using a generalization of phase-contrast MRI (PC-MRI).
MATERIALS AND METHODS: Four 27-mm aortic valves (Björk-Shiley Monostrut tilting-disc, St. Jude Medical Standard bileaflet, Medtronic Mosaic stented and Freestyle stentless porcine valve) were tested under steady inflow conditions in a Plexiglas phantom. Three-dimensional PC-MRI data were acquired to measure the mean velocity field and the turbulent kinetic energy (TKE), a direction-independent measure of TI.
RESULTS: Velocity and TI estimates could be obtained up- and downstream of the valves, except where metallic structure in the valves caused signal void. Distinct differences in the location, extent, and peak values of velocity and TI were observed between the valves tested. The maximum values of TKE varied between the different valves: tilting disc, 100 J/m(3); bileaflet, 115 J/m(3); stented, 200 J/m(3); stentless, 145 J/m(3).
CONCLUSION: The TI downstream from a prosthetic heart valve is dependent on the specific valve design. Generalized PC-MRI can be used to quantify velocity and TI downstream from prosthetic heart valves, which may allow assessment of these aspects of prosthetic valvular function in postoperative patients. Copyright 2010 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 20432341     DOI: 10.1002/jmri.22163

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  15 in total

1.  Near valve flows and potential blood damage during closure of a bileaflet mechanical heart valve.

Authors:  L H Herbertson; S Deutsch; K B Manning
Journal:  J Biomech Eng       Date:  2011-09       Impact factor: 2.097

2.  Haemodynamic outcome at four-dimensional flow magnetic resonance imaging following valve-sparing aortic root replacement with tricuspid and bicuspid valve morphology.

Authors:  Edouard Semaan; Michael Markl; S Chris Malaisrie; Alex Barker; Bradley Allen; Patrick McCarthy; James C Carr; Jeremy D Collins
Journal:  Eur J Cardiothorac Surg       Date:  2013-12-06       Impact factor: 4.191

3.  Viscous energy loss in the presence of abnormal aortic flow.

Authors:  Alex J Barker; Pim van Ooij; Krishna Bandi; Julio Garcia; Mazen Albaghdadi; Patrick McCarthy; Robert O Bonow; James Carr; Jeremy Collins; S Chris Malaisrie; Michael Markl
Journal:  Magn Reson Med       Date:  2013-10-02       Impact factor: 4.668

4.  Reduction of aberrant aortic haemodynamics following aortic root replacement with a mechanical valved conduit.

Authors:  Eric J Keller; S Chris Malaisrie; Jane Kruse; Patrick M McCarthy; James C Carr; Michael Markl; Alex J Barker; Jeremy D Collins
Journal:  Interact Cardiovasc Thorac Surg       Date:  2016-05-30

Review 5.  4D flow imaging: current status to future clinical applications.

Authors:  Michael Markl; Susanne Schnell; Alex J Barker
Journal:  Curr Cardiol Rep       Date:  2014-05       Impact factor: 2.931

6.  In vitro evaluation of flow patterns and turbulent kinetic energy in trans-catheter aortic valve prostheses.

Authors:  Daniel Giese; Kilian Weiss; Bettina Baeßler; Navid Madershahian; Yeong-Hoon Choi; David Maintz; Alexander C Bunck
Journal:  MAGMA       Date:  2017-09-18       Impact factor: 2.310

7.  Measurement of cardiac valve and aortic blood flow velocities in stroke patients: a comparison of 4D flow MRI and echocardiography.

Authors:  Thomas Wehrum; Felix Guenther; Alexander Fuchs; Florian Schuchardt; Anja Hennemuth; Andreas Harloff
Journal:  Int J Cardiovasc Imaging       Date:  2018-01-11       Impact factor: 2.357

Review 8.  Comprehensive 4D velocity mapping of the heart and great vessels by cardiovascular magnetic resonance.

Authors:  Michael Markl; Philip J Kilner; Tino Ebbers
Journal:  J Cardiovasc Magn Reson       Date:  2011-01-14       Impact factor: 5.364

Review 9.  Advanced flow MRI: emerging techniques and applications.

Authors:  M Markl; S Schnell; C Wu; E Bollache; K Jarvis; A J Barker; J D Robinson; C K Rigsby
Journal:  Clin Radiol       Date:  2016-03-02       Impact factor: 2.350

10.  Comparison of Hemodynamics After Aortic Root Replacement Using Valve-Sparing or Bioprosthetic Valved Conduit.

Authors:  Jeremy D Collins; Edouard Semaan; Alex Barker; Patrick M McCarthy; James C Carr; Michael Markl; S Chris Malaisrie
Journal:  Ann Thorac Surg       Date:  2015-07-23       Impact factor: 4.330

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.