Dosimetry of permanent interstitial prostate brachytherapy for an interoperative procedure, using O-arm based CT and TRUS.

Hiromichi Ishiyama1, Akane Sekiguchi1, Takefumi Satoh2, Hideyasu Tsumura2, Kouji Takenaka1, Shogo Kawakami1, Ken-Ichi Tabata2, Kentaro Kobayashi2, Masatsugu Iwamura2, Kazushige Hayakawa1.   

Abstract

PURPOSE: The aim of this report is dosimetric evaluation for an intraoperative fusion computed tomography (CT) as a superior predictor of 1-month CT based dosimetry in comparison to transrectal ultrasound (TRUS) in permanent interstitial prostate brachytherapy.
MATERIAL AND METHODS: Data of 65 patients treated with seed implantation were analyzed. All procedures has been performed with patients in the lithotomy position inside the O-arm system. An end-fine probe is used as a landmark to fuse TRUS and O-arm-based CT images. There was no difference in the patient's position, probe position, and timing of image acquisition between the two imaging modalities. Dose-volume histogram (DVH) parameters such as the dose to 90% of prostate volume (D90) has been analyzed.
RESULTS: The area under the curve of the receiver operating characteristic tended to be larger on fusion CT than on TRUS for most DVH parameters (71.85% vs. 59.59% for D90; p = 0.07). Significant relationships between fusion CT and 1-month CT were confirmed using Pearson's correlation coefficients for most DVH parameters (R = 0.48, p < 0.01 for D90), although the relationship between TRUS and 1-month CT was poor. Large dose reduction (35 Gy for D90) was seen from TRUS to fusion CT, especially in patients with high body weight and small prostate volume.
CONCLUSIONS: Intraoperative fusion CT appears to have higher predictive power for 1-month CT-based dosimetry than TRUS. A prospective trial using fusion CT-based planning is warranted.

Entities:  

Keywords:  O-arm system; brachytherapy; intraoperative CT; low dose rate; prostate cancer

Year:  2016        PMID: 26985192      PMCID: PMC4793069          DOI: 10.5114/jcb.2016.57817

Source DB:  PubMed          Journal:  J Contemp Brachytherapy        ISSN: 2081-2841


Purpose

Transrectal ultrasound (TRUS) is the standard imaging tool for interstitial prostate brachytherapy [1]. The prostate, urethra, and rectum are usually contoured on TRUS images, and treatment is planned based on these contours. In addition, these contours could be modified in a real-time manner during surgery. Meanwhile, it is a well-known fact that ultrasound is not suitable for imaging implanted seeds [2, 3]. Although computed tomography (CT) is not available in the usual operating room, the guideline of the American Brachytherapy Society recommended CT as the gold standard for detecting seed position and calculating post-implant dose volume histograms (DVHs) [4]. The O-arm® surgical imaging system (Medtronic, Dublin, Ireland) was developed to provide real-time, intraoperative CT imaging with a large field-of-view. This system permits patients to be in the lithotomy position even during image acquisition because the bore diameter of this system (965 mm) is significantly larger than that of conventional CT (700-800 mm). However, soft tissues such as the prostate or rectum are difficult to delineate with this system because of its lower contrast resolution compared to conventional CT, although high-density structures such as bone or seeds that are made from titanium can be clearly imaged. Therefore, we combined O-arm-based CT and TRUS during surgery as a new strategy for intraoperative dosimetric evaluation. With this fusion dosimetry, we can compensate for the shortcomings of each modality because accurate seed position can be detected by CT images, and accurate contours can be delineated by ultrasound (US) images. In addition, highly matched fusion images can be expected since there is no difference in the patient's position, probe position, and timing of image acquisition between the two modalities. The purpose of this study was to evaluate whether fusion CT is better than TRUS for intraoperative dosimetry as a predictor of 1-month CT-based dosimetry.

Material and methods

Patients

The institutional review board approved this study (B15-01). Data of 65 patients treated with seed implantation during July 2014 to April 2015 were analyzed. Patients who received additional external beam radiotherapy were excluded. The patients’ characteristics are shown in Table 1.
Table 1

Patients’ characteristics

Age (y)69.3 (6.6)
T stage
 1c33
 2a15
 2b7
 2c10
iPSA (ng/ml)7.52 (3.1)
Gleason score
 3+321
 3+422
 3+52
 4+315
 4+44
 4+51
Hormonal therapy
 Yes15
 No50
Height (cm)165.3 (5.8)
Weight (kg)64.9 (8.0)

Values are given as means (standard deviation) or numbers

iPSA – initial prostate-specific antigen

Patients’ characteristics Values are given as means (standard deviation) or numbers iPSA – initial prostate-specific antigen

Set-up

After anesthesia, patients were placed in the O-arm system installed in our operating room. Zelefsky et al. have already reported the details of this system [5]. All procedures were performed with patients in the lithotomy position inside the O-arm system. This position was kept throughout the whole operation, except when re-positioning was needed or in emergent situations.

Image acquisition and implantation

Transrectal ultrasound images of the whole prostate gland were acquired using a biplane transrectal ultrasound probe (HI VISION Preirus, Hitachi Aloka Medical, Ltd., Tokyo, Japan). Peripheral needles were placed based on these 1st TRUS images, and then 2nd TRUS images were acquired after peripheral needle insertion. The prostate contour was modified based on the 2nd TRUS images because of swelling and deformation of the prostate due to needle insertion. All treatment plans were developed using Variseed software version 8.0.2 (Varian Medical Systems, Palo Alto, CA). The prescribed dose to the prostate with a 3- to 5-mm margin was set at 145 Gy. Two types of 125I source (OncoSeed® model 6711, GE Health care Medi-Physics, Inc, Arlington Hts, IL, or BrachySource® model STM125I, CR BARD, Murray Hill, NJ) were used. Both loose-seed and intraoperatively built custom-linked (IBCL) seeds were used for patients in this study. Loose seeds were placed one by one transperineally through needles attached to a Mick applicator (Eckert & Ziegler BEBIG, Berlin, Germany). Intraoperatively built custom-linked seeds were connected to each other using the Quick-link system (CR BARD, Murray Hill, NJ) and inserted through a relay system [6]. It has been reported that there is no dosimetric difference between loose seeds and IBCL seeds [7]. At the last phase of the operation, whole prostate contours of all slices were revised by the urologist and radiation oncologist, and matched to real-time prostate images through adjusting probe position, and then an end-fire probe was set at the most cranial slice of the prostate. Computed tomography images were then acquired using the O-arm system at 120 kV, 50 mA and 200 mAs.

Image fusion

Acquired CT images were transferred to Variseed software. Figure 1 shows a representative case of image fusion technique. The end-fire probe was used as a landmark to fuse TRUS and O-arm-based CT images because it can be recognized in both modalities. In addition, a Foley catheter that was inserted in the urethra was visualized with both modalities and used for fine tuning. After image fusion, the contours of the prostate, urethra, and rectal wall were copied from TRUS to CT images. Therefore, the same contours were available on both modalities. A phantom study was done before data acquisition and confirmed fusion accuracy less than 1 mm deviation.
Fig. 1

Image fusion of ultrasound (US) and O-arm-based computed tomography (CT). The end-fire probe is shown as a half-moon shaped defect area on the US image (A). Meanwhile, this area is shown as a high-contrast round structure on the CT image (B). Therefore, this end-fire probe was used as a landmark for fusing US and CT images. As the end-fire is covered with 1-mm-thick sheets, a corresponding space is made between the half-moon defect area on US and the high-density area on CT (C). On the sagittal image, the center of the end-probe is matched to the most cranial slice of the prostate contours (D). A Foley catheter is also used for fine tuning of CT (D) and transrectal ultrasound (TRUS) (E) images

Image fusion of ultrasound (US) and O-arm-based computed tomography (CT). The end-fire probe is shown as a half-moon shaped defect area on the US image (A). Meanwhile, this area is shown as a high-contrast round structure on the CT image (B). Therefore, this end-fire probe was used as a landmark for fusing US and CT images. As the end-fire is covered with 1-mm-thick sheets, a corresponding space is made between the half-moon defect area on US and the high-density area on CT (C). On the sagittal image, the center of the end-probe is matched to the most cranial slice of the prostate contours (D). A Foley catheter is also used for fine tuning of CT (D) and transrectal ultrasound (TRUS) (E) images

Dose-volume histogram analysis

Post-implanted CT analysis was done for all patients based on conventional CT images acquired 24 hours and 1 month after implantation. Urethral contouring was based on the outer rim of the urethral catheter, except for 1-month CT, in which the center of the prostate was used as a surrogate for urethral position. The rectal wall including sphincter muscle was fully contoured on 24-h and 1-month CT images, but only the anterior one-third excluding the lumen (body of the TRUS probe) was contoured on TRUS and fusion CT. The urethra and rectum were contoured in the same slices as the prostate contour. Dose-volume histogram parameters including dose to 90% of prostate volume (D90), prostate volume receiving at least 100% dose (V100), prostate volume receiving at least 150% dose (V150), dose to 90% of urethral volume (UD90), dose to 30% of urethral volume (UD30), rectal volume receiving at least 100% dose (RV100), and rectal volume receiving at least 150% dose (RV150) were collected from TRUS, fusion CT, 24-h CT, and 1-month CT.

Statistical analysis

Statistical analyses were performed using R software, version 3.2.0. Dosimetry of 1-month CT was defined as the gold standard [8]. Pearson's or Spearman's correlation coefficients were calculated for TRUS and fusion CT-based dosimetry in relation to the gold standard on 1-month CT. In addition, the diagnostic accuracies of TRUS and fusion CT to predict dosimetry based on 1-month CT were compared using receiver operating characteristic (ROC) analysis. Parameters of D90 > 170 Gy and V100 > 95%, V150 < 65% and RV100 < 1 cc on 1-month CT were defined as the reference for “good quality implant”, although there is no validated definition. The paired t-test was used to compare the DVH parameters of each modality. Risk factors related to the degree of discrepancy of D90 between TRUS and fusion CT were assessed using Pearson's correlation coefficient.

Results

Table 2 shows dosimetric results based on TRUS, fusion CT, 24-h CT, and 1-month CT images. Significant differences between TRUS and fusion CT were detected, although the contours were the same. All DVH parameters except the rectum were significantly decreased on fusion CT compared to TRUS. Surprisingly, the value of D90 dropped by 35 Gy from TRUS to fusion CT. Table 3 shows patient-related risk factors that had a significant effect on the degree of discrepancy between the two. Interestingly, patients with high body weight and small prostate volume tended to have a larger discrepancy between TRUS and fusion CT. As a matter of fact, DVH parameters that indicate a high dose on TRUS and a low dose on fusion CT were also detected as risk factors for a large discrepancy between TRUS and fusion CT (data not shown). Dose-volume histogram parameters of the prostate and urethra increased gradually from fusion CT through 24-h CT to 1-month CT. Meanwhile, DVH parameters of the rectum dropped temporarily on 24-h CT but increased again on 1-month CT (Table 2).
Table 2

Comparison between ultrasound- and computed tomography-based dosimetries

ParameterTRUS p valueFusion CT p value24-h CT P value1-month CT p value
D90 (Gy)191 (14.1)< 0.001156 (14.4)< 0.001163 (13.9)< 0.001183 (17.8)0.004
V100 (%)98.8 (1.50)< 0.00193.0 (4.51)0.00295.3 (5.90)< 0.00197.6 (3.27)0.004
V150 (%)74.6 (7.97)< 0.00149.6 (9.03)n.s.50.5 (10.4)< 0.00168.8 (11.5)< 0.001
UD90 (Gy)174 (17.1)< 0.001134 (17.4)0.002143 (21.7)< 0.001172 (22.7)n.s.
UD30 (%)212 (14.9)< 0.001181 (14.8)n.s.178 (15.7)< 0.001212 (18.9)n.s.
RV100 (cc)0.28 (0.26)< 0.0010.57 (0.55)0.0030.31 (0.42)< 0.0010.64 (0.60)< 0.001
RV150 (cc)0.02 (0.12)< 0.0010.05 (0.08)n.s.0.04 (0.07)n.s.0.06 (0.11)< 0.001

Values are given as means (standard deviation). Each p value between columns corresponds to comparison of the two columns. P value on right edge corresponds to comparison of 1-month CT and TRUS.

D90 – dose to 90% of prostate volume, V100 – prostate volume receiving at least 100% of prescription dose, V150 – prostate volume receiving at least 150% of prescription dose, UD90 – dose to 90% of urethral volume, UD30 – dose to 30% of urethral volume, RV100 – rectal volume receiving at least 100% of prescription dose, RV150 – rectal volume receiving at least 150% of prescription dose

TRUS – transrectal ultrasound, CT – computed tomography, n.s. – not significant

Table 3

Risk factors for a discrepancy in D90 (dose to 90% of prostate volume) between transrectal ultrasound (TRUS) and fusion computed tomography (CT)

ParameterCorrelation coefficient95% CI p value
Weight0.310.070.510.013
Prostate volume*–0.82–0.97–0.180.024

Measurement before implantation

Comparison between ultrasound- and computed tomography-based dosimetries Values are given as means (standard deviation). Each p value between columns corresponds to comparison of the two columns. P value on right edge corresponds to comparison of 1-month CT and TRUS. D90 – dose to 90% of prostate volume, V100 – prostate volume receiving at least 100% of prescription dose, V150 – prostate volume receiving at least 150% of prescription dose, UD90 – dose to 90% of urethral volume, UD30 – dose to 30% of urethral volume, RV100 – rectal volume receiving at least 100% of prescription dose, RV150 – rectal volume receiving at least 150% of prescription dose TRUS – transrectal ultrasound, CT – computed tomography, n.s. – not significant Risk factors for a discrepancy in D90 (dose to 90% of prostate volume) between transrectal ultrasound (TRUS) and fusion computed tomography (CT) Measurement before implantation Table 4 shows the changes in the volumes of each structure. Volumes on TRUS and fusion CT were well matched because the same contouring was used. Meanwhile, rectal volume on 24-h CT was significantly larger than that of TRUS-based volume since contouring methods were different.
Table 4

Comparison of volumes of each structure between modalities

ParameterTRUSFusion CT24-h CT1-month CT
Prostate volume (cc)32.539.9632.559.9633.349.2528.137.38
Urethral volume (cc)0.740.160.730.151.270.280.360.05
Rectal volume (cc)8.492.248.502.2325.9111.3222.857.41

Values are given as means (standard deviation)

TRUS – transrectal ultrasound, CT – computed tomography

Comparison of volumes of each structure between modalities Values are given as means (standard deviation) TRUS – transrectal ultrasound, CT – computed tomography Figure 2 shows a representative case demonstrating discrepancies among TRUS, fusion CT, 24-h CT, and 1-month CT. Although seeds seemed to be implanted inside the prostate on TRUS images, corresponding seeds were implanted outside of the prostate on fusion CT images. Interestingly, on 24-h CT, prostate contouring was affected to a large degree by implanted seeds because of the “seeds-must-be-inside-the-prostate” bias. Prostate volume was slightly decreased on 1-month CT compared to 24-h CT.
Fig. 2

Discrepancies among transrectal ultrasound (TRUS), fusion computed tomography (CT), 24-h CT, and 1-month CT-based dosimetries. Although most seeds seem to be implanted inside the prostate capsule on TRUS images (A), fusion CT shows that at least 4 seeds have been implanted outside the capsule (B). Due to streaking artifact from high-density parts of the probe, seed detection in the posterior-center part of prostate is difficult on fusion CT images (B). Therefore, manual adjustments by investigators are needed in many cases. Interestingly, the prostate contour is highly affected by implanted seeds on the 24-h CT (C). It seems that investigators tend to contour the prostate including all seeds. The prostate contour is slightly reduced on 1-month CT, although seed positions have not moved. Red line – prostate, dark blue line – rectum, light green or green circle – Foley catheter, green triangle – urethra, blue line – 145 Gy, pink line – 160 Gy, yellow line – 217.5 Gy

Discrepancies among transrectal ultrasound (TRUS), fusion computed tomography (CT), 24-h CT, and 1-month CT-based dosimetries. Although most seeds seem to be implanted inside the prostate capsule on TRUS images (A), fusion CT shows that at least 4 seeds have been implanted outside the capsule (B). Due to streaking artifact from high-density parts of the probe, seed detection in the posterior-center part of prostate is difficult on fusion CT images (B). Therefore, manual adjustments by investigators are needed in many cases. Interestingly, the prostate contour is highly affected by implanted seeds on the 24-h CT (C). It seems that investigators tend to contour the prostate including all seeds. The prostate contour is slightly reduced on 1-month CT, although seed positions have not moved. Red line – prostate, dark blue line – rectum, light green or green circle – Foley catheter, green triangle – urethra, blue line – 145 Gy, pink line – 160 Gy, yellow line – 217.5 Gy Figure 3 shows relationships among TRUS, fusion CT, and 1-month CT for prostate D90, V100, and V150. Significant relationships were seen between fusion CT and 1-month CT for all parameters. Only V150 showed a relationship between TRUS and 1 month CT. R values between fusion CT and 1-month CT were higher than between TRUS and 1-month CT.
Fig. 3

Relationships among transrectal ultrasound (TRUS), fusion computed tomography (CT), and 1-month CT for prostate D90, V100, and V150. The left column shows the relationship between TRUS and 1-month CT. The right column shows the relationship between fusion CT and 1-month CT. The regression line is also depicted in each figure. Significant relationships are seen between fusion CT and 1-month CT, although no relationship is seen between TRUS and 1-month CT, except V150. R values between fusion CT and 1-month CT are higher than between TRUS and 1-month CT

Relationships among transrectal ultrasound (TRUS), fusion computed tomography (CT), and 1-month CT for prostate D90, V100, and V150. The left column shows the relationship between TRUS and 1-month CT. The right column shows the relationship between fusion CT and 1-month CT. The regression line is also depicted in each figure. Significant relationships are seen between fusion CT and 1-month CT, although no relationship is seen between TRUS and 1-month CT, except V150. R values between fusion CT and 1-month CT are higher than between TRUS and 1-month CT Figure 4 shows relationships among TRUS, fusion CT, and 1-month CT for UD90 and UD30. There were significant relationships among TRUS, fusion CT, and 1-month CT for UD30. R values between fusion CT and 1-month CT were higher than between TRUS and 1-month CT. There were no significant relationships among any modalities for RV100 and RV150 (Figure 5).
Fig. 4

Relationships among transrectal ultrasound (TRUS), fusion computed tomography (CT), and 1-month CT for UD90 and UD30. Significant relationships are seen among the three modalities for UD30. R values are higher between fusion CT and 1-month CT than between TRUS and 1-month CT

Fig. 5

Relationships among transrectal ultrasound (TRUS), fusion computed tomography (CT), and 1-month CT for RV100 and RV150. There are no significant relationships among any modalities for rectal V100 and V150, although the relationship between fusion CT and 1-month CT for RV100 has a relatively small p value

Relationships among transrectal ultrasound (TRUS), fusion computed tomography (CT), and 1-month CT for UD90 and UD30. Significant relationships are seen among the three modalities for UD30. R values are higher between fusion CT and 1-month CT than between TRUS and 1-month CT Relationships among transrectal ultrasound (TRUS), fusion computed tomography (CT), and 1-month CT for RV100 and RV150. There are no significant relationships among any modalities for rectal V100 and V150, although the relationship between fusion CT and 1-month CT for RV100 has a relatively small p value Figure 6 shows the areas under the ROC curves of TRUS and fusion CT for the defined thresholds on 1-month CT. Area under the curve (AUC) values tended to be higher on fusion CT than on TRUS, except for V100 but not significantly. Table 5 shows the threshold for each DVH parameter derived from ROC, and their diagnostic performance with the best accuracy for the defined good quality implantation.
Fig. 6

Receiver operating characteristic curves for D90, pV100, pV150, and RV100. Area under the curve (AUC) values tend to be higher on fusion computed tomography (CT) than on transrectal ultrasound (TRUS), except for V100 but not significantly. Bold line – fusion CT, dotted line – TRUS

Table 5

Diagnostic performance with the best accuracy for good quality implant on 1-month computed tomography

D90 > 170 (Gy)V100 > 95 (%)V150 < 65 (%)RV100 < 1 (cc)
Fusion CTTRUSFusion CTTRUSFusion CTTRUSFusion CTTRUS
Threshold154.7183.289.999.144.578.90.260.42
Sensitivity (%)67.983.089.765.555.095.040.774.1
PPV (%)92.386.394.597.468.841.395.788.9
Specificity (%)75.041.757.185.788.940.090.954.5
NPV (%)34.635.740.023.181.694.723.830.0

CT – computed tomography, TRUS – transrectal ultrasound, PPV – positive predictive value, NPV – negative predictive value

Receiver operating characteristic curves for D90, pV100, pV150, and RV100. Area under the curve (AUC) values tend to be higher on fusion computed tomography (CT) than on transrectal ultrasound (TRUS), except for V100 but not significantly. Bold line – fusion CT, dotted line – TRUS Diagnostic performance with the best accuracy for good quality implant on 1-month computed tomography CT – computed tomography, TRUS – transrectal ultrasound, PPV – positive predictive value, NPV – negative predictive value

Discussion

Several papers have reported the usefulness of intraoperative CT-based dosimetry for interstitial prostate brachytherapy [5, 9, 10]. However, they did not match the patient's body position, inserted probe position, timing of image acquisition, and contours between TRUS and intraoperative CT as in the present study. Steggerda et al. have already reported the same fusion technique with satisfactory accuracy (average deviation less than 1 mm) [11]. Their paper, however, focused on the geometrical accuracy of fusion imaging of CT and TRUS and showed little information about DVH parameters. The inaccuracy of TRUS-based dosimetry has been reported in several papers. Nag et al. compared intraoperative US dosimetry with postoperative CT dosimetry acquired a few hours after operation [12]. As in the present study, they reported 16% and 10% differences in D90 and V100, respectively, and significant underestimation of rectal dose. Igidbashian et al. also reported the poor predictive value of intraoperative TRUS dosimetry for 1-month CT-based dosimetry [13]. When comparing the dosimetries between TRUS and fusion CT, we can consider that the discrepancy was caused by “implant errors” because the contours of both modalities were the same. As shown in Figure 2, seed positions on TRUS were different from those on fusion CT. Especially for seeds in peripheral needles, they tended to unintentionally deposit in outer positions. This may be caused by obliquely inserted needles. Although the needle path was a completely horizontal line in the planning software (Variseed), most of the actual needles were obliquely inserted to varying degrees. Therefore, seed position would stray from the ideal path on the software to an unintentional position, especially for deeply implanted seeds. Patients with high body weight tended to have thick peritoneal soft tissue that would lengthen the distance between the template and the prostate. In addition, dose parameters such as D90 would be highly affected by implant errors when calculated from a small target volume. Therefore, the doses to the prostate and urethra were significantly reduced from TRUS to fusion CT, especially in patients with high body weight and small prostate volume. Doses to the rectum would be inversely increased because the seeds were implanted near the rectal wall in these cases. When comparing dosimetries between fusion CT and 24-h CT, we can consider that the discrepancy was caused by “deformation” due to probe insertion and “contouring errors” of investigators, since seed positions and prostate volumes were probably not changing for 24 hours. As shown in Figure 2, prostate contours could be erroneously delineated by investigators. It has been reported that prostates are significantly larger on CT images than on MRI or US images [14, 15], especially the seminal vesicles and apex. In addition, significant interobserver variation was also reported in prostate delineation on CT images [16, 17]. Meanwhile, US images have a high degree of accuracy and a strong correlation with pathologic prostate volume [18]. In the present study, the value of D90 and V100 increased slightly from fusion CT to 24-h CT. This was probably caused by erroneous delineation biased by seed position, as shown in Figure 2. The area under the curve of TRUS was higher than that of fusion CT only for V100 in the present study (Figure 6).This was probably caused by erroneous delineation on 1-month CT used as a standard. If the standard (1-month CT) was overestimated, the overestimating predictor (TRUS) would have higher prediction power compared to a more precise predictor (fusion CT). Prostate deformation caused by probe insertion may have some effects on DVH, although its impact would be small, except for the rectal wall. The value of RV100 decreased slightly from fusion CT to 24-h CT (Table 2). Though the anterior rectal wall on fusion CT was anteriorly deviated due to probe insertion, the deviation was resolved by removing the probe for 24-h CT. In addition, rectal contouring methods were different between the two modalities, as described above. These differences probably caused the DVH differences of RV100. The current standard for postoperative dosimetric analysis was based on CT images after implantation, although recommended timing was not defined. However, it has been suggested that more than 30 days after implant would be needed to precisely evaluate the dose distribution, considering the decaying dose rate of 125I and prostate volume change [8]. Therefore, we defined the 1-month CT as the gold standard. When comparing dosimetry between 24-h CT and 1-month CT, we can consider that the discrepancy was caused by “volume change”. Dose to every structure was significantly increased from 24-h CT to 1-month CT because of significant prostate volume reduction. Since the distances between the rectum and implanted seeds were reduced, dose to the rectal wall also increased on 1-month CT. The present study showed that fusion CT had a higher predictive power for 1-month CT dosimetry than TRUS. Nevertheless, the predictive power of fusion CT was not satisfactory, as shown in Table 5. Sensitivity and specificity for predicting D90 > 170 Gy were only 67.9% and 75.0%, respectively. Because 1-month CT images included not only implant errors but also contouring errors and volume changes, perfect prediction was probably impossible. At this time, we believe that reducing implant errors using fusion CT images is the best way to improve total implant quality. Especially for patients with high body weight and a small prostate, who tended to have large implant errors, as the present study suggested, fusion CT-based planning would be useful for high quality implantation. The present study has several limitations. Prostate movement during TRUS image acquisition may cause some problems. Because mechanical shifting of the end-fire probe from the base to the apex in a number of steps usually causes small prostate movement, reconstructed 3D images include not only prostate anatomy but also prostate movement. Therefore, contours on TRUS images may not perfectly match the actual prostate on CT images, because the probe was not moving during O-arm CT acquisition. Another problem was streaking artifacts caused by seeds and radio-opaque parts inside the TRUS probe. Since the automatic seed-finding program of the planning software could not handle these strong artifacts perfectly, manual adjustment of seed detection was needed for almost all patients, although it took only 10 minutes. The bore size of the O-arm system may be problematic for American or European people who are relatively taller than Japanese people. Although most of our patients could be in the lithotomy position, some patients who were relatively taller did not undergo CT due to positioning difficulty.

Conclusions

The present study showed that intraoperative fusion CT has higher predictive power for 1-month CT-based dosimetry than TRUS. In addition, the fusion CT-based dosimetry permits evaluation of “implant errors” without disturbance from contouring errors and volume changes. The present results also suggest that patients with high body weight and small prostate volume tended to have large implant error. To improve the quality of permanent interstitial prostate brachytherapy, especially for these high-risk patients, a prospective trial of fusion CT-based planning is warranted.
  18 in total

Review 1.  The American Brachytherapy Society recommendations for permanent prostate brachytherapy postimplant dosimetric analysis.

Authors:  S Nag; W Bice; K DeWyngaert; B Prestidge; R Stock; Y Yu
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-01-01       Impact factor: 7.038

2.  Comparison of intraoperatively built custom linked seeds versus loose seed gun applicator technique using real-time intraoperative planning for permanent prostate brachytherapy.

Authors:  A Jason Zauls; Michael S Ashenafi; Georgiana Onicescu; Harry S Clarke; David T Marshall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-10-01       Impact factor: 7.038

Review 3.  Review of intraoperative imaging and planning techniques in permanent seed prostate brachytherapy.

Authors:  Alfredo Polo; Carl Salembier; Jack Venselaar; Peter Hoskin
Journal:  Radiother Oncol       Date:  2010-01-13       Impact factor: 6.280

4.  American Brachytherapy Society consensus guidelines for transrectal ultrasound-guided permanent prostate brachytherapy.

Authors:  Brian J Davis; Eric M Horwitz; W Robert Lee; Juanita M Crook; Richard G Stock; Gregory S Merrick; Wayne M Butler; Peter D Grimm; Nelson N Stone; Louis Potters; Anthony L Zietman; Michael J Zelefsky
Journal:  Brachytherapy       Date:  2012 Jan-Feb       Impact factor: 2.362

5.  Real-time intraoperative computed tomography assessment of quality of permanent interstitial seed implantation for prostate cancer.

Authors:  Michael J Zelefsky; Mick Worman; Gilad N Cohen; Xin Pei; Marisa Kollmeier; Josh Yamada; Brett Cox; Zhigang Zhang; Eva Bieniek; Lawrence Dauer; Marco Zaider
Journal:  Urology       Date:  2010-11       Impact factor: 2.649

6.  Real-time computed tomography dosimetry during ultrasound-guided brachytherapy for prostate cancer.

Authors:  Irving D Kaplan; Paul Meskell; Nicklas E Oldenburg; Brian Saltzman; Gary P Kearney; Edward J Holupka
Journal:  Brachytherapy       Date:  2006 Jul-Sep       Impact factor: 2.362

7.  Intraoperative adaptive brachytherapy of iodine-125 prostate implants guided by C-arm cone-beam computed tomography-based dosimetry.

Authors:  Hendrik Westendorp; Carel J Hoekstra; Arie van't Riet; André W Minken; Jos J Immerzeel
Journal:  Brachytherapy       Date:  2007 Oct-Dec       Impact factor: 2.362

8.  Definition of the prostate in CT and MRI: a multi-observer study.

Authors:  C Rasch; I Barillot; P Remeijer; A Touw; M van Herk; J V Lebesque
Journal:  Int J Radiat Oncol Biol Phys       Date:  1999-01-01       Impact factor: 7.038

9.  A comparison of CT scan to transrectal ultrasound-measured prostate volume in untreated prostate cancer.

Authors:  S Christopher Hoffelt; Lynn M Marshall; Mark Garzotto; Arthur Hung; John Holland; Tomasz M Beer
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-09-01       Impact factor: 7.038

10.  Multisector dosimetry in the immediate post-implant period: significant under dosage of the prostate base.

Authors:  Austin N Kirschner; Vythialingam Sathiaseelan; Yunkai Zhang; James David; John A Kalapurakal
Journal:  J Contemp Brachytherapy       Date:  2014-04-03
View more
  8 in total

1.  Intraoperative Registered Ultrasound and Fluoroscopy (iRUF) for dose calculation during prostate brachytherapy: Improved accuracy compared to standard ultrasound-based dosimetry.

Authors:  Junghoon Lee; Omar Y Mian; Yi Le; Hee Joon Bae; E Clif Burdette; Theodore L DeWeese; Jerry L Prince; Daniel Y Song
Journal:  Radiother Oncol       Date:  2017-06-21       Impact factor: 6.280

2.  Comparison of post-implant dosimetrics between intraoperatively built custom-linked seeds and loose seeds by sector analysis at 24 hours and 1 month for localized prostate cancer.

Authors:  Eiichiro Okazaki; Katsuyuki Kuratsukuri; Tomohiro Nishikawa; Tomoaki Tanaka; Mai Sakagami; Keiko Shibuya
Journal:  J Contemp Brachytherapy       Date:  2020-08-21

3.  Evaluation of transrectal ultrasound-based dosimetry for brachytherapy of prostate cancer: a single-center experience.

Authors:  Changzhao Yang; Zhengtong Lv; Lingxiao Chen; Jie Wang; Xiheng Hu; Harripersaud Chand; Xi Sun; Guyu Tang; Congyi Tang; Huichuan Jiang; Yuan Li
Journal:  J Contemp Brachytherapy       Date:  2020-08-21

4.  Superior Postimplant Dosimetry Achieved Using Dynamic Intraoperative Dosimetry for Permanent Prostate Brachytherapy.

Authors:  Tanmay Singh; Junghoon Lee; Marianna Zahurak; Hee Joon Bae; Tamey Habtu; Robert Hobbs; Yi Le; Everette C Burdette; Daniel Y Song
Journal:  Pract Radiat Oncol       Date:  2021-03-13

5.  Is intraoperative real-time dosimetry in prostate seed brachytherapy predictive of biochemical outcome?

Authors:  Daniel Taussky; Levon Igidbashian; David Donath; Dominic Béliveau-Nadeauv; Renée X Larouche; Yanick Hervieux; Guila Delouya
Journal:  J Contemp Brachytherapy       Date:  2017-06-22

6.  Automatic needle detection using improved random sample consensus in CT image-guided lung interstitial brachytherapy.

Authors:  Yongnan Zheng; Shan Jiang; Zhiyong Yang; Lin Wei
Journal:  J Appl Clin Med Phys       Date:  2021-03-25       Impact factor: 2.102

7.  A cold spot compensation technique using a combination of trans-rectal ultrasonography and intraoperative computed tomography for interstitial permanent prostate brachytherapy: a single-arm prospective trial.

Authors:  Hiromichi Ishiyama; Hideyasu Tsumura; Shogo Kawakami; Takefumi Satoh; Akane Sekiguchi; Ken-Ichi Tabata; Masatsugu Iwamura; Kazushige Hayakawa
Journal:  J Contemp Brachytherapy       Date:  2018-02-28

8.  Correlations of post-implant regional dosimetric parameters at 24 hours and one month, with clinical results of low-dose-rate brachytherapy for localized prostate cancer.

Authors:  Eiichiro Okazaki; Katsuyuki Kuratsukuri; Kentaro Ishii; Tomoaki Tanaka; Ryo Ogino; Tomohiro Nishikawa; Hideyuki Morimoto; Masako Hosono; Yukio Miki
Journal:  J Contemp Brachytherapy       Date:  2017-12-30
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.