| Literature DB >> 32436115 |
Ali Ahmadi1, Ran Klein2, Howard C Lewin3, Rob S B Beanlands1, Robert A deKemp4.
Abstract
INTRODUCTION: Strontium-82/Rubidium-82 (82Sr/82Rb) generators are used widely for positron emission tomography (PET) imaging of myocardial perfusion. In this study, the 82Rb isotope yield and production efficiency of two FDA-approved 82Sr/82Rb generators were compared.Entities:
Keywords: Myocardial perfusion imaging; daily quality assurance; positron emission tomography; rubidium-82 generator
Year: 2020 PMID: 32436115 PMCID: PMC7599151 DOI: 10.1007/s12350-020-02200-6
Source DB: PubMed Journal: J Nucl Cardiol ISSN: 1071-3581 Impact factor: 5.952
Figure 1Components of the RUBY-FILL® 82Rb generator and elution system. The CardioGen-82® system is similar but uses a syringe pump instead of peristaltic pump and does not include the generator-bypass valve/tubing for saline-push
Figure 2Example 82Rb generator activity output profiles from a RUBY-FILL® test elution performed at 20 mL·min−1. 82Rb eluate activity (MBq/s, blue) as measured at the output of the patient infusion line over time. Total activity delivered to the patient or calibration vial (MBq, solid orange line) increases over-time according to the integrated area under the 82Rb eluate time–activity curve. Activity measured in a dose-calibrator vial (CardioGen-82®) increases initially over time, but then reaches a steady-state maximum value (MBq, dashed orange line) which is lower than the integrated total activity
Activity, yield and efficiency methods for each elution system
| Generator system | Daily QA reported 82Rb activity | Reported 82Rb yield | Calculated efficiency |
|---|---|---|---|
| RUBY-FILL® | Integrated total from calibration/breakthrough elution | Integrated total from calibration/breakthrough elution | Integrated total 82Rb yield/82Sr parent |
| CardioGen-82® | Dose Cal maximum from daily QA breakthrough elution | Integrated total estimated from daily QA and RUBY-FILL test elutions (1.3 × Max) | Integrated total 82Rb yield/82Sr parent |
Figure 382Rb daily QA activity (solid green lines) measured on RUBY-FILL® (A) and CardioGen-82® (B) generator systems. Activity is shown in mCi (GBq ÷ 37) on the left axis, and as a percent of the 82Sr parent activity available on each day of use (solid red lines) on the right axis
82Rb generator daily QA activity (n = 9 each)
| Generator system | Integral (yield) | Maximum (steady-state) | ||
|---|---|---|---|---|
| GBq (mCi) on first day of use | GBq (mCi) on last day of use | GBq (mCi) on first day of use | GBq (mCi) on last day of use | |
| RUBY-FILL®* | 3.04 ± 0.17 (82.2 ± 4.6) on day 4 | 0.80 ± 0.01 (21.7 ± 0.1) on day 56 | 2.31 ± 0.13 (62.6 ± 3.5) on day 4 | 0.61 ± 0.01 (16.5 ± 0.08) on day 56 |
| CardioGen-82®** | 3.03 ± 0.21 (82.0 ± 5.6) on day 0 | 1.07 ± 0.06 (28.9 ± 1.6) on day 42 | 2.31 ± 0.16 (62.4 ± 4.3) on day 0 | 0.81 ± 0.04 (22.0 ± 1.2) on day 42 |
Values are mean ± standard deviation. mCi values = GBq × 27 mCi/GBq
* RUBY-FILL Maximum values = Integral values ÷ 1.314 correction-factor
**CardioGen-82 Integral values = Maximum values × 1.314 correction-factor
Figure 4RUBY-FILL® test elution results (n = 6 each). At each elution flow-rate, the dose-calibrator maximum activity (A) is a constant fraction (slope) of the integrated total QA activity. The corresponding ratios of Dose-calibrator Maximum/Integrated Total calibration activity (B) are extrapolated to the value of 0.761 that would be expected using a 50 mL @50 mL·min−1 elution. Standard deviation error bars (~ 1%) are smaller than the plotted symbol size
RUBY-FILL® test elution activity ratios (n = 6 each)
| Elution rate | Elution maximum ÷ calibration total | Elution maximum ÷ elution total |
|---|---|---|
| 15 mL·min−1 | 0.536 ± 0.002 | 0.830 ± 0.005 |
| 20 mL·min−1 | 0.592 ± 0.003 | 0.775 ± 0.005 |
| 25 mL·min−1 | 0.626 ± 0.003 | 0.724 ± 0.007 |
| 30 mL·min−1 | 0.649 ± 0.004 | 0.684 ± 0.007 |
| 50 mL·min−1 | 0.761 ± 0.008 * | 0.51 ± 0.07 * |
*extrapolated using mean ± SD of logarithmic and power function models
Figure 582Rb isotope production efficiency of the RUBY-FILL® and CardioGen-82® systems over the clinical shelf-life of N = 9 generators each. 82Rb yield values were measured for RUBY-FILL® (and estimated for CardioGen-82®) using 35 mL elution at 20 mL·min−1