| Literature DB >> 24790620 |
Austin N Kirschner1, Vythialingam Sathiaseelan1, Yunkai Zhang1, James David1, John A Kalapurakal1.
Abstract
PURPOSE: While there are several reports of prostate multisector dosimetry data obtained from CT or MRI scans performed at intervals ranging from 14-70 days after prostate brachytherapy (PB), there are no reports on multisector dosimetry performed in the immediate post-implant period. This study was undertaken to determine the results of prostate multisector dosimetry performed in the immediate post-implant period on day 1 post-implant dosimetry after (125)I PB.Entities:
Keywords: CT-based dosimetry; brachytherapy; multisector; prostate cancer
Year: 2014 PMID: 24790620 PMCID: PMC4003430 DOI: 10.5114/jcb.2014.42023
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1Pre-implant ultrasound dosimetry axial scans spaced 5 mm apart. The prostate gland is outlined in cyan, 144 Gy (100%) isodose line in yellow and 216 Gy (150%) isodose line in red. The multisector dosimetric parameters are: Base (top row): D90 = = 168.3 Gy, V100 = 99.8%; Mid gland (middle row): D90 = 177.1 Gy, V100 = 100.0%; Apex (bottom row): D90 = 162.2 Gy, V100 = 98.9%
Fig. 2Day 1 post-implant CT scan dosimetry of the patient in Figure 1 after I125 brachytherapy. The 3 mm thick CT axial slices are spaced 3 mm apart. The prostate gland is outlined in yellow and the rectum in blue, 144 Gy (100%) isodose line in green, and 216 Gy (150%) isodose line in red. The multisector dosimetric parameters are: base (top row): D90 = 95.9 Gy, V100 = 71.0%; mid gland (middle row): D90 = 147.6 Gy, V100 = 90.7%; apex (bottom row): D90 = 168.5 Gy, V100 = 95.7%
Comparison of V100 and D90 of the prostate base relative to the entire prostate gland, mid-gland, and apex on pre-implant ultrasound multisector dosimetry
| Dosimetry parameter | Base vs. Entire gland | Base vs. Mid-gland | Base vs. Apex | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Base | Entire gland |
| Base | Mid-gland |
| Base | Apex |
| ||
| V100 | mean | 98.9% | 99.2% | 0.20 | 98.9% | 99.5% | < 0.01 | 98.9% | 98.5% | 0.36 |
| SD | 1% | 1% | 1% | 1% | 1% | 4% | ||||
| D90 | mean | 166.1 | 174.1 | < 0.01 | 166.1 | 178.7 | < 0.01 | 166.1 | 170.3 | 0.02 |
| SD | 6.63 | 8.14 | 6.63 | 8.25 | 6.63 | 1.42 | ||||
V100 – Percentage volume of the prostate gland that receives 100% of prescription dose, D90 – dose (Gy) that treats 90% of the prostate gland volume, SD – standard deviation
Comparison of V100 and D90 of the prostate base relative to the entire prostate gland, mid-gland, and apex on day 1 post-implant CT scan multisector dosimetry
| Dosimetry parameter | Base vs. Entire gland | Base vs. Mid-Gland | Base vs. Apex | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Base | Entire gland |
| Base | Mid-gland |
| Base | Apex |
| ||
| V100 | mean | 77.9% | 90.2% | < 0.01 | 77.9% | 95.1% | < 0.01 | 77.9% | 90.6% | < 0.01 |
| SD | 14% | 6% | 14% | 5% | 14% | 12% | ||||
| D90 | mean | 125.7 | 148.7 | < 0.01 | 125.7 | 170.3 | < 0.01 | 125.7 | 157.4 | < 0.01 |
| SD | 23.3 | 22.1 | 23.3 | 26.0 | 23.3 | 29.5 | ||||
V100 – Percentage volume of the prostate gland that receives 100% of prescription dose, D90 – dose (Gy) that treats 90% of the prostate gland volume, SD – standard deviation