Literature DB >> 27184903

Limitations of Rb-82 weight-adjusted dosing accuracy at low doses.

Jonathan B Moody1, Kerri M Hiller2, Benjamin C Lee1, James R Corbett2, Edward P Ficaro1,2, Venkatesh L Murthy3.   

Abstract

INTRODUCTION: Weight-adjusted dosing is important to maintain accurate quantification for dynamic cardiac three-dimensional positron emission tomography (PET). However, the manufacturer of the only Food and Drug Administration approved rubidium-82 (Rb-82) generator (CardioGen-82, Bracco Diagnostics, Inc.) recommends recalibration after each change in dose which is inefficient in a busy clinical PET lab. The objective of this study was to evaluate the accuracy of Rb-82 dosing without this recalibration.
METHODS: After daily calibration to either 30 mCi (1110 MBq) or 12 mCi (444 MBq), Rb-82 doses between 6 and 45 mCi (222-1665 MBq) were eluted and measured on an external dose calibrator. This was repeated for four generators at weeks 1, 2, or 4 of the generator cycle. The measurements were compared with values reported by the infusion system.
RESULTS: For requested doses less than 30 mCi, the measured dose was consistently lower than the requested dose, ranging from -3.2 mCi at 25 mCi to -5.5 mCi at 10 mCi. The error exceeded 10% for doses less than or equal to 26 mCi for the 30 mCi calibration. Residual activity in the infusion system dead volume accounted for the discrepancy between requested and delivered doses for calibration to 30 mCi but not 12 mCi.
CONCLUSIONS: The CardioGen-82 infusion system is capable of accurate weight-adjusted doses without recalibration for Rb-82 doses as low as 26 mCi when calibrated to 30 mCi. For doses less than 26 mCi, the generator-reported residual activity within the infusion system can be used to correct the delivered dose.

Entities:  

Keywords:  Weight-adjusted dosing; cardiac PET; rubidium-82

Mesh:

Substances:

Year:  2016        PMID: 27184903     DOI: 10.1007/s12350-016-0531-2

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  10 in total

1.  Approaches to enhancing radiation safety in cardiovascular imaging: a scientific statement from the American Heart Association.

Authors:  Reza Fazel; Thomas C Gerber; Stephen Balter; David J Brenner; J Jeffrey Carr; Manuel D Cerqueira; Jersey Chen; Andrew J Einstein; Harlan M Krumholz; Mahadevappa Mahesh; Cynthia H McCollough; James K Min; Richard L Morin; Brahmajee K Nallamothu; Khurram Nasir; Rita F Redberg; Leslee J Shaw
Journal:  Circulation       Date:  2014-09-29       Impact factor: 29.690

2.  Recommendations for reducing radiation exposure in myocardial perfusion imaging.

Authors:  Manuel D Cerqueira; Kevin C Allman; Edward P Ficaro; Christopher L Hansen; Kenneth J Nichols; Randall C Thompson; William A Van Decker; Marko Yakovlevitch
Journal:  J Nucl Cardiol       Date:  2010-08       Impact factor: 5.952

3.  Precision-controlled elution of a 82Sr/82Rb generator for cardiac perfusion imaging with positron emission tomography.

Authors:  R Klein; A Adler; R S Beanlands; R A Dekemp
Journal:  Phys Med Biol       Date:  2007-01-11       Impact factor: 3.609

4.  Use of the 82Sr/82Rb generator in clinical PET studies.

Authors:  G B Saha; R T Go; W J MacIntyre; T H Marwick; A Beachler; J L King; D R Neumann
Journal:  Int J Rad Appl Instrum B       Date:  1990

5.  Ionizing radiation in cardiac imaging: a science advisory from the American Heart Association Committee on Cardiac Imaging of the Council on Clinical Cardiology and Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention.

Authors:  Thomas C Gerber; J Jeffrey Carr; Andrew E Arai; Robert L Dixon; Victor A Ferrari; Antoinette S Gomes; Gary V Heller; Cynthia H McCollough; Michael F McNitt-Gray; Fred A Mettler; Jennifer H Mieres; Richard L Morin; Michael V Yester
Journal:  Circulation       Date:  2009-02-02       Impact factor: 29.690

Review 6.  Approaches to reducing radiation dose from radionuclide myocardial perfusion imaging.

Authors:  Sharmila Dorbala; Ron Blankstein; Hicham Skali; Mi-Ae Park; Jolene Fantony; Charles Mauceri; James Semer; Stephen C Moore; Marcelo F Di Carli
Journal:  J Nucl Med       Date:  2015-03-12       Impact factor: 10.057

7.  Radioanalysis of 82Rb generator eluates.

Authors:  G P Gennaro; B C Bergner; P S Haney; R H Kramer; M D Loberg
Journal:  Int J Rad Appl Instrum A       Date:  1987

8.  2012 ACCF/AHA/ACP/AATS/PCNA/SCAI/STS Guideline for the diagnosis and management of patients with stable ischemic heart disease: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines, and the American College of Physicians, American Association for Thoracic Surgery, Preventive Cardiovascular Nurses Association, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons.

Authors:  Stephan D Fihn; Julius M Gardin; Jonathan Abrams; Kathleen Berra; James C Blankenship; Apostolos P Dallas; Pamela S Douglas; Joanne M Foody; Thomas C Gerber; Alan L Hinderliter; Spencer B King; Paul D Kligfield; Harlan M Krumholz; Raymond Y K Kwong; Michael J Lim; Jane A Linderbaum; Michael J Mack; Mark A Munger; Richard L Prager; Joseph F Sabik; Leslee J Shaw; Joanna D Sikkema; Craig R Smith; Sidney C Smith; John A Spertus; Sankey V Williams
Journal:  J Am Coll Cardiol       Date:  2012-11-19       Impact factor: 24.094

9.  Assessment of a protocol for routine simultaneous myocardial blood flow measurement and standard myocardial perfusion imaging with rubidium-82 on a high count rate positron emission tomography system.

Authors:  Deborah Tout; Christine M Tonge; Sivakumar Muthu; Parthiban Arumugam
Journal:  Nucl Med Commun       Date:  2012-11       Impact factor: 1.690

Review 10.  Will 3-dimensional PET-CT enable the routine quantification of myocardial blood flow?

Authors:  Robert A deKemp; Keiichiro Yoshinaga; Rob S B Beanlands
Journal:  J Nucl Cardiol       Date:  2007 May-Jun       Impact factor: 5.952

  10 in total
  2 in total

1.  Rubidium-82 generator yield and efficiency for PET perfusion imaging: Comparison of two clinical systems.

Authors:  Ali Ahmadi; Ran Klein; Howard C Lewin; Rob S B Beanlands; Robert A deKemp
Journal:  J Nucl Cardiol       Date:  2020-05-20       Impact factor: 5.952

2.  (82)Rb PET imaging of myocardial blood flow-have we achieved the 4 "R"s to support routine use?

Authors:  Robert A deKemp; Ran Klein; Rob S B Beanlands
Journal:  EJNMMI Res       Date:  2016-09-20       Impact factor: 3.138

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.