Effect of fluorescein dye staining of the tear film on Scheimpflug measurements of central corneal thickness.

Nino Hirnschall1, Alja Crnej, Vinod Gangwani, Oliver Findl.   

Abstract

PURPOSE: To assess whether fluorescein dye in the tear film results in a clinically relevant change in central corneal thickness (CCT), as measured with Scheimpflug imaging (Pentacam HR; Oculus, Wetzlar, Germany). The study was conducted at Moorfields Eye Hospital NHS Foundation Trust.
METHODS: In this prospective, randomized, controlled, single-center study on healthy volunteers, CCT was measured using rotating Scheimpflug imaging and partial coherence interferometry technique. After 3 baseline measurements of both eyes, a drop of 0.25% fluorescein with 0.5% proxymetacaine hydrochloride dye, as used for measuring intraocular pressure with applanation Goldmann tonometry, was instilled in the first eye and measurements were repeated after 1 minute and 5, 10, 20, 40, and 60 minutes. A fluorescein drop was then instilled in the second eye, and 1 measurement was performed after 1 minute. Immediately afterwards, the second eye was rinsed with 2 ml of balanced saline solution and the CCT measurement was repeated. On a different day, a Schirmer II test was performed.
RESULTS: Thirty eyes of 15 volunteers were included in this study. The average CCT measured with Scheimpflug imaging was 539.1 μm (SD, 32.2 μm), and after instilling fluorescein, CCT values increased by 46.6 μm (SD, 11.4 μm; P < 0.01, paired t test). In 10 of 15 subjects, the fluorescein layer was still present 40 minutes after instilling the drop. Rinsing the eye decreased the fluorescein layer by 49.6% and a value of 24.8 μm (SD, 16.8 μm).
CONCLUSIONS: Because CCT measurements are critical for corneal refractive surgery, Scheimpflug imaging should not be performed after use of fluorescein dye for staining of the corneal epithelium or for applanation tonometry.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22001815     DOI: 10.1097/ICO.0b013e31821eea97

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  5 in total

1.  Evaluation of corneal parameters in patients with Parkinson's disease.

Authors:  Seden Demirci; Alime Gunes; Hasan Rifat Koyuncuoglu; Levent Tok; Ozlem Tok
Journal:  Neurol Sci       Date:  2016-04-07       Impact factor: 3.307

2.  Effect of 2% fluorescein on Scheimpflug central corneal thickness measurements.

Authors:  Stella Briggs; Marwa Bin Moammar
Journal:  Int J Ophthalmol       Date:  2016-02-18       Impact factor: 1.779

3.  Factors Influencing Efficacy of Peripheral Corneal Relaxing Incisions during Cataract Surgery.

Authors:  Nino Hirnschall; Jörg Wiesinger; Petra Draschl; Oliver Findl
Journal:  J Ophthalmol       Date:  2015-06-25       Impact factor: 1.909

4.  Effect of Proparacaine 0.375%-Sodium Fluorescein 0.25% Eye Drop Mixture and Fluorescein Strip on Anterior Segment Parameters.

Authors:  Mustafa Doğan; Mehmet Cem Sabaner; Mehmet Akif Erol
Journal:  J Ophthalmol       Date:  2018-09-05       Impact factor: 1.909

5.  Evaluation of Central and Peripheral Corneal Thicknesses in Patients with Systemic Lupus Erythematosus.

Authors:  Iman M Eissa; Ghada A Nassar; Shaimaa A Arfeen; Ahmed A Dahab
Journal:  Clin Ophthalmol       Date:  2020-02-28
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.