Literature DB >> 27195789

Standardized Loads Acting in Hip Implants.

Georg Bergmann1, Alwina Bender1, Jörn Dymke1, Georg Duda1, Philipp Damm1.   

Abstract

With the increasing success of hip joint replacements, the average age of patients has decreased, patients have become more active and their expectations of the implant durability have risen. Thus, pre-clinical endurance tests on hip implants require defining realistic in vivo loads from younger and more active patients. These loads require simplifications to be applicable for simulator tests and numerical analyses. Here, the contact forces in the joint were measured with instrumented hip implants in ten subjects during nine of the most physically demanding and frequent activities of daily living. Typical levels and directions of average and high joint loads were extracted from the intra- and inter-individually widely varying individual data. These data can also be used to analyse bone remodelling at the implant-bone interface, evaluate tissue straining in finite element studies or validate analytical loading predictions, among other uses. The current ISO standards for endurance tests of implant stems and necks are based on historic analytical data from the 1970s. Comparisons of these test forces with in vivo loads unveiled that their unidirectional orientations deviate from the time-dependent in vivo directions during walking and most other activities. The ISO force for testing the stem is substantially too low while the ISO force for the neck better matches typical in vivo magnitudes. Because the magnitudes and orientations of peak forces substantially vary among the activities, load scenarios that reflect a collection of time-dependent high forces should be applied rather than using unidirectional forces. Based on data from ten patients, proposals for the most demanding activities, the time courses of the contact forces and the required cycle numbers for testing are given here. Friction moments in the joint were measured in addition to the contact forces. The moment data were also standardized and can be applied to wear tests of the implant. It was shown that friction only very slightly influences the stresses in the implant neck and shaft.

Entities:  

Mesh:

Year:  2016        PMID: 27195789      PMCID: PMC4873223          DOI: 10.1371/journal.pone.0155612

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Total joint replacement of the hip is one of the most successful surgical procedures; in Germany alone, it is performed 200,000 times per year. Success rates are still very high, despite the fact that the patients have become substantially younger over the last two decades. In addition, patient expectations after surgery are high; many of them anticipate returning to their full activity level prior to the onset of arthritis and thus challenge the mechanics of the total hip replacement substantially. In the 1970s, many patients were happy to return to walking and stair climbing, while today patients are likely to practice more athletic activities. The currently available data do not reflect the increased activity level of patients, but realistic loads are essential for the pre-clinical assessments that lead to total hip replacement innovations.

Relevance for a reference load spectrum

Pre-clinical endurance tests must be performed on new implant designs of various implant sizes to prevent mechanical failure, for example, at the taper of modular designs. Finite element studies investigating interface mechanics can enable the development of novel implant geometries and surfaces that improve the adaptation of bone to the implant and prevent loosening, but only if a realistic load spectrum to which the constructs are subjected is known. Friction at the cup, in addition to contact forces, may endanger its fixation stability; this necessitates a more detailed characterisation of the load transfer in the joint. Analytical models of loads in joints, bones and muscles require various model parameters to be defined, but many of them are difficult to identify accurately; thus, the obtained results need to be verified by in vivo data. Realistic loads can be measured in subjects by using instrumented implants with telemetric data transmission. However, the loads vary within and between subjects and are highly dependent on the kind and level of activity. Rather than using an arbitrary single instance of in vivo data, e.g., from walking, it is therefore favourable to develop a standardized comprehensive reference load spectrum for the many conceivable test scenarios, of which only a select few were mentioned above.

ISO endurance tests on hip implants

Endurance tests of the stem are performed according to the ISO7206-4 standard. A sinusoidal force of 2300 N is applied under angles of 10° from medial and 9° from anterior relative to the stem axis. In the following, this test force is referred to as the ‘ISO stem force’. A similar approach is defined in ISO7206-6 for testing the endurance of the neck, but the force is increased to 5340 N (‘ISO neck force’). These load directions and magnitudes are based on historic joint contact force calculations [1-3], and a comparison to previously reported in vivo data [4, 5] suggests that the ISO test conditions need to be revised to better reflect the real in vivo loading conditions. In both ISO tests, 5*106 cycles have to be applied; in practice, 107 cycles are typically exerted. Fatigue of implant metals is determined by the most frequently acting loads, which cause cumulative damage over time. Implant metals can withstand static stresses that are approximately 2 to 3 times higher than the fatigue limits [6]. This means that rarely acting extreme forces, such as those during stumbling [7, 8], may cause a sudden fracture of the implant if their amplitudes are 2 to 3 times higher than those during frequent routine activities. A loading frequency between 1 Hz and 30 Hz is specified by the ISO tests. This indicates that the loading rate has no or only a minor influence on the fatigue properties of the stem and neck.

Previous standardised data on hip contact forces

In 1998, a first set of four patients was assessed, and our team measured telemetrically the hip contact forces and synchronous gait data during some activities of daily living [5]. The complete data were also published on a compact disc [4]. Typical ‘average’ and ‘high’ forces for testing hip implants were determined. However, the results were based on only four subjects, including one subject with an impaired walking ability. Now, an improved instrumented implant [9] enables contact force measurements in more physically capable subjects and additionally allows synchronous measurement of the friction moments in the joint. The movement data in [4] had been used by [10] to transform the load directions from the femur to the pelvis system. The transformed data and the appropriate loads from an averaged subject [4] are meant to be used for wear tests of hip implants. However, the standardized loads described here are much more representative than the data from 2001. The applicable relative movements between the pelvis and femur will be published in an additional paper.

Friction moments

In addition to high contact forces, friction moments in the joint may endanger the fixation of acetabular cups, as demonstrated by the frequent implant failures caused by cup loosening [11-13]. For an under-reaming of only 1 mm, a moment of 8.8 Nm was reported to loosen cementless cups after only 5 loading cycles [14], but even values as small as 2.2 Nm have been associated with cup failures [15]. Further statistical analyses suggested that even a more substantial under-reaming of 2 mm, and a moment as little as 3.3 Nm would be sufficient to cause fixation failure in 4.6% of the subjects [16]. Therefore, it is important to compare such stability limits with the real in vivo moments to identify daily activities that may affect cup fixation. In addition to the contact forces, friction moments also affect the loads within the femoral component. The in vivo friction data allow quantification of this effect and help determine whether the friction moments should be considered when testing neck or stem.

Goals of this study

Using the in vivo load data captured during demanding and frequent activities of daily living from ten subjects, this study had two main goals: To define reference loading conditions at the hip joint and report typical time courses and peak values of average and high contact forces and moments. To compare the high forces with those from the ISO standards for endurance tests on the neck and stem and, if required, propose more realistic test forces for the ISO standards, based on the new in vivo load spectrum.

Methods

Ethics statement

The study was approved by the Charité Ethics committee (EA2/057/09) and registered at the ‘German Clinical Trials Register’ (DRKS00000563). All patients gave written informed consent prior to participation in this study and to have their images published.

Investigated subjects

Ten subjects participated in this study (Table 1), all of whom suffered from coxarthrosis. Implantations were performed using a standard clinical procedure through a lateral approach. At the time of measurement (Table 1), the participants were in good physical condition and performed all activities without restriction. Only subject H5L had pain at the replaced contralateral hip joint, but with mild pain medication, she was able to perform all exercises without obvious limitations.
Table 1

Subject characteristics.

Age taken at time of implantation.

SubjectH1LH2RH3LH4LH5LH6RH7RH8LH9LH10R
Age[years]55615950626852555453
Sex[f / m]mmmmfmmmmf
Weight[kg]737592858784958011898
Height[cm]178172168178168176179178181162

Subject characteristics.

Age taken at time of implantation.

Measurements

Measurements were taken during the most frequent and highly demanding activities of daily living (Table 2). Based on activity measurements among patients with hip implants [17], these were walking on level ground at a self-selected speed, going up or down a staircase without using the handrail, sitting down or standing up and standing in a one-legged stance. Slow jogging at 7 km/h was additionally studied, as many enjoy this athletic activity and subjects increasingly perform it even after hip replacement [18]. Furthermore, cycling at a moderate power expenditure of 90 W was analysed because it is often practiced by patients with implants as an alternative to walking long distances or as fitness training [19]. Both activities may be performed with a high number of loading cycles and may therefore have a potential influence on implant fatigue.
Table 2

Activities.

ActivityMeasurement Conditions
CyclingPower = 90 Watts, Speed = 40 rpm, adapted saddle height
Sit Down / Stand UpWithout use of arm rest. Seat height = 45 cm
Knee Bend (Squat)Max. knee flexion angles = 51° to 95°, average = 73°
WalkingLevel walking, speed = 1.0–1.3 m/s; average = 1.1 m/s
StanceShifting weight from both to one leg and back
Stairs Up / Stairs DownWithout use of hand rail. Step height = 19.8 cm, width = 26.3 cm
Jogging7 km/h on treadmill.
All activities, measurement conditions, numbers of analysed trials and postoperative times of measurement are listed in Tables 2 and 3. Most data were captured 10 to 13 months after surgery. For subject H1L, three activities were not measured due to time restrictions of the patient, and no data from jogging were collected for three subjects who were not able to run with confidence at the target speed of 7 km/h. Throughout all activities, synchronously acquired motion capture data were used to determine walking speed and hip flexion angles.
Table 3

Postoperative time (in months) of measurement and number of investigated trials.

X/Y = X months postoperative / Y trials.

Activity → SubjectCyclingSit DownStand UpKnee BendWalkingStanceStairs UpStairs DownJogging
H1L13/134---13/3113/513/513/4-
H2R12/4212/712/711/812/4812/512/712/715/38
H3L18/2512/512/510/412/10112/412/812/7-
H4L12/113/63/6¾12/4712/612/612/712/55
H5L12/4012/412/410/712/4312/512/712/719/22
H6R12/1712/612/57/512/4412/612/1212/612/24
H7R15/8110/710/710/912/4212/512/1412/812/13
H8L18/3212/812/93/512/5112/712/1612/818/8
H9L13/2113/513/613/313/7213/513/713/713/53
H10R12/2912/512/58/812/7312/6---

Postoperative time (in months) of measurement and number of investigated trials.

X/Y = X months postoperative / Y trials. In our subjects, the average resultant contact force during walking became by 1% lower from the first to the twelfth postoperative month [20]. Within the following 21 months (average), it rose by less than 7% (not published). Therefore, the data presented here are assumed to be representative for the whole postoperative time.

Instrumented implant

A clinically proven titanium implant (CTW, Merete Medical, Berlin, Germany) with a 32 mm ceramic head that articulated with a press-fit cup with XPE inlay was modified to measure the forces and moments acting at the head. The electronics in the hollow implant neck was powered inductively by a coil around the hip joint and was equipped with six strain gauges and a 9-channel amplifier with telemetric data transfer. The electronics were safely encapsulated by welding. This system provided real-time monitoring of the 3D forces and moments with an accuracy of 1–2%. The subject’s activities were video-taped synchronously to the measured load data. The CCD angle of the implants was always 135°; the neck length L was adapted individually (Fig 1, Table 4). The implant, measuring system and calibration are described elsewhere in detail [9, 21, 22].
Fig 1

Coordinate system of right femur and implant.

x, y, z = axes of femur coordinate system. x = parallel to posterior contour of condyles. P1 = intersection of neck axis and femoral midline. P2 = middle of intercondylar notch. z = straight femur axis between P1 and P2. Force components Fx, Fy and Fz act in directions x, y and z. Moment components Mx, My and Mz turn clockwise around x, y and z. The implant is turned clockwise by angles αz, αy and αx around the femur axes z, y and x. αz = anteversion of neck (negative). x’, y’, z’ = axes of implant. xn, yn, zn = coordinate system at distal end of implant neck. xs, ys, zs = coordinate system of stem 80 mm below head centre.

Table 4

Rotation angles for transforming loads from the femur- to the implant-based coordinate system and implant neck lengths.

αx, αy and αz = orientation of the implant relative to the femur coordinate system in frontal, sagittal and transverse planes (see Fig 1). L = neck length from head centre to intersection of neck and stem axis (see Fig 1). Data from 10 individual subjects and their average.

Implant OrientationNeck Length
Relative to Femur System
Subjectαx [°]αy [°]αz [°]L [mm]
H1L2.3-2.3-15.055.6
H2R4.10.6-13.859.3
H3L4.0-3.0-13.855.6
H4L7.5-1.7-18.963.3
H5L4.0-2.3-2.355.6
H6R5.8-1.7-31.055.6
H7R6.3-1.7-2.463.3
H8L4.6-1.7-15.559.3
H9L4.60.6-2.359.3
H10R1.7-1.2-9.759.6
Average4.5-1.4-12.558.7

Coordinate system of right femur and implant.

x, y, z = axes of femur coordinate system. x = parallel to posterior contour of condyles. P1 = intersection of neck axis and femoral midline. P2 = middle of intercondylar notch. z = straight femur axis between P1 and P2. Force components Fx, Fy and Fz act in directions x, y and z. Moment components Mx, My and Mz turn clockwise around x, y and z. The implant is turned clockwise by angles αz, αy and αx around the femur axes z, y and x. αz = anteversion of neck (negative). x’, y’, z’ = axes of implant. xn, yn, zn = coordinate system at distal end of implant neck. xs, ys, zs = coordinate system of stem 80 mm below head centre.

Rotation angles for transforming loads from the femur- to the implant-based coordinate system and implant neck lengths.

αx, αy and αz = orientation of the implant relative to the femur coordinate system in frontal, sagittal and transverse planes (see Fig 1). L = neck length from head centre to intersection of neck and stem axis (see Fig 1). Data from 10 individual subjects and their average.

Measured load components and coordinate systems

The implant measured the three contact force components Fx, Fy and Fz in Newton and the three moment components Mx, My and Mz in Nm (Fig 1). The moments are caused by friction in the joint and act in addition to the forces. The resultant force Fres and the resultant friction moment Mres were calculated from the vector sums of their components. For the remainder of the study, the term ‘load’ refers to the complete set of all six components and their two resultants, unless otherwise stated. If a force is mentioned without further description, it refers to the peak value during the whole loading cycle. All loads are reported in the x, y, z coordinate system, which is defined relative to a right-sided femur (Fig 1). This is advantageous because then the loads relative to the implant can be recalculated from that data for any prosthesis which is differently oriented relative to the femur, e.g., has a different anteversion. Data from left-sided implants were mirrored to the right side. The origin of this coordinate system is located at the centre of the femoral head. The +z axis points upward and is defined by the line connecting the two points where the curved femoral mid-line intersected with the neck axis (P1) and where it passes the intercondylar notch (P2). The +x axis points laterally and is oriented parallel to the proximal contour of the condyles. The +y axis points in the anterior direction. This definition is in accordance with the ISB recommendations [23] but deviates from our earlier definition [24] for the left hip joint, which defined the +x axis to point medially [4, 5, 8].

Transformation of loads

To test fatigue of a non-implanted prosthesis, as in the ISO tests, the loads are mostly applied in an implant-based coordinate system. To set up such a system, the orientations of all implants relative to the femur were determined from CT images and inter-individually averaged. The implant coordinate system (xi, yi and zi) is fixed at the implant head but has rotated axes relative to the femur system (Fig 1). It is rotated clockwise by the angles αz, αy and αx around the +z, +y and +x axes. The angle αz is the negative anteversion angle of the implant; the other two angles are related to the curvature of the femur in the frontal and sagittal plane. To obtain the force vector Fi and the moment vector Mi in the implant system, the femur-based force F and moment M must be multiplied with the transformation matrix T: T is given by The implant rotations for the individual subjects and their averages are given in Table 4. The average anteversion was -αz = 12.5°, which was nearly identical to 12.6°, reported as being ‘normal’ (Sutter et al., 2015). To transform the reported loads from the femur to the implant system of our average subject, the transformation matrix T takes the following form:

Calculation of average loads AVER75

The following calculations were performed separately for all activities. First, all measured loads were linearly adjusted from the individual patient’s weight to an average body weight (BW) of 75 kg; this weight approximately corresponds to the average weight of the European population above the age of 60 years (Nordic Council of Ministers, 2011). Then, the time-functions of Fres from all trials from the same subject and activity were averaged. The employed ‘time warping’ procedure averaged the cycle durations and then distorted the single trial times such that the summed squared error between all trials and their average was minimized [25]. This algorithm was formulated to best preserve the maxima and minima of the input signals. After morphing the trial times, all patterns of Fres were averaged arithmetically. Morphing was based on Fres because it contains time-dependent characteristics of all three force components. For each trial the obtained time distortions of Fres were then applied to all six corresponding force and moment components to preserve their synchronicity before they were also averaged. In the remainder of this study, forces and moments from single subjects always refer to these individual averages. The same procedure was then repeated with the individual averages of Fres (Fig 2). The resulting six load components were the average loads ‘AVER75’ for an average subject with a BW of 75 kg.
Fig 2

Average force AVER75 and calculation of high loads HIGH100.

Resultant forces Fres from walking. Coloured lines = forces from 10 subjects with normalized BW = 75 kg. Black solid line = AVER75 force, represents subject of 75 kg with average load levels. Subject H7R had the highest peak value (red circle). AVER75 was multiplied by factor f1, to obtain a peak equal to the peak of H7R (black dotted line). The obtained force was further multiplied by f2 = 1.33 = 100 kg / 75 kg to obtain the HIGH100 load (black dashed line). The factor fAH = f1*f2, obtained from the analysis of Fres, was then applied to all 6 load components. HIGH100 represents subjects of 100 kg with high load levels. For explanation of factors f1 and f2, please also see Fig 3.

Average force AVER75 and calculation of high loads HIGH100.

Resultant forces Fres from walking. Coloured lines = forces from 10 subjects with normalized BW = 75 kg. Black solid line = AVER75 force, represents subject of 75 kg with average load levels. Subject H7R had the highest peak value (red circle). AVER75 was multiplied by factor f1, to obtain a peak equal to the peak of H7R (black dotted line). The obtained force was further multiplied by f2 = 1.33 = 100 kg / 75 kg to obtain the HIGH100 load (black dashed line). The factor fAH = f1*f2, obtained from the analysis of Fres, was then applied to all 6 load components. HIGH100 represents subjects of 100 kg with high load levels. For explanation of factors f1 and f2, please also see Fig 3.
Fig 3

Peak forces Fres from nine activities.

Coloured lines = individual loads from 10 subjects with normalized BW = 75 kg. Black solid line = average loads AVER75 for BW = 75 kg. Black dashed line = high loads HIGH100 for BW = 100 kg. Factors f1 and f2 are shown for the data from standing up; for other activities, they are different. See also Fig 2.

Calculation of high loads HIGH100

Separately for each activity the highest peak value was identified among the individual load-time functions of Fres (red circles in Figs 2 and 3). The AVER75 function was then multiplied by a factor, f1, which raised its maximum to this peak value. Multiplication by the additional factor f2 = 1.33 = 100kg/75kg yielded the high loads ‘HIGH100’. HIGH100 data represent a subject which has the highest peak force relative to all others and a body weight of 100 kg. The resultant multiplication factor fAH = f1 * f2 (Table 5, bottom), determined from the analysis of Fres, was then applied to all six load components.
Table 5

Magnitudes and directions of AVER75 and HIGH100 peak forces and moments and the conversion factor fAH.

Average = average of peak force Fres from all subjects. Range = range from 10 single subjects. Range [%] = Range in percent of average. αfront, αsagit and αtrans = angles of the peak force vector in frontal, sagittal and transverse planes (see Fig 1). These angles are the same for AVER75 and HIGH100 loads. fAH = factor for calculation of AVER75 from HIGH100 loads for all 6 load components and their resultants: AVER75 = HIGH100 / fAH. a/b = absolute values except from cycling and jogging. * = Mres is low at the first peak of Fres.

Activity → Load TypeCyclingSit DownStand UpKnee BendWalkingStanceStairs UpStairs DownJogging
Peak Value of Fres
 AVER75 Average [N]7311360b160016991925207722322300a3065
 AVER75 Range [N]416–942891–22021286–28801122–23591613–21611753–25052046–27051988–29072613–3630
 AVER75 Range [%]72961007328 b36304033
 HIGH100 Average [N]12562935383931452880 b334036063875 a4839
Mres at Instant of Peak Fres
 HIGH100 Average [Nm]1.111.922.51a2.490.931.031.930.66 b*0.80
Peak Value of Mres
 HIGH100 Average [Nm]1.152.092.77a2.571.761.722.002.73*1.60
Angle αfront of Fres in Frontal Plane
 Range [°]15–2713–2616–2917–289–218–2114–2513–2413–20
 Average [°]202024 a231716 b2016 b15
Angle αsagit of Fres in Sagittal Plane
 Range [°]9–292–192–144–250–19-2–810–236–187–22
 Average [°]171188114 b18 a1115
Angle αtrans of Fres in Transverse Plane
 Range [°]22–566–534–347–52-1–49-2–3424–5619–5026–61
 Average [°]402817183312 b42 a3344
Factor fAH [1]1.732.172.43 a1.851.55 b1.641.661.711.59

Peak forces Fres from nine activities.

Coloured lines = individual loads from 10 subjects with normalized BW = 75 kg. Black solid line = average loads AVER75 for BW = 75 kg. Black dashed line = high loads HIGH100 for BW = 100 kg. Factors f1 and f2 are shown for the data from standing up; for other activities, they are different. See also Fig 2.

Magnitudes and directions of AVER75 and HIGH100 peak forces and moments and the conversion factor fAH.

Average = average of peak force Fres from all subjects. Range = range from 10 single subjects. Range [%] = Range in percent of average. αfront, αsagit and αtrans = angles of the peak force vector in frontal, sagittal and transverse planes (see Fig 1). These angles are the same for AVER75 and HIGH100 loads. fAH = factor for calculation of AVER75 from HIGH100 loads for all 6 load components and their resultants: AVER75 = HIGH100 / fAH. a/b = absolute values except from cycling and jogging. * = Mres is low at the first peak of Fres. In the remainder of this study, HIGH100 loads are generally reported, but the spatial directions of the AVER75 forces and moments are identical to those of the HIGH100 loads. The magnitudes and time courses of all AVER75 components and resultants can be obtained by dividing the HIGH100 data by fAH. This factor differs between the activities and depends on the inter-subject variability of the Fres peak values. The more uniformly an exercise is performed across the cohort, the smaller the difference is between the AVER75 and HIGH100 load magnitudes. This influence of activity is apparent when comparing standing up and walking. Standing up showed larger peak force variations and therefore larger differences between the AVER75 and HIGH100 load magnitudes than walking (Fig 3). The BW used to calculate the HIGH100 loads was not unrealistically high. The 95% percentile of BW for the average European population from 65–74 years was reported to be 100.5 kg (Nordic Council of Ministers, 2011). Moreover, the 85% percentile of BW for the average male population (70–79 years) in the United States was reported to be 100 kg [26]. If the 95% percentile of BW for the male US population (60–69 years) shall be respected, then even a BW of approximately 125 kg would be realistic, i.e., all HIGH100 should be increased by 25%. The friction moments were averaged in the same way as the forces, based on the same time distortions of Fres. The maxima of the HIGH100 friction moments were compared to published stability limits of cementless acetabular cups to determine whether they were high enough to potentially cause cup loosening and which activities were the most dangerous ones.

Internal moments in neck and stem when applying ISO or HIGH100 forces

Three exemplary internal moments in the femoral component were calculated for all activities (Fig 1, right side): The bending moment Mbn at the distal end of the neck is given by the vector sum of moments Mxn around axis xn plus Myn around yn. Mbn quantifies most of the stress in natural, fracture-stabilized or implant necks and should be generated by the ISO neck test. The torque Mts around the stem axis zs may influence the initial torsional stability of cementless implants in the femur. The total bending moment Mbs in the stem is determined 80 mm below the head centre and is most decisive for the stem endurance in the case of a proximally resorbed femur. Mts and Mbs should both be realistically generated by the ISO stem test. The three moments depend not only on the magnitude but also on the spatial direction of Fres relative to the implant. Mbs is especially sensitive to small directional changes in the three planes (Fig 1) because it depends on the spatial lever arm between Fres and the location for which Mbs is determined. This dependency on the force directions relative to the femur is equivalent to orientation changes of the implant relative to the bone (αx, αy, and αz; Fig 1). The first two angles depend on the curvature of the femur in two planes, and αz is the anteversion chosen by the surgeon. To quantify this sensitivity, all three angles were modified by ±3° and the largest changes of the three moments during walking were determined, caused by any combined angle modification.

Influence of friction moments on loads in neck and stem

When fatigue testing femoral components, it is unknown whether friction moments can be neglected. To answer this, the three moments in the implant were calculated for all activities under the assumption that only the contact forces were acting at the head. Then, their changes due to the additional friction moments were determined. This was performed for the instance of peak Fres, based on the average geometry and orientation of the implants (Table 4).

Comparison of ISO test forces with in vivo HIGH100 forces

The magnitude and orientation of the ISO test forces were compared to those of the HIGH100 forces from all activities. Additionally the peak values of Mbn, Mbs and Mts in the implant, which are caused by the ISO forces, were compared with their in vivo values during the various activities. This shows whether the test forces realistically simulate the in vivo loading in the neck and shaft.

Results

Peak forces AVER75

The AVER75 peak values of Fres varied considerably between subjects (example for walking in Fig 2) and activities (Fig 3; Table 5). AVER75 values during double-legged activities (sitting down, standing up, knee bend) were lower than those during all single-legged exercises, except cycling. Among the single-legged activities, walking caused the lowest forces. The force acting when going down stairs was slightly higher than that during stair climbing. For the single-legged activities, the ranges of individual peak forces were 28%– 40% of the average AVER75 values (Table 5). The smallest range (28%) was measured during walking, the most frequent cyclic activity. Cycling had a small average AVER75 value and therefore its individual AVER75 range (73%) was much higher. For the double-legged activities, the range of peak Fres was 73%– 100% of the averages, which reflected the large individual variability in left or right support during these exercises.

Peak forces HIGH100

The HIGH100 peak values of Fres consistently exceeded their corresponding AVER75 values by more than a factor f2 of 1.33 (Table 5), the expected rise when only increasing the BW from 75kg to 100kg (Figs 2 and 3). Across the different activities, the factor fAH varied between 1.55 and 2.43 and reflected the amount of variation in peak force between subjects. The multiplier fAH was smallest for walking and generally larger for the three double-legged activities. The increased values were caused by the AVER75 forces of subject H5L (the patient with a somewhat painful contralateral hip joint), which were much higher than for all other subjects (Fig 3). As a result, the HIGH100 peak force of 2880 N during walking was the lowest from all daily activities except cycling. When excluding cycling and jogging, the HIGH100 peak forces all laid in a range from 2880 N to 3875 N. The force during jogging (4839 N) was 68% higher than during walking and reached nearly five times the BW.

Time course of AVER75 forces

The individual magnitude of Fres during walking varied substantially, but the general shape of the time courses were very similar across all subjects (Fig 2). When walking, a double-peak pattern was observed for Fres, with the first maximum, on average, only slightly higher than the second one. The first force peak occurred approximately at the time of contralateral toe-off, while the second peak was associated with the contralateral heel strike. The AVER75 time courses of all six load components during walking and all other activities are identical to those of the HIGH100 patterns if the latter are divided by the activity-specific factor fAH (Table 5).

Time courses of HIGH100 forces

In Fig 4, the three HIGH100 force components and Fres during all activities are charted over time; the maxima and minima of the components are listed in Table 6. The extrema of single force components did not always occur at the same time as the maximum of Fres (Table 5).
Fig 4

High100 forces during nine activities and ISO forces.

High forces for a simulated subject with a body weight of 100 kg. Components Fx, Fy and Fz and resultant force Fres in the femur coordinate system.

Table 6

Minima and maxima of HIGH100 force and moment components.

Minima/Maxima within whole loading cycles. a/b = values from all activities except cycling and jogging. All moment components changed their signs within the cycle. See also Fig 4.

Activity → Load ComponentCyclingSit DownStand UpKnee BendWalkingStanceStairs UpStairs DownJogging
ForceFx323/420442/1002472/1578a428/1202265/837b347/956379/1164329/1085311/1241
[N]Fy-357/-58-541/-88-612/70-455/-52-536/24-203b/-50-1067a/-68-855/-78-1222/-21
-Fz392/1131728/2709b753/34801069/2882235/2709b644/3198558/3876a405/3662229/4519
MomentMx-0.60/1.11-2.03/1.57-1.11b /2.61a-1.97/2.39-1.28/1.47-1.33/1.41-1.21/1.82-2.06a /1.19b-1.44/0.97
[Nm]My-0.22/0.190.00/0.77b-0.70/0.780.31/0.93-0.37/1.21-0.71/1.71a-0.51/1.24-0.88a /1.50-0.49/1.07
Mz-0.25/0.25-0.57/0.34-0.22b/0.70-0.63/0.43-0.50/0.63-0.55/0.93a-0.44/0.44-1.04a/0.30b-0.72/0.13

High100 forces during nine activities and ISO forces.

High forces for a simulated subject with a body weight of 100 kg. Components Fx, Fy and Fz and resultant force Fres in the femur coordinate system.

Minima and maxima of HIGH100 force and moment components.

Minima/Maxima within whole loading cycles. a/b = values from all activities except cycling and jogging. All moment components changed their signs within the cycle. See also Fig 4. Changes in the forces generally followed smooth patterns, except for the single-legged stance, which required considerable muscle activity to maintain balance. Walking generated the smallest absolute magnitudes of the components Fx and Fz (Table 6, first and third lines) when cycling was excluded. For Fy, walking generated the third smallest absolute magnitudes, behind those generated during stance and knee bend. Fx was always positive throughout all activities (Fig 4), indicating that the contact force permanently acted from medial on the joint. The highest value of Fx was not found for the activity with the highest Fres (jogging) but for standing up, followed by jogging and knee bend (Table 6). It must be noted that HIGH100 loads describe the average across subjects. Individual data fluctuated around the average numbers. Fy was generally negative throughout the whole cycle except during walking and standing up, when very small positive values were observed for short periods of time. The joint was therefore nearly always loaded from anterior. When going up stairs, the peaks of Fy were close to the magnitudes during jogging. Because the contact force always acts downward, -Fz was consistently positive. The HIGH100 time course patterns of all load components for walking, the most frequent high-demand daily activity, are combined in Fig 5.
Fig 5

HIGH100 loads during one gait cycle of walking.

Top = 3 force components and resultant force. Bottom = 3 moment components and resultant moment. To obtain average loads AVER75 instead of the high loads HIGH100, all components must be divided by the factor fAH (Table 5).

HIGH100 loads during one gait cycle of walking.

Top = 3 force components and resultant force. Bottom = 3 moment components and resultant moment. To obtain average loads AVER75 instead of the high loads HIGH100, all components must be divided by the factor fAH (Table 5).

HIGH100 friction moments

The moment components (Fig 6) differed much more between the activities than the force components. All components changed their signs, or at least reached zero, during the loading cycles of all activities (Table 6).
Fig 6

High100 moments during nine activities.

High loads for a simulated subject with a body weight of 100 kg. Components Fx, Fy and Fz and resultant force Fres.

High100 moments during nine activities.

High loads for a simulated subject with a body weight of 100 kg. Components Fx, Fy and Fz and resultant force Fres. The instability of the single-legged stance was reflected by excessive, random variations of all moment components. The multiple high moment peaks in all three planes did not occur at identical time points. Therefore, these high peak components did not result in equally excessive peaks of the resultant moment, Mres. Positive values of Mx are defined as occurring during extension. The highest positive Mx values were found for standing up, getting up from knee bend and stair ascending, i.e., for activities involving large extending movements (Table 6, Fig 6). Often, the absolute minima of Mx were in the same order of magnitude as the maxima or even larger. During jogging, the maxima of Mx were lower than during all other activities and even lower than during cycling. Six activities generated values similar to or higher than the (absolute) Mx minimum during jogging. The highest positive values of My were observed when standing, followed by going down and up a staircase. During several activities, the minima and maxima of My exceeded those during jogging. The highest negative value of Mz occurred when going down stairs; it far exceeded the values during jogging and all other activities. The highest positive value of Mz was observed during the single-leg stance. In contrast to the force components, where the axial component–Fz dominated the two transverse components and nearly equalled the resultant force Fres, all three moment components had similar magnitudes and equally affected the resultant moment Mres. The highest peak values of Mres were found for standing up, descending stairs and knee bend (Table 5). If cycling was disregarded, jogging caused a lower Mres peak than all other activities; here, the moment was 6% or 9% lower than during walking or standing, respectively. Mres did not always reach its maximum at the same instant as Fres (Fig 6). When going down stairs, Mres was very low at that instant (* in Table 5).

Force directions

During all activities, the directions αfront and αsagit of peak Fres did not change much throughout the single loading cycles. This limited range of load directions was most pronounced in the frontal plane and for high forces. During walking), for instance, high forces above half of their peak value had directions which only varied by 8° in the frontal and 12° in the sagittal plane(dashed lines in Fig 7. The lower the force, the less its direction remained constant.
Fig 7

Peak force and path of force vector during one cycle of walking.

HIGH100 forces from the average subject. Left = frontal plane. Middle = sagittal plane. Right = transverse plane. The vector acts from the green trajectory towards the head centre. ‘Start’ and circle = point on trajectory at the instant of heel strike. Red dots = directions of the peak force. Dashed lines = limits of force directions throughout the cycle for forces > 1500 N. Angles αfront, αsagit, αtrans = angles of the force vector in 3 planes.

Peak force and path of force vector during one cycle of walking.

HIGH100 forces from the average subject. Left = frontal plane. Middle = sagittal plane. Right = transverse plane. The vector acts from the green trajectory towards the head centre. ‘Start’ and circle = point on trajectory at the instant of heel strike. Red dots = directions of the peak force. Dashed lines = limits of force directions throughout the cycle for forces > 1500 N. Angles αfront, αsagit, αtrans = angles of the force vector in 3 planes. Small differences in the load directions were also found when the average peak forces from different activities were compared (Fig 8). The load angles varied by only 9° in the frontal and 14° in the sagittal plane. In the transverse plane, the range was considerably larger, reaching a value of 33°.
Fig 8

Magnitude and direction of HIGH100 peak forces during 9 activities.

High loads HIGH100 from the average subject. Left = frontal plane. Middle = sagittal plane. Right = transverse plane. Vectors of peak forces act from the symbols (different activities) towards the head centre. Red vector = direction during walking. Small red dots = AVER75 peak forces from individual subjects during walking, indicating the inter-individual variation of force directions for the same activity.

Magnitude and direction of HIGH100 peak forces during 9 activities.

High loads HIGH100 from the average subject. Left = frontal plane. Middle = sagittal plane. Right = transverse plane. Vectors of peak forces act from the symbols (different activities) towards the head centre. Red vector = direction during walking. Small red dots = AVER75 peak forces from individual subjects during walking, indicating the inter-individual variation of force directions for the same activity. However, the extent of force direction variation was even larger between the single subjects than between different activities. This is illustrated by the small red dots in Fig 8, indicating the individual peak force directions during walking. For the various activities, the force directions individually varied by 7° to 13° in the frontal and 6° to 21° in the sagittal plane (Table 5). The peak magnitudes of the resultant and all three components of the ISO stem force were much smaller than in the HIGH100 data during every activity except cycling (Fig 4). They did not even simulate the magnitudes during walking. In contrast, the magnitude of the ISO neck force was even larger than the highest in vivo magnitude, found during jogging. However, especially the Fx component of the ISO neck force was substantially smaller than in vivo. The inadequacy of Fx, as defined in the ISO tests, was further reflected in the direction of the ISO force in the frontal plane (Fig 8, left). While the ISO test proposes that the force should form an angle of 10° with the z-axis, the in vivo data suggested that this angle should be between 15° and 24° (Table 5). For simulating walking, the most frequent and demanding activity, this angle should have been 17°. In the sagittal plane (Fig 8, middle) the load angle of both ISO forces (7°) fell within the in vivo range determined across all activities (4° to 18°), but was smaller than the in vivo value during walking (11°). The internal torque Mts = 16.0Nm in the stem (Table 7), caused by the ISO stem force, was only sufficient to simulate the in vivo conditions during cycling, knee bend and stance. However, it was up to 66% smaller than the in vivo magnitudes during walking, ascending or descending stairs. Similar findings were observed for the bending moment Mbs in the stem, which was 69.6 Nm using the ISO stem force. Although it was large enough to simulate walking, it was substantially lower than the in vivo values when jogging or using a staircase.
Table 7

Bending of implant neck and torque around stem.

HIGH100 data at the instance of peak force Fres. Mbn = bending moment at base of neck (Fig 1) without friction, only caused by Fres. Mts = torque around stem without friction. Mbs = bending moment in stem without friction. ΔMbn, ΔMts and ΔMbs = changes due to additionally acting friction moments at head. ISO Neck/Stem = ISO 7206-6/4 test forces.

Activity → MomentsCyclingSit DownStand UpKnee BendWalkingStanceStairs UpStairs DownJoggingISO StemISO Neck
Mbn [Nm] in Neck24.354.459.852.8b59.369.871.281.5a107.1(56.7)131.7
ΔMbn [%]-0.71.20.51.5a0.50.80.0b0.60.3------
Mts[Nm] in Stem15.522.919.315.624.811.7b47.0a32.457.416.0(37.0)
ΔMts [%]1.0-1.8b2.91.70.87.2a0.9-0.3-0.4------
Mbs [Nm] in Stem36.857.046.436.8b65.955.4113.1a90.2149.169.6(161.5)
ΔMbs [%]2.7-3.3a5.25.5a0.6-1.00.00.20.2------

a/b = absolute values during activities except cycling and jogging.

Bending of implant neck and torque around stem.

HIGH100 data at the instance of peak force Fres. Mbn = bending moment at base of neck (Fig 1) without friction, only caused by Fres. Mts = torque around stem without friction. Mbs = bending moment in stem without friction. ΔMbn, ΔMts and ΔMbs = changes due to additionally acting friction moments at head. ISO Neck/Stem = ISO 7206-6/4 test forces. a/b = absolute values during activities except cycling and jogging. The ISO neck force produced a bending moment Mbn of 131.7 Nm in the neck; this was higher than any in vivo value, including that from jogging (Table 7). Regarding the local moments in the implant neck and stem (Table 7), the most strenuous in vivo activities were ascending or descending stairs (jogging excluded). The bending moment Mbs in the stem was nearly the same for the ISO stem test (69.6 Nm, Table 7) and for walking (65.9 Nm), although Fres was lower for the ISO stem test (2300N) than for walking (2880N, Table 5). When changing the orientation of the implant relative to the femur by modifying all three angles αx, αy and αz (Fig 1) by ±3°, the largest change of Mbs during walking, caused by any combined angle variations, was ±40%. If the contact forces were considered without friction, the highest bending moment Mbn in the neck was 81.5 Nm when going down a staircase (Table 7, cycling and jogging excluded). If the friction moments were added to the analysis, Mbn rose by a maximum of +1.5%. Without friction moments, the highest torque Mts around the implant stem was 47.0 Nm during stair climbing. Friction increased Mts by 7.2% during stance, but changes for the other activities only reached 2.9%. If only the contact forces acted, Mbs was highest at 113.1 Nm when ascending stairs (jogging excluded). Changes of Mbs due to friction varied between +5.5% and -3.3%.

Data for internet download

The data obtained from this study and additional information can be downloaded as numerical files and graphics from the public database www.Orthoload.com. The AVER75, HIGH100, individual load courses, and video records of the individual subjects during all investigated activities are available. If required, the numerical data can be used directly as input for analytical or experimental investigations.

Discussion

Comparison of data with our previous publication

The data presented were based on a much larger group of ten subjects and included younger and more physically active individuals compared with our previous study [8]. This earlier reference was based on in vivo measured contact forces from only four older and less active subjects. Furthermore, in the previous study only the contact forces but not the moments could be measured. For all activities (cycling and jogging excluded), the peak values of Fres in the AVER75 data were larger in the current than in the previous study. The differences were between 11% and 25% and even 60% for knee bend. This 60% increase was probably caused by the higher hip flexion angles in the new group. The differences in the remaining activities may be explained in part by the faster execution of the exercises. Because the previous study included only four subjects, statistical significances of the differences could not be tested.

Extreme loads, not included in the standardized data

In the previous study [8] extremely ‘high forces’ up to 11000 N during stumbling were reported. Because such uncontrolled movements could not be investigated systematically, even higher occasional contact forces cannot be excluded. Loads of 11000 N are 2.3 times higher than the HIGH100 loads during jogging and 3.8 times above those during walking, as reported here. Because the static strength of metal implant materials is 2 to 3 times higher than the fatigue strength [6], these rare extreme forces could endanger the neck and stem of the implant. Therefore, it is suggested that static tests or tests with low cycle numbers for such extreme forces should be performed in addition to the fatigue tests. The standardized HIGH100 loads describe high load levels in subjects with a BW of 100 kg. It had already been mentioned that these loads should be increased by 25% to encompass the high BW of 60–69 year old male patients in the United States. It is not possible to define loads for extremely obese subjects, during accidents or loads caused by frequent and excessive activities, such as fast running or playing tennis. For such conditions, any upper limitation and standardisation would be arbitrary. Such loading conditions will always cause an increased risk of implant failure that cannot be quantified.

Load transformations for endurance tests of the stem

The presented forces and moments all refer to the coordinate system x, y, z of the femur (Fig 1). For endurance tests of the femoral component, however, these loads are preferably applied in the x’, y’, z’ system of the implant. For such a load transformation, the three orientation angles, αz, αy, and αx, of the implant have to be defined and the HIGH100 force and moment components must be re-calculated according to Eqs 1 to 3. For the implant orientation of our average subject (Table 4), the transformation matrix from Eq 4 can be applied.

Suggestion on approaches to realistically test hip implants

The HIGH100 loads are based on data from only ten subjects. Due to inter-individual variations, the loads would probably increase if data from more subjects were included. This is illustrated by the atypically high AVER75 forces in one subject (H5L) during the double-legged activities (Fig 3). It is likely that even higher loads than those reported here among the general population of joint replacement patients are possible. Additionally, it is common technical practice to perform fatigue tests at a load level that is raised by a safety factor above its expected maximum. This would suggest to test the implants at even higher load levels than reported here. Most endurance tests of the implant stem are currently performed by applying uni-directional sinusoidal forces, as defined in the ISO standards. If this practice is to be continued, the direction and amplitude of the test force should be adapted to realistic HIGH100 forces. One possibility could be to combine the highest activity amplitude of Fres (3875 N in stair descent) with the average force directions αfront and αsagit from the most frequent activity (walking) and the most demanding activity (ascending/descending stairs), which would be 17.5° and 13.5°, respectively (Table 5). If only walking should be simulated, a maximum force of Fres = 2880 N would have to act at αfront = 17° and αsagit = 11°. A more realistic test method would be to apply a combination of HIGH100 peak forces from different activities with sinusoidal time courses. Such a test would either require a 3D loading apparatus or multiple re-arrangements of the stem. Based on the duration and frequency of daily activities in total hip replacement patients [17], the following activity numbers were proposed [8]: 1000 (Walking) + 55 (StairsUp) + 55 (StairsDown) + 25 (StandUp) + 25 (SitDown) + 50 (Stance). For a total cycle number of 107, these numbers must be multiplied by 104. If the testing machine enabled loading of the femoral component under varying directions, the real HIGH100 load-time course during different activities could be applied instead of forces with constant directions. The data from walking (Fig 8, small dots) and the other activities (Table 5) indicate that the individual force directions differ considerably from the average directions of either AVER75 or HIGH100 forces. These individual variations are disregarded when applying unidirectional forces or a combination of time-dependent HIGH100 forces, as proposed for a simplified test. Thus, it may be of interest to test endurance at a defined location at the neck or stem under the worst possible loading conditions. For such a scenario, the absolute maxima of each of the six internal force and moment components at this location can be determined from all individual loads and for a special collection of activities. Such a collection may include the proposed combination of the six most demanding activities or just walking, for example. Because all three force components at the head act in one axis direction and don’t change their signs (Figs 4 and 8), one specific unidirectional force at the head can be calculated, which exactly generates three of these six maximum internal load components. In most applications, this will be the three moment components, as they have the greatest impact on the local stresses. A test force derived in this manner would be best suited to mimic the worst stress condition at the selected location. A ‘worst condition at a given location’ test force could be especially valuable for testing implants of different sizes or the endurance of the taper in modular implants. We plan to support such developments by creating a dedicated, freely available computer program for that purpose in the future.

Analysing the mechanical situation around the implant

The ISO stem test assumes a totally resorbed proximal femur. A different situation exists if the femoral component is proximally well fixed and the torsional stem stability or bone remodelling around the stem have to be tested or analysed. Then the muscle forces acting at the upper femur must also be taken into account as they widely influence bone deformation and stresses and strains in the implant and interface [27, 28]. Knowledge about these forces during various activities, however, requires verified musculoskeletal analyses [29-31]. The data presented here can be used to improve predictive power and accuracy of such simulation models. Friction moments are important for the stability of the cup in the acetabulum, especially for cementless fixations. However, the stresses in the bone surrounding the cup additionally depend on the magnitude and direction of the contact forces relative to the cup. Higher contact forces are likely to increase the torsional strength of the cup fixation due to a clamping effect. Although a detailed analysis of the acetabular cup loading was beyond the scope of the current paper, gait data were captured in all subjects to determine hip joint kinematics and to allow the necessary transformation of contact forces and moments from the femur to the acetabulum. Future publications will focus on the analyses of the hip kinematics and the orientation of the cup loads because this information is needed to test wear and stability of the cup. If quasi-static problems are investigated, e.g., the sudden failure at a porous stem interface during the first postoperative weeks, extreme joint contact forces must be considered. Such excessive loads may act during stumbling, for example. During such events, when substantial levels of muscle co-contraction prevail in an attempt to stabilise the joint, joint contact forces of up 870% BW were measured [7], corresponding to 11,000 N in a 100 kg subject with high load levels [8]. Such extreme forces were observed only by chance and never in planned laboratory experiments or by maximal arbitrary muscle activities. However, because the load directions during stumbling were found to be similar to those during walking, it seems reasonable to approximate such loading conditions by applying the HIGH100 force directions from walking but increasing the force magnitude to 11,000 N. The ISO stem force generated torque in the stem that was smaller than that during most in vivo activities, including walking (Table 7). The bending moment by the ISO stem force was just sufficient to mimic the conditions during walking but was 38% smaller than that observed when ascending stairs. Amplitude as well as direction of the ISO stem force are therefore insufficient for testing the endurance of the stem. The use of the new in vivo load data presented here should permit a better approximation of the mechanical in vivo situation of the stem. When modifying all three implant orientations relative to the femur at the same time by ±3°, the bending moment Mbs changed by up to ±40% during walking. This shows that the stresses in a proximally resorbed implant not only depend on the magnitude of the contact force but also on a) the curvature of the femur in two planes, b) the anteversion of the implant and c) the chosen location on the stem. Fractures of the stem are rare. A proximally resorbed femur is more likely to cause a periprosthetic fracture (0.47% risk for uncemented stems) than a stem failure [32]. This clinical observation and the susceptibility of the moments to the factors a) to c) call into question the relevance of the stem tests. Although the internal moments in neck and stem are influenced by the presence of additional friction moments, their influence is small. The largest increase of any internal moment due to friction was 7.2%. For some activities, friction even slightly reduced the local moments. Compared to the influence of the contact forces, the impact of friction on the internal implant loads is small. Given the complexity of modelling friction in a fatigue test, the effect of friction could be alternatively modelled by increasing the contact force by approximately 7%.

Moments at the acetabular cup

The highest HIGH100 friction moments Mres reached values of up to 2.8 Nm during standing up, descending stairs and knee bending. Our internal data collection contains occasionally measured moments up to 6.1 Nm (not published). Moreover, three months after surgery, the friction moments were between 13% and 72% (47% average in 10 subjects) higher than they were a year after the operation [20] and exhibited substantial individual variability. Therefore, peak in vivo moments can exceed the reported minimum fixation strength limit for the cup, which lay at 2.2 Nm or 3.3 Nm. Thus, the fixation of cementless cups may be at risk, especially soon after the operation, if the components are implanted with only minor under-reaming in subjects with high load levels, a high body weight or higher than average friction moments. This risk is highest when standing up, descending stairs and during knee bending (Table 5), but the mechanics during other activities are only marginally less strenuous. The observation that the friction-induced moment components My and Mz were higher during static stance than during nearly all dynamic activities, including jogging, is of special interest. This probably indicates that friction during stance is high because the synovia is squeezed out of the joint by the permanent static load. This worsens the lubrication and relatively small forces can then generate high moments.
  25 in total

1.  Physiologically based boundary conditions in finite element modelling.

Authors:  Andrew D Speirs; Markus O Heller; Georg N Duda; William R Taylor
Journal:  J Biomech       Date:  2006-12-12       Impact factor: 2.712

Review 2.  Which type of exercise therapy is effective after hip arthroplasty? A systematic review of randomized controlled trials.

Authors:  M Di Monaco; C Castiglioni
Journal:  Eur J Phys Rehabil Med       Date:  2013-10-30       Impact factor: 2.874

3.  Design and calibration of load sensing orthopaedic implants.

Authors:  G Bergmann; F Graichen; A Rohlmann; P Westerhoff; B Heinlein; A Bender; R Ehrig
Journal:  J Biomech Eng       Date:  2008-04       Impact factor: 2.097

4.  Influence of muscle forces on femoral strain distribution.

Authors:  G N Duda; M Heller; J Albinger; O Schulz; E Schneider; L Claes
Journal:  J Biomech       Date:  1998-09       Impact factor: 2.712

5.  Determination of typical patterns from strongly varying signals.

Authors:  A Bender; G Bergmann
Journal:  Comput Methods Biomech Biomed Engin       Date:  2011-07-04       Impact factor: 1.763

6.  Jogging after total hip arthroplasty.

Authors:  Hirohito Abe; Takashi Sakai; Takashi Nishii; Masaki Takao; Nobuo Nakamura; Nobuhiko Sugano
Journal:  Am J Sports Med       Date:  2013-10-10       Impact factor: 6.202

7.  Hip contact forces and gait patterns from routine activities.

Authors:  G Bergmann; G Deuretzbacher; M Heller; F Graichen; A Rohlmann; J Strauss; G N Duda
Journal:  J Biomech       Date:  2001-07       Impact factor: 2.712

8.  Duration and frequency of every day activities in total hip patients.

Authors:  M Morlock; E Schneider; A Bluhm; M Vollmer; G Bergmann; V Müller; M Honl
Journal:  J Biomech       Date:  2001-07       Impact factor: 2.712

9.  Anthropometric reference data for children and adults: United States, 2007-2010.

Authors:  Cheryl D Fryar; Qiuping Gu; Cynthia L Ogden
Journal:  Vital Health Stat 11       Date:  2012-10

10.  Postoperative changes in in vivo measured friction in total hip joint prosthesis during walking.

Authors:  Philipp Damm; Alwina Bender; Georg Bergmann
Journal:  PLoS One       Date:  2015-03-25       Impact factor: 3.240

View more
  38 in total

1.  Improving Musculoskeletal Model Scaling Using an Anatomical Atlas: The Importance of Gender and Anthropometric Similarity to Quantify Joint Reaction Forces.

Authors:  Ziyun Ding; Chui K Tsang; Daniel Nolte; Angela E Kedgley; Anthony M J Bull
Journal:  IEEE Trans Biomed Eng       Date:  2019-03-28       Impact factor: 4.538

Review 2.  [Tribology in hip arthroplasty : Benefits of different materials].

Authors:  J Philippe Kretzer; Maximilian Uhler; Sebastian Jäger; Therese Bormann; Robert Sonntag; Mareike Schonhoff; Stefan Schröder
Journal:  Orthopade       Date:  2021-02-25       Impact factor: 1.087

3.  Effect of a collar on subsidence and local micromotion of cementless femoral stems: in vitro comparative study based on micro-computerised tomography.

Authors:  Valérie Malfroy Camine; Hannes A Rüdiger; Dominique P Pioletti; Alexandre Terrier
Journal:  Int Orthop       Date:  2017-06-07       Impact factor: 3.075

4.  Design of an Affordable, Modular Implant Device for Soft Tissue Tension Assessment and Range of Motion Tracking During Total Hip Arthroplasty.

Authors:  Jonathan C J Wei; Bryan Blaauw; Dieter G M Van Der Pol; Mauricio Cruz Saldivar; Chung-Feng Lai; Jenny Dankelman; Tim Horeman
Journal:  IEEE J Transl Eng Health Med       Date:  2022-05-12

5.  The effects of a functionally-graded scaffold and bone marrow-derived mononuclear cells on steroid-induced femoral head osteonecrosis.

Authors:  Masahiro Maruyama; Akira Nabeshima; Chi-Chun Pan; Anthony W Behn; Timothy Thio; Tzuhua Lin; Jukka Pajarinen; Toshiyuki Kawai; Michiaki Takagi; Stuart B Goodman; Yunzhi Peter Yang
Journal:  Biomaterials       Date:  2018-09-20       Impact factor: 12.479

6.  Analysis of hip joint loading during walking with different shoe types using instrumented total hip prostheses.

Authors:  Y Palmowski; S Popović; D Kosack; P Damm
Journal:  Sci Rep       Date:  2021-05-12       Impact factor: 4.379

7.  Gluteal muscle damage leads to higher in vivo hip joint loads 3 months after total hip arthroplasty.

Authors:  Philipp Damm; Jip Zonneveld; Sophie Brackertz; Florian Streitparth; Tobias Winkler
Journal:  PLoS One       Date:  2018-01-09       Impact factor: 3.240

8.  In Vivo Damage of the Head-Neck Junction in Hard-on-Hard Total Hip Replacements: Effect of Femoral Head Size, Metal Combination, and 12/14 Taper Design.

Authors:  Massimiliano Baleani; Paolo Erani; Barbara Bordini; Federica Zuccheri; Mateusz Kordian Mąkosa; Dalila De Pasquale; Alina Beraudi; Susanna Stea
Journal:  Materials (Basel)       Date:  2017-07-01       Impact factor: 3.623

9.  Bone preserving level of osteotomy in short-stem total hip arthroplasty does not influence stress shielding dimensions - a comparing finite elements analysis.

Authors:  Rene Burchard; Sabrina Braas; Christian Soost; Jan Adriaan Graw; Jan Schmitt
Journal:  BMC Musculoskelet Disord       Date:  2017-08-07       Impact factor: 2.362

10.  The efficacy of core decompression for steroid-associated osteonecrosis of the femoral head in rabbits.

Authors:  Masahiro Maruyama; Tzuhua Lin; Nicolas I Kaminow; Timothy Thio; Hunter W Storaci; Chi-Chun Pan; Zhenyu Yao; Michiaki Takagi; Stuart B Goodman; Yunzhi P Yang
Journal:  J Orthop Res       Date:  2020-11-06       Impact factor: 3.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.