XiaoBo Zheng1, FangJun Bao1, Brendan Geraghty2, JinHai Huang1, Ayong Yu1, QinMei Wang1.
Abstract
BACKGROUD: To evaluate the symmetry of corneal biomechanical metrics, measured using an ocular response analyzer (ORA) and self-built corneal inflation test platform, in bilateral rabbit corneas and to investigate their relationship with physical intraocular pressure (IOPp).Entities:
Keywords: Corneal biomechanical metrics; Corneal stiffness; Intercorneal symmetry; Ocular response analyzer
Year: 2016 PMID: 26949709 PMCID: PMC4779259 DOI: 10.1186/s40662-016-0037-7
Source DB: PubMed Journal: Eye Vis (Lond) ISSN: 2326-0254
Average and standard deviation values of corneal shape parameters between right and left eyes
| Ocular parameters | Right eyes | Left eyes | t | P |
|---|---|---|---|---|
| R (mm) | 5.9 ± 0.89 | 5.97 ± 0.92 | −0.23 | 0.83 |
| p | 0.59 ± 0.36 | 0.6 ± 0.37 | −0.13 | 0.90 |
| CCT (μm) | 360.7 ± 12.1 | 358.7 ± 15.7 | 0.32 | 0.76 |
| PCT (μm) | 356.3 ± 13.4 | 351.4 ± 15.2 | 1.06 | 0.32 |
| CD (mm) | 12.87 ± 0.47 | 12.98 ± 0.47 | −1.87 | 0.10 |
R= corneal radius, p= corneal shape factor, CCT= central corneal thickness, PCT= peripheral corneal thickness, CD= corneal diameter
Average and standard deviation values of CH and CRF between right and left eyes at different pressure levels
| Biomechanical parameters | Pressure | Right eye | Left eye | Interocular differences |
|---|---|---|---|---|
| CH | 7.5 mmHg | 4.07 ± 1.07 | 3.83 ± 0.63 | 0.24 ± 1.39 |
| 15.0 mmHg | 5.02 ± 1.12 | 5.05 ± 0.97 | −0.04 ± 1.72 | |
| 22.5 mmHg | 5.41 ± 1.92 | 4.87 ± 1.46 | 0.54 ± 2.4 | |
| 30.0 mmHg | 5.68 ± 1.45 | 4.95 ± 2.16 | 0.73 ± 2.41 | |
| 37.5 mmHg | 5.86 ± 2.77 | 4.56 ± 1.80 | 1.02 ± 3.25 | |
| CRF | 7.5 mmHg | 0.79 ± 0.74 | 0.71 ± 0.56 | 0.31 ± 0.79 |
| 15.0 mmHg | 3.38 ± 1.02 | 3.12 ± 0.74 | 0.32 ± 1.42 | |
| 22.5 mmHg | 5.30 ± 1.54 | 5.01 ± 1.09 | 0.28 ± 1.59 | |
| 30.0 mmHg | 7.81 ± 1.30 | 7.11 ± 1.80 | 0.7 ± 1.68 | |
| 37.5 mmHg | 10.01 ± 2.01 | 8.64 ± 1.70 | 1.11 ± 2.39 |
CH= corneal hysteresis, CRF= corneal resistance factor
Fig. 1Pressure-rise behavior of corneal specimens between right and left eyes, "a-j" means specimen 1 to specimen 10
Average and standard deviation values of tangent modulus between right and left eyes at different stress levels
| Biomechanical parameters | Stress (MPa) | Right eye | Left eye | Interocular differences |
|---|---|---|---|---|
| Tangent Modulus | 0.001 | 0.5 ± 0.39 | 0.86 ± 0.83 | −0.36 ± 0.87 |
| 0.002 | 0.6 ± 0.39 | 0.97 ± 0.82 | −0.37 ± 0.89 | |
| 0.003 | 0.71 ± 0.4 | 1.09 ± 0.82 | −0.38 ± 0.92 | |
| 0.004 | 0.82 ± 0.42 | 1.21 ± 0.83 | −0.39 ± 0.95 | |
| 0.005 | 0.94 ± 0.43 | 1.34 ± 0.83 | −0.41 ± 0.98 |
Fig. 2Comparison of average stress-strain behavior between right and left eyes. The error bars represent the standard deviation of stress values