[Effects of central corneal thickness and corneal curvature on measurement of intraocular pressure with Goldmann applanation tonometer and non-contact tonometer].

Yang Zhang1, Jia-liang Zhao, Ai-ling Bian, Xiao-li Liu, Yu-mei Jin.   

Abstract

OBJECTIVE: To investigate the effects of central corneal thickness (CCT) and corneal curvature (CC) on intraocular pressure (IOP) measurements by the Goldmann applanation tonometer (GAT) and the non-contact tonometer (NCT).
METHODS: One hundred and twenty patients were recruited from the clinic of Peking Union Medical College Hospital. The CCT was measured by ultrasound pachymetry and the mean radius of CC by using Canon PK-5 refractometer. The IOP of each eye was measured by both GAT and NCT. Linear regression was used to compare the measurements of GAT and NCT; multi regression was used to analyze the relationships between CCT, CC and the measurements of GAT and NCT. Bland-Altman method was used to compare the effect of NCT and GAT on the IOP measurements.
RESULTS: The results of the right eyes were reported in this paper. The mean and standard deviation of IOP measured by GAT and NCT was (18.4 + or - 4.0) mm Hg (1 mm Hg = 0.133 kPa) and (17.0 + or - 4.6) mm Hg, respectively, the difference was statistically significant (r = 0.835, P = 0.000). IOP measured using GAT increased by 0.039 mm Hg per microm increase in CCT. IOP measured using NCT increased by 0.064 mm Hg per microm increase in CCT. For an increase of 1 mm of mean corneal curvature there was decrease in IOP of 2.648 mm Hg measured by the GAT and of 3.190 mm Hg measured by the NCT. Compared to the GAT, NCT underestimated at low IOP level and overestimated at higher IOP level.
CONCLUSIONS: The IOP measurement obtained with both GAT and NCT varied with CCT and CC. CCT affected IOP measurements by NCT more than that by GAT.

Entities:  

Mesh:

Year:  2009        PMID: 20021884

Source DB:  PubMed          Journal:  Zhonghua Yan Ke Za Zhi        ISSN: 0412-4081


  5 in total

1.  Comparison of intraocular pressures at different points in human's cornea before and after laser in situ keratomileusis with tono-pen tonometer.

Authors:  Xinyu Li; Guigang Li; Lei Liu; Jing Li
Journal:  J Huazhong Univ Sci Technolog Med Sci       Date:  2011-02-19

2.  Effect of manual upper eyelid elevation on intraocular pressure measurement by noncontact tonometer.

Authors:  Raziye Dönmez Gün; Şaban Şimşek
Journal:  Int Ophthalmol       Date:  2022-05-13       Impact factor: 2.029

3.  Ocular biometric parameters are associated with non-contact tonometry measured intraocular pressure in non-pathologic myopic patients.

Authors:  Yazhen Ma; Yanping Ma; Chenli Feng; Minqian Shen; Yuanzhi Yuan
Journal:  Int Ophthalmol       Date:  2019-12-02       Impact factor: 2.031

4.  Comparability of three intraocular pressure measurement: iCare pro rebound, non-contact and Goldmann applanation tonometry in different IOP group.

Authors:  Min Chen; Lina Zhang; Jia Xu; Xinyi Chen; Yuxiang Gu; Yuping Ren; Kaijun Wang
Journal:  BMC Ophthalmol       Date:  2019-11-14       Impact factor: 2.209

5.  Effects of Central Corneal Stromal Thickness and Epithelial Thickness on Intraocular Pressure Using Goldmann Applanation and Non-Contact Tonometers.

Authors:  Marvin Lee; Jaehong Ahn
Journal:  PLoS One       Date:  2016-03-21       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.