Is intraoperative real-time dosimetry in prostate seed brachytherapy predictive of biochemical outcome?

Daniel Taussky1,2, Levon Igidbashian3, David Donath1, Dominic Béliveau-Nadeauv1, Renée X Larouche1, Yanick Hervieux1, Guila Delouya1,2.   

Abstract

PURPOSE: To analyze intraoperative (IO) dosimetry using transrectal ultrasound (TRUS), performed before and after prostate low-dose-rate brachytherapy (LDR-BT), and compare it to dosimetry performed 30 days following the LDR-BT implant (Day 30). <br> MATERIAL AND METHODS: A total of 236 patients underwent prostate LDR-BT using 125I that was performed with a three-dimensional TRUS-guided interactive inverse preplanning system (preimplant dosimetry). After the implant procedure, the TRUS was repeated in the operating room, and the dosimetry was recalculated (postimplant dosimetry) and compared to dosimetry on Day 30 computed tomography (CT) scans. Area under curve (AUC) statistics was used for models predictive of dosimetric parameters at Day 30. <br> RESULTS: The median follow-up for patients without BF was 96 months, the 5-year and 8-year biochemical recurrence (BR)-free rate was 96% and 90%, respectively. The postimplant median D90 was 3.8 Gy lower (interquartile range [IQR], 12.4-0.9), and the V100 only 1% less (IQR, 2.9-0.2%) than the preimplant dosimetry. When comparing the postimplant and the Day 30 dosimetries, the postimplant median D90 was 9.6 Gy higher (IQR [-] 9.5-30.3 Gy), and the V100 was 3.2% greater (0.2-8.9%) than Day 30 postimplant dosimetry. The variables that best predicted the D90 of Day 30 was the postimplant D90 (AUC = 0.62, p = 0.038). None of the analyzed values for IO or Day 30 dosimetry showed any predictive value for BR. <br> CONCLUSIONS: Although improving the IO preimplant and postimplant dosimetry improved dosimetry on Day 30, the BR-free rate was not dependent on any dosimetric parameter. Unpredictable factors such as intraprostatic seed migration and IO factors, prevented the accurate prediction of Day 30 dosimetry.

Entities:  

Keywords:  biochemical recurrence; brachytherapy; intraoperative dosimetry; prostate cancer

Year:  2017        PMID: 28951748      PMCID: PMC5611451          DOI: 10.5114/jcb.2017.68467

Source DB:  PubMed          Journal:  J Contemp Brachytherapy        ISSN: 2081-2841


Purpose

In 2001, the American Brachytherapy Society (ABS) provided their critical assessment on intraoperative (IO) planning in prostate permanent seed low-dose-rate brachytherapy (LDR-BT) with 125I, and gave recommendation for an online real-time IO dosimetry [1]. In 2010, the Groupe Européen de Curiethérapie and the European Society for Radiotherapy & Oncology (GEC-ESTRO) group discussed the various ways to use real-time dosimetry and described the basic approaches involving IO planning within the operating room (OR). This consisted of interactive planning with modification of the treatment based on the feedback of the real-time tracking of the actual needle positions and a continuous feedback of the seed positions as they were implanted [2]. In 2008, we described our results with an interactive planning technique using real-time three-dimensional (3D) transrectal ultrasound (TRUS) for dose planning and IO navigation [3], and reported the clinical outcome, toxicity [4,5], and seed loss [6] for low to high activity seeds. We described that IO dosimetric parameters on TRUS could predict a good quality implant on the Day 30 computed tomography (CT) scan. We also discovered that a good IO preimplant plan was the best predictive factor of a good implant on Day 30 [3]. In this present study, we updated our experience with a larger than previously described cohort and investigated whether IO TRUS-based dosimetry was predictive of biochemical recurrence (BR).

Material and methods

Of the first 251 patients who were treated with LDR-BT at our department, complete data were available for 238 (95%) patients. A postoperative dosimetry was not obtained due to technical difficulties in the other 5% of patients. In all patients, TRUS (BK Medical Systems, Harlev, Denmark) was performed immediately after the implant of two fixation needles. This TRUS was used for IO pre-planning (preimplant IO dosimetry). The planning and seed delivery were performed using the Nucletron FIRST system (Nucletron, an Elekta company, Elekta AB, Stockholm, Sweden) with IO interactive planning, dynamic dose calculation, virtual needle guidance, robotic seed delivery, and needle retraction system. Immediately after implantation, an additional TRUS study was performed with the fixation needles still in place. Prostate delineation was repeated within the planning system a few days after the implant on the postoperative TRUS images, and the dosimetry to the prostate was recalculated (postimplant IO dosimetry). Slice thickness on both TRUS studies was 2.5 mm. All patients had CT-based dosimetry performed at 30 days (Day 30) after the implant. The implant technique and dosimetry were performed as previously described [3]. Slice thickness was 3 mm.

Statistical analysis

Correlations between variables were calculated using the Pearson correlation coefficient. The value of IO dosimetry to predict Day 30 dosimetry was calculated using area under the receiver operating characteristic (AUROC) curves. BR was defined according to the Phoenix definition (PSA = nadir + 2 ng/ml). Survival analyses were performed using the Kaplan-Meier method and were compared using the log-rank test.

Results

Table 1 lists patient and dosimetric characteristics of the 236 patients analyzed. Compared to the preimplant IO dosimetry, the postimplant D90 median was 3.8 Gy lower (interquartile range [IQR], 12.4-0.9), and the V100 was 1% less (IQR, 2.9-0.2%). The postimplant IO D90 was a median of 9.6 Gy higher (IQR, 9.5-30.3 Gy) than on Day 30, and the postimplant V100 was 3.2% higher (0.2-8.9%) than on Day 30. See Table 2 for differences in dosimetry at the different time-points. The quality of the implant measured with the D90 and V100 did not change significantly over time. Only the postimplant V100 value increased slightly with the year of implantation (r = 0.19, p = 0.004), but not the D90 without any effect on preimplant or Day 30 values (data not shown).
Table 1

Patient and dosimetric characteristics (n = 236)

FactorMedian (IQR)
Age median65 (61-69)
Seed activity (mCi)0.59 (0.44-0.61)
Seeds/needles2.6 (2.4-2.8)
ADT4%
PSA5.7 (4.4-7.3)
Gleason score 6/7/887.5%/12.2%/0.4%
Prostate volume (cc)38.6 (32-45)
Preimplant V100(%)99 (98-100)
Preimplant D90 (Gy)178 (173-182)
Postimplant V100 (%)98 (95-99)
Postimplant D90 (Gy)171 (163-177)
Day 30 V100 (%)93 (89-97)
Day 30 D90 (Gy)157 (140-177)

IQR – interquartile range, ADT – androgen deprivation therapy, PSA – prostate-specific antigen, V100% – volume of the anatomic volume receiving 100% of the prescribed dose, D90 – percent of the prescription dose covering 90% of the CTV

Table 2

Differences in dosimetry

FactorMedianIQR
D90 (Gy) Δ postimplant – preimplant–3.8–12.4 to –0.9
D90 Δ postimplant – Day 309.6–9.5 to 30.3
D90 Δ preimplant – Day 3020.1–0.8 to 36.8
V100 (%) Δ postimplant – preimplant–0.95–2.9 to –0.2
V100 Δ postimplant – Day 303.20.2 to 8.9
V100 Δ preimplant – Day 305.41.9 to 9.6

IQR – interquartile range, Δ – difference, V100% – volume of the anatomic volume receiving 100% of the prescribed dose, D90 – percent of the prescription dose covering 90% of the CTV

Patient and dosimetric characteristics (n = 236) IQR – interquartile range, ADT – androgen deprivation therapy, PSAprostate-specific antigen, V100% – volume of the anatomic volume receiving 100% of the prescribed dose, D90 – percent of the prescription dose covering 90% of the CTV Differences in dosimetry IQR – interquartile range, Δ – difference, V100% – volume of the anatomic volume receiving 100% of the prescribed dose, D90 – percent of the prescription dose covering 90% of the CTV Figure 1 illustrates the correlation between the preimplant and the postimplant D90, and Figure 2 shows the correlation between the postimplant D90 and the D90 on Day 30 as a scatterplot. Both the postimplant IO D90 (AUC = 0.62, p = 0.038) and the preimplant IO D90 (AUC = 0.61, p = 0.046) were significant predictors of the D90 at Day 30. However, the false positive rate to predict for a D90 ≥ 130 Gy for different cut-offs was too high to identify ideal cut-offs. The D90 on Day 30 weakly correlated with the postimplant D90 (r = 0.13, p = 0.05) and with the prostate volume (r = 0.26, p < 0.001). On Day 30, the D90 was < 130 Gy in 13% of the patients.
Fig. 1

Correlation between the preimplant D90 and the postimplant D90 as a scatterplot

Fig. 2

Correlation between the postimplant D90 and the D90 on Day 30 as a scatterplot

Correlation between the preimplant D90 and the postimplant D90 as a scatterplot Correlation between the postimplant D90 and the D90 on Day 30 as a scatterplot Neither the preimplant nor the postimplant V100 showed a clear correlation with the V100 on Day 30. The postimplant V100 had the better predictive ability (AUC = 0.64, p = 0.015) for a V100 of < 85% (14% of patients) on Day 30 than the preimplant V100 (p = 0.08). The median follow-up for patients without BR was 96 months (IQR, 66-108). The 5-year and 8-year BR-free rate was 96% and 90%, respectively. To identify cut-off values predictive of BR, we used quartiles of the D90 of the preoperative, postoperative, and Day 30 values. None of these quartiles showed any predictive value. This was also true when applied for a cut-off of 130 Gy for the D90 of Day 30 (detailed results not shown, all p ≥ 0.2). The year of the implant also did not have an influence on BR (p = 0.8). To account for a possible learning-curve, we repeated the analysis, this time excluding the first 10% (n = 23) of patients. None of the tested quartiles or the D90 on Day 30 was predictive of BR. Neither were such factors as Gleason score, PSA, or PSA-density (results not shown).

Discussion

We present one of the largest studies on IO planning and dose-delivery for prostate LDR-BT with a long follow-up. We discovered that a high D90 on the preplanning as well as on the postoperative TRUS were predictive of a good (≥ 130 Gy) D90. In addition, a high V100 on the postoperative TRUS dosimetry was significantly (p = 0.015) predictive of a V100 ≥ 85%, although the AUC (0.64) was low, making this measurement an unreliable factor. However, due to the relatively low AUC value (< 0.7) of all values and a lack of correlation between the values on Day 30 and the IO values, we determined that the IO dosimetry cannot reliably predict for Day 30 dosimetry. Therefore, Day 30 dosimetry remains the gold-standard in the evaluation of implant quality. In fact, many studies published before about 2010 have shown Day 30 dosimetry to predict for BR [7]. With a median follow-up of 8 years, we sought to identify cut-offs from IO dosimetry to predict BR. The fact that the quartiles, including the worst versus best quartiles of IO and Day 30, did not predict for BR, is likely due to the small sample size of 238 patients. Others have described positive results in patients treated with intraoperative planning [8]. Indeed, our results are similar to a recent publication by Shukla et al. [9] using real-time intraoperative planning. They describe an eight-year BR-free rate of 87-92%, comparable to our 90% and, as in our present study, an absence of any dosimetric parameters predicting BR. Another reason for the lack of predictive value of dosimetry might be that there are many other factors that may play a role in outcome determination such as classical risk factors for biochemical recurrence including PSA and Gleason score as well as PSA density, which was recently described by our group [10]. Day 30 dosimetry is more accurate and less interobserver-dependent when performed with magnetic resonance imaging (MRI)-CT fusion than with CT alone, and is more accurate for prostate-sector analysis [11,12]. In fact, the use of MRI in the evaluation of postimplant dosimetry has recently been recommended by the American College of Radiology (ACR) [13]. The reasons why dosimetry decreases from preimplant to postimplant, and then to Day 30 are manifold. First, the decrease in dosimetry from preimplant to postimplant is influenced by factors such as non-anticipated pubic-arch interference and prostate swelling during the implant. The decrease to Day 30 is influenced by difficulties during the procedure in correctly identifying the prostate, and probably most importantly, intraprostatic migration after the implant [14] and seeds being displaced by the movement of the needle, or the creation of a vacuum or tunnel in the prostatic tissue by the needle. Another factor influencing dosimetry could be urinary catheter placement [15], but no catheter was placed on Day 30, although catheters are in place intraoperatively. Furthermore, we didn’t evaluate interobserver or intraobserver variability in contouring the prostate on TRUS, which is lower than on CT and has a smaller effect on dosimetry. This is due to the better soft-tissue contrast at the prostate border on TRUS [16,17]. It is known that very small changes on CT contouring can have a large effect on dosimetry [18]. Unfortunately, with the present software, we find it very difficult to analyze which seeds were displaced and by how much. Evaluation and comparison of prostate contours before and after the implant are difficult. In our study, the preimplant dosimetry was done after the implantation of 2 fixation needles, which caused significant prostate edema. Although we previously found that the edema caused by the implant compared to the prostate volume after the insertion of the fixation needles was small (mean 1.9 cc) [19]; the prostate borders are more difficult to accurately define after the implant because of the edema and other artifacts caused by the procedure. The best way to analyze the reasons for the decline in dosimetry would be to use magnetic resonance imaging (MRI) immediately after the implant as well as on Day 30. Another more feasible option to improve intraoperative prediction of Day 30 dosimetry is the use of intraoperative fusion computed tomography [20]. The major limitation of IO planning with most currently available IO planning systems is the inability to adequately update the delivered seed position [1]. This may explain the absence of a meaningful intraoperative prediction of Day 30 dosimetry and BR in our present study. Other weak points of our study are that there was no IO protocol for the several physicists in the OR. Therefore, planning practices as well as when and how to update the needle position, may have differed between physicists. The treated patients represent the first cases of LDR-BT in our department. Although learning-curve effect on the implant quality and technique is probable, our long-term biochemical results are comparable with the literature. The same physician contoured the prostates on TRUS as well as on CT on Day 30. We therefore did not account for interobserver variability in contouring. Intraoperative planning techniques were developed to overcome some of the shortfalls of the classic preplanning approach. These shortfalls include difficulties in matching the prostate TRUS image from the preplan to the IO position, adjusting for prostate shape and volume changes, and the additional work and discomfort to the patients caused by preplanning study. As such, IO planning has many advantages. Real-time IO treatment planning software with inverse planning optimization has been shown to achieve excellent implants [21,22]. Several centers have shown better dosimetric results from IO planning, as reviewed by Polo et al. [2]. To our knowledge, this is the first study with an exceptionally long follow-up, analyzing the impact of IO preimplant and postimplant dosimetry on BR. Intraoperative 3D dose-planning has been shown to reduce the dose to the rectum and urethra, and allows for a reduction in numbers of seeds and mCi delivered [23]. IO with inverse optimization, as used in our cohort, has also been shown to produce less urethral dose, and a better implant than computer-assisted optimization methods modified peripheral and geometric optimization. Furthermore, it has been shown to result in a higher median V10 and D90 with fewer needles and seeds [24].

Conclusions

In conclusion, we found that IO dosimetry was predictive of Day 30 dosimetry, but with a low predictive value. We could not identify IO dosimetry that would predict biochemical outcome. We, therefore, conclude that postoperative CT- and MRI-based dosimetry remains the gold standard for evaluation of implant quality [12,25]. The incorporation of functional imaging, MRI-based IO planning, or use of robotics in seed delivery requires further study to determine if they will improve IO planning.

Disclosure

Authors report no conflict of interest.
  25 in total

1.  Urethra-sparing, intraoperative, real-time planned, permanent-seed prostate brachytherapy: toxicity analysis.

Authors:  Thomas Zilli; Daniel Taussky; David Donath; Hoa Phong Le; Renée-Xavière Larouche; Dominique Béliveau-Nadeau; Yannick Hervieux; Guila Delouya
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-04-20       Impact factor: 7.038

2.  MR and CT image fusion for postimplant analysis in permanent prostate seed implants.

Authors:  Alfredo Polo; Federica Cattani; Andrea Vavassori; Daniela Origgi; Gaetano Villa; Hugo Marsiglia; Massimo Bellomi; Giampiero Tosi; Ottavio De Cobelli; Roberto Orecchia
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-12-01       Impact factor: 7.038

3.  Prostate seed implantation using 3D-computer assisted intraoperative planning vs. a standard look-up nomogram: Improved target conformality with reduction in urethral and rectal wall dose.

Authors:  Adam Raben; Hansen Chen; Arnold Grebler; Jules Geltzeiler; Matt Geltzeiler; Ira Keselman; Sam Litvin; Sang Sim; Alex Hanlon; Jack Yang
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-12-01       Impact factor: 7.038

4.  The effect of interobserver variability on transrectal ultrasonography-based postimplant dosimetry.

Authors:  Jinyu Xue; Frank Waterman; Jay Handler; Eric Gressen
Journal:  Brachytherapy       Date:  2006 Jul-Sep       Impact factor: 2.362

5.  Bypassing the learning curve in permanent seed implants using state-of-the-art technology.

Authors:  Luc Beaulieu; Dee-Ann Radford Evans; Sylviane Aubin; Steven Angyalfi; Siraj Husain; Ian Kay; André-Guy Martin; Nicolas Varfalvy; Eric Vigneault; Peter Dunscombe
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-09-25       Impact factor: 7.038

Review 6.  Review of intraoperative imaging and planning techniques in permanent seed prostate brachytherapy.

Authors:  Alfredo Polo; Carl Salembier; Jack Venselaar; Peter Hoskin
Journal:  Radiother Oncol       Date:  2010-01-13       Impact factor: 6.280

7.  Comparison of CT and MR-CT fusion for prostate post-implant dosimetry.

Authors:  Kristina L Maletz; Ronald D Ennis; Jason Ostenson; Alexander Pevsner; Alexander Kagen; Iddo Wernick
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-05-06       Impact factor: 7.038

8.  The Quadrella: a novel approach to analyzing optimal outcomes after permanent seed prostate brachytherapy.

Authors:  Audrey Tétreault-Laflamme; Thomas Zilli; Aliza Meissner; Sandra Larrivée; Marie-Pierre Sylvestre; Guila Delouya; Daniel Taussky
Journal:  Radiother Oncol       Date:  2014-02-20       Impact factor: 6.280

9.  Seed loss in prostate brachytherapy : Operator dependency and impact on dosimetry.

Authors:  Nancy El-Bared; Natanel Sebbag; Dominic Béliveau-Nadeau; Yannick Hervieux; Renée Larouche; Daniel Taussky; Guila Delouya
Journal:  Strahlenther Onkol       Date:  2016-02-29       Impact factor: 3.621

10.  MRI-CT fusion to assess postbrachytherapy prostate volume and the effects of prolonged edema on dosimetry following transperineal interstitial permanent prostate brachytherapy.

Authors:  Juanita Crook; Michael McLean; Ivan Yeung; Theresa Williams; Gina Lockwood
Journal:  Brachytherapy       Date:  2004       Impact factor: 2.362

View more
  6 in total

1.  Evaluation of transrectal ultrasound-based dosimetry for brachytherapy of prostate cancer: a single-center experience.

Authors:  Changzhao Yang; Zhengtong Lv; Lingxiao Chen; Jie Wang; Xiheng Hu; Harripersaud Chand; Xi Sun; Guyu Tang; Congyi Tang; Huichuan Jiang; Yuan Li
Journal:  J Contemp Brachytherapy       Date:  2020-08-21

2.  Superior Postimplant Dosimetry Achieved Using Dynamic Intraoperative Dosimetry for Permanent Prostate Brachytherapy.

Authors:  Tanmay Singh; Junghoon Lee; Marianna Zahurak; Hee Joon Bae; Tamey Habtu; Robert Hobbs; Yi Le; Everette C Burdette; Daniel Y Song
Journal:  Pract Radiat Oncol       Date:  2021-03-13

3.  Acute and late toxicities in localized prostate cancer patients treated with low-dose 125I brachytherapy (110 Gy) in combination with external beam radiation therapy versus brachytherapy alone (160 Gy).

Authors:  Yuki Mukai; Narihiko Hayashi; Izumi Koike; Hisashi Kaizu; Shoko Takano; Madoka Sugiura; Eiko Ito; Mizuki Sato; Hiroji Uemura; Masahiro Yao; Masaharu Hata
Journal:  J Contemp Brachytherapy       Date:  2018-10-31

4.  Plan reproducibility of intraoperatively custom-built linked seeds compared to loose seeds for prostate brachytherapy.

Authors:  Tomoya Kaneda; Toshio Ohashi; Masanori Sakayori; Shinya Sutani; Shoji Yamashita; Tetsuo Momma; Shinichi Takahashi; Takashi Hanada; Naoyuki Shigematsu
Journal:  J Contemp Brachytherapy       Date:  2018-08-31

5.  Dosimetric outcomes of preoperative treatment planning with intraoperative optimization using stranded seeds in prostate brachytherapy.

Authors:  Jason Joon Bock Lee; Eungman Lee; Won Hoon Choi; Jihun Kim; Kyung Hwan Chang; Dong Wook Kim; Han Back Shin; Tae Hyung Kim; Hwa Kyung Byun; Jaeho Cho
Journal:  PLoS One       Date:  2022-03-30       Impact factor: 3.240

6.  Correlations of post-implant regional dosimetric parameters at 24 hours and one month, with clinical results of low-dose-rate brachytherapy for localized prostate cancer.

Authors:  Eiichiro Okazaki; Katsuyuki Kuratsukuri; Kentaro Ishii; Tomoaki Tanaka; Ryo Ogino; Tomohiro Nishikawa; Hideyuki Morimoto; Masako Hosono; Yukio Miki
Journal:  J Contemp Brachytherapy       Date:  2017-12-30
  6 in total