| Literature DB >> 27167121 |
Sandra O'Connell1,2, Gearóid ÓLaighin2,3,4, Lisa Kelly1, Elaine Murphy1, Sorcha Beirne1, Niall Burke1, Orlaith Kilgannon1, Leo R Quinlan1,4.
Abstract
INTRODUCTION: Physical activity is a vitally important part of a healthy lifestyle, and is of major benefit to both physical and mental health. A daily step count of 10,000 steps is recommended globally to achieve an appropriate level of physical activity. Accurate quantification of physical activity during conditions reflecting those needed to achieve the recommended daily step count of 10,000 steps is essential. As such, we aimed to assess four commercial activity monitors for their sensitivity/accuracy in a prescribed walking route that reflects a range of surfaces that would typically be used to achieve the recommended daily step count, in two types of footwear expected to be used throughout the day when aiming to achieve the recommended daily step count, and in a timeframe required to do so.Entities:
Mesh:
Year: 2016 PMID: 27167121 PMCID: PMC4864247 DOI: 10.1371/journal.pone.0154956
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Details of the different activity monitors tested.
| Activity Monitor | Specifications | Outputs | Interface | Location of attachment | Validation Studies |
|---|---|---|---|---|---|
| ActivPAL micro ™ | Step count; time sitting/lying, standing, stepping; number of transitions from standing to sitting; energy expenditure | Need to connect to PC to download and view data | Anterior aspect of the thigh | Ryan et al., 2006 [ | |
| Withings Smart Activity Monitor Tracker (Pulse O2) ™ | Step count; distance; elevation; calories; sleep cycle; heart rate & blood O2 | Screen and smart phone application | On the wrist (detachable wristband); clipped on at the hip (detachable clip); in a pocket | Ferguson et al., 2015 [ | |
| Fitbit One ™ | Step count; distance; elevation; calories; sleep cycle | Screen and smart phone application | Clipped on at hip or chest level; in a pocket; in a wristband (at night) | Ferguson et al., 2015 [ | |
| Garmin Vivofit ™ | Step count; distance; calories; sleep cycle; heart rate; sitting, standing | Screen and smart phone application | -Wristband | ||
| NL-2000 pedometer ™ | Step count; elevation; calories | Screen | -Clipped on at the hip level | Crouter et al., 2003 [ |
Fig 1Position of the activity monitors.
The ActivPAL micro ™ activity monitor on the thigh (right), the NL-2000 ™ pedometer on the waist (right), the Garmin Vivofit ™ on the wrist (left), the Withings Smart Activity Monitor Tracker (Pulse O2) ™ on the waist (left), the Fitbit One ™ activity monitor on the chest.
Fig 2Prescribed walking route.
Breakdown of surfaces and distances throughout the walking route.
Mean step counts observed and registered on the four activity monitors tested over the different surface types of the walking route.
| Observed Step Count | Withings ™ | Garmin ™ | NL-2000 ™ | Fitbit ™ | P Value | |
|---|---|---|---|---|---|---|
| Natural Lawn Grass | 1937 ± 166 | 1945 ± 190 | 1921 ± 156 | 1888 ± 222 | 1928 ± 169 | 0.195 |
| Gravel | 2422 ± 158 | 2408 ± 143 | 2359 ± 150 | 2388 ± 251 | 2424 ± 159 | 0.131 |
| Linoleum | 1979 ± 182 | 1982 ± 193 | 1987 ± 175 | 1922 ± 268 | 1984 ± 191 | 0.106 |
| Tarmacadam/ Asphalt | 2234 ± 234 | 2186 ± 251 | 2185 ± 235 | 2214 ± 239 | 2210 ± 303 | 0.286 |
| Ceramic Tile | 469 ± 31 | 455 ± 50 | 465 ± 21 | 436 ± 108 | 473 ± 41 | 0.457 |
| Ramp | 954 ± 63 | 965 ± 97 | 951 ± 78 | 952 ± 87 | 959 ± 83 | 0.591 |
| Stairs Up | 111 ± 9 | 112 ± 8 | 114 ± 10 | 102 ± 19 | 110 ± 11 | 0.083 |
| Stairs Down | 117 ± 4 | 111 ± 14 | 115 ± 11 | 113 ± 9 | 118 ± 23 | 0.129 |
| Total Step Count | 10950 ± 1209 | 10866 ± 1246 | 10950 ± 1364 | 10695 ± 1281 | 10875 ± 1236 | 0.213 |
There was no statistically significant difference in step count between the observed step count and the step count registered on each activity monitor over the different surfaces tested. Figures represent mean ± SD.
Fig 3Overall walking route step count.
Independent of surface type, there was no statistically significant difference in step count versus the observed step count or versus the ActivPAL ™ activity monitor. All data represent mean ± SEM.
Mean absolute percentage error (MAPE) and error range for the four activity monitors tested irrespective of footwear type.
| Withings ™ | Garmin ™ | NL-2000 ™ | Fitbit ™ | |||||
|---|---|---|---|---|---|---|---|---|
| MAPE | Error Range | MAPE | Error Range | MAPE | Error Range | MAPE | Error Range | |
| Natural Lawn Grass | 1.70 | -6.34 to 15.96 | 3.99 | -16.41 to 14.09 | 3.68 | -25.95 to 7.41 | 1.24 | -11.99 to 2.99 |
| Gravel | 1.81 | -17.89 to 9.01 | 4.55 | -16.29 to 11.67 | 2.83 | -42.59 to 8.96 | 0.69 | -1.68 to 4.70 |
| Linoleum | 1.32 | -6.51 to 5.61 | 3.15 | -9.54 to 7.65 | 4.17 | -49.13 to 5.16 | 1.34 | -7.78 to 7.20 |
| Tarmacadam/ Asphalt | 2.40 | -23.58 to 3.05 | 4.66 | -33.67 to 3.57 | 1.28 | -6.84 to 3.13 | 3.26 | -37.59 to 19.04 |
| Ceramic Tile | 3.20 | -22.20 to 0.22 | -16.89 to 13.23 | -63.43 to 0.00 | 1.06 | -1.04 to 7.02 | ||
| Ramp | 3.52 | -29.69 to 22.53 | 4.10 | -13.98 to 16.13 | 3.08 | -13.76 to 22.00 | 2.18 | -15.76 to 24.55 |
| Stairs Up | -22.94 to 79.41 | -14.29 to 80.88 | - 47.71 to 2.86 | 2.93 | -20.69 to 8.62 | |||
| Stairs Down | -47.46 to 2.61 | -10.53 to 5.58 | 4.43 | -25.62 to 5.45 | -16.36 to 99.16 | |||
| Total Step Count | 1.36 | -9.55 to 3.42 | 4.61 | -21.73 to 9.46 | 2.82 | -18.61 to 3.13 | 1.44 | -8.01 to 3.89 |
Eight mean absolute percentage errors were found to be outside of our selected 5% error zone (in bold): the Withings Smart Activity Monitor Tracker (Pulse O2) ™ step count error on the stairs down and stairs up sections, the Garmin ™ step count error on the ceramic tile and stairs up and stairs down sections, NL-2000 pedometer ™ step count error on the ceramic tile and stairs up sections, and the Fitbit ™ step count error on the stairs down section of the walking route. The error range shows the variation in percentage error for each activity monitor over each surface type.
Mean absolute percentage error (MAPE) and error range for the four activity monitors tested: all values ± 2SD from the mean observed step count excluded.
| Withings ™ | Garmin ™ | NL-2000 ™ | Fitbit ™ | |||||
|---|---|---|---|---|---|---|---|---|
| MAPE | Error Range | MAPE | Error Range | MAPE | Error Range | MAPE | Error Range | |
| Natural Lawn Grass | 1.21 | -6.34 to 7.47 | 2.18 | -14.31 to 5.19 | 2.23 | -14.40 to 7.41 | 1.24 | -11.99 to 2.99 |
| Gravel | 1.25 | -5.88 to 9.01 | 2.75 | -10.06 to 11.67 | 1.46 | -7.17 to 8.96 | 0.69 | -1.68 to 4.70 |
| Linoleum | 1.32 | -6.51 to 5.61 | 3.15 | -9.54 to 7.65 | 2.02 | -8.07 to 5.16 | 1.34 | -7.78 to 7.20 |
| Tarmacadam/ Asphalt | 1.67 | -6.91 to 3.05 | 3.10 | -14.45 to 3.57 | 1.28 | -6.84 to 3.13 | 2.07 | -8.33 to 19.04 |
| Ceramic Tile | 0.82 | -1.46 to 0.22 | 1.36 | -0.64 to 6.41 | 0.69 | -1.60 to 0 | 1.06 | -1.04 to 7.02 |
| Ramp | 1.91 | -4.04 to 10.99 | 2.96 | -13.98 to 7.28 | 2.43 | -13.76 to 6.99 | 0.90 | -4.36 to 4.11 |
| Stairs Up | 3.46 | -13.04 to 14.02 | -14.29 to 16.22 | 2.65 | -8.04 to 2.86 | 2.32 | -9.09 to 8.62 | |
| Stairs Down | 2.20 | -7.29 to 2.61 | -7.27 to 5.88 | 1.91 | -6.25 to 5.45 | 2.10 | -7.14 to 1.65 | |
| Total Step Count | 1.36 | -9.55 to 3.42 | 4.58 | -21.73 to 9.46 | 2.82 | -18.61 to 3.13 | 1.44 | -8.01 to 3.89 |
With all values lying ±2SD from the mean observed step count removed, only two mean absolute percentage errors were found to be outside of our selected 5% error zone (in bold): both for the Garmin ™ device, on the stairs up and stairs down sections. The error range shows the variation in percentage error for each activity monitor over each surface type.
Fig 4Bland Altman plots for step counts of the A. Withings ™, B. Garmin ™, C. Fitbit ™, D. ActivPAL ™, and E. NL-2000 ™ activity monitors.
The solid line represents the mean step count difference between the step count observed from the video recordings and each activity monitor for the overall step count. The dashed lines indicate mean ± limits of agreement (1.96*SD).