| Literature DB >> 35372697 |
Manachai Nonpassopon1, Passara Jongkhajornpong1, Prae Phimpho1, Nontawat Cheewaruangroj1, Kaevalin Lekhanont1, Varintorn Chuckpaiwong1.
Abstract
Objective: To assess agreement and repeatability of white-to-white (WTW) and anterior chamber depth (ACD), and agreement of implantable collamer lens (ICL) size using these measurements from different devices. Methods and analysis: A retrospective review of 83 eyes with ICL implantation (42 patients) was conducted. The agreement of WTW (measured with WaveLight Topolyzer and Orbscan IIz) and ACD (measured with WaveLight Oculyzer and Orbscan IIz) was analysed. Correlation of ICL sizes and difference of eyes with unacceptable vaults between two data sets (WaveLight platform; Topolyzer and Oculyzer and Orbscan IIz) were assessed.Entities:
Keywords: diagnostic tests/investigation; treatment surgery
Mesh:
Year: 2022 PMID: 35372697 PMCID: PMC8905975 DOI: 10.1136/bmjophth-2021-000941
Source DB: PubMed Journal: BMJ Open Ophthalmol ISSN: 2397-3269
Preoperative characteristics of 83 eyes
| Characteristics | Mean (SD) |
| Sphere (diopters) | −8.9 (2.5) |
| Cylinder (diopters), median (range) | −2.0 (-5.8, 0.0) |
| Spherical equivalent (diopters) | −9.9 (2.3) |
| UDVA LogMAR | 2.0 (0.2) |
| BCVA LogMAR | 0.03 (0.09) |
| IOP (mm Hg) | 13.7 (2.4) |
| Endothelial cell count (cells/mm2) | 2838.1 (226.1) |
| Central corneal thickness (mm) | 0.52 (0.03) |
| ICL power (diopters) | −11.8 (2.2) |
| ICL size (mm), median (range) | 13.2 (12.6, 13.2) |
BCVA, best-corrected visual acuity; ICL, implantable collamer lens; IOP, intraocular pressure; UDVA, uncorrected distant visual acuity.
Postoperative results after ICL implantation
| Parameters mean (SD) | 1 month | 3 months | 6 months | 12 months | Last follow-up |
| N=79 | N=57 | N=28 | N=24 | N=83 | |
| UDVA (logMAR) | 0.01 (0.13) | −0.01 (0.12) | −0.05 (0.09) | 0.00 (0.11) | 0.00 (0.12) |
| BCVA (logMAR) | −0.05 (0.08) | −0.06 (0.07) | −0.08 (0.06) | −0.06 (0.05) | −0.06 (0.07) |
| Spherical equivalent (D) | −0.12 (0.27) | −0.11 (0.20) | −0.12 (0.20) | −0.22 (0.24) | −0.16 (0.26) |
| Actual vaults (µm) | 708 (212) | 674 (213) | 613 (196) | 653 (186) | 671 (206) |
BCVA, best-corrected visual acuity; ICL, implantable collamer lens; UDVA, uncorrected distant visual acuity.
Figure 1(A, B): Bland-Altman plots of differences between measurements 1 and 2; (A) White-to-white (WTW) by Topolyzer. (B) Anterior chamber depth (ACD) by Oculyzer, (C, D): Bland-Altman plots of the measurement differences between two devices; (C) WTW between Orbscan IIz and Topolyzer (D) ACD between Orbscan IIz and Oculyzer.
Results of Deming regression analysis to investigate the agreement between two different devices
| Model | Deming regression slope | Intercept (95% CI) | |
| Estimate (95% CI) | SE | ||
| Mean WTW_Orbscan (X), mean WTW_Topolyzer (Y) | 1.15 (1.01 to 1.30) | 1.15 | −1.76 (−3.45 to 0.07) |
| Mean ACD_Orbscan (X), mean ACD_Oculyzer (Y) | 0.99 (0.89 to 1.08) | 0.05 | 0.08 (−0.21 to 0.38) |
ACD, anterior chamber depth; WTW, white-to-white.
Characteristics, ICL information and postoperative vaults of five eyes with unacceptable postoperative vaults
| Case number | Laterality | WTW | WTW | ACD | ACD | ICL power | Toric ICL | ICL size based on dataset 1 | ICL size based on dataset 2 (mm) | Actual postoperative vault* | Presumed postoperative vault† |
| 1 | OD | 11.8 | 11.8 | 3.36 | 3.37 | −10 | 1.5 | 13.20 | 13.20 | 1060 | 1060 |
| 1 | OS | 12 | 11.9 | 3.36 | 3.33 | −13.5 | 2 | 13.20 | 13.20 | 1100 | 1100 |
| 2 | OS | 11.7 | 11.6 | 2.99 | 2.99 | −12.5 | 3 | 12.60 | 12.60 | 180 | 180 |
| 3 | OS | 12.1 | 11.9 | 3.91 | 3.91 | −12.5 | 5 | 13.20 | 13.20 | 1080 | 1080 |
| 3 | OD | 12 | 11.9 | 3.95 | 3.95 | −11.5 | 5 | 13.20 | 13.20 | 1240 | 1240 |
*Calculating ICL size based on dataset 1 (from Orbscan IIz).
†=Calculating ICL size based on dataset 2 (from Topolyzer and Oculyzer).
ACD, anterior chamber depth; ICL, implantable collamer lens; OD, oculus dexter (right eye); OS, oculus sinister (left eye); WTW, white-to-white.
Characteristics, ICL information and postoperative vaults of eight eyes with different ICL size based on two datasets
| Case number | Laterality | WTW | WTW | ACD | ACD | ICL power | Toric ICL | ICL size based on dataset 1 | ICL size based on dataset 2 (mm) | Actual postoperative vault* | Presumed postoperative vault† |
| 1 | OD | 11.5 | 11.7 | 3.12 | 3.17 | −11.0 | 1.0 | 12.60 | 13.20 | 780 | 1330 |
| 1 | OS | 11.4 | 11.7 | 3.08 | 3.10 | −11.5 | 0 | 12.60 | 13.20 | 800 | 1350 |
| 2 | OS | 11.4 | 11.1 | 3.20 | 3.23 | −9.5 | 1.0 | 12.60 | 12.10 | 390 | 0 |
| 3 | OD | 11.6 | 11.7 | 3.12 | 3.16 | −8.0 | 0 | 12.60 | 13.20 | 490 | 1040 |
| 4 | OD | 11.6 | 11.7 | 3.33 | 3.40 | −11.0 | 0 | 12.60 | 13.20 | 610 | 1160 |
| 4 | OS | 11.5 | 11.8 | 3.35 | 3.38 | −14.0 | 0 | 12.60 | 13.20 | 450 | 1000 |
| 5 | OD | 11.5 | 11.7 | 3.39 | 3.46 | −15.0 | 1.5 | 12.60 | 13.20 | 790 | 1340 |
| 6 | OS | 11.8 | 11.6 | 3.00 | 3.05 | −13.0 | 4.5 | 13.20 | 12.60 | 670 | 120 |
*Calculating ICL size based on dataset 1 (from Orbscan IIz),
†=Calculating ICL size based on dataset 2 (from Topolyzer and Oculyzer).
ACD, anterior chamber depth; ICL, implantable collamer lens; OD, oculus dexter (right eye); OS, oculus sinister (left eye); WTW, white-to-white.
Figure 2Diagram showing agreement of ICL sizes of 83 eyes when using measurements from two datasets (dataset 1; WTW and ACD from Orbscan IIz and dataset 2; WTW from Topolyzer and ACD from Oculyzer). ACD, anterior chamber depth; ICL, implantable collamer lens; WTW, white-to-white.