Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy.

A Fogliata1, R Garcia, T Knoos, G Nicolini, A Clivio, E Vanetti, C Khamphan, L Cozzi.   

Abstract

PURPOSE: Flattening filter free (FFF) beams generated by medical linear accelerators have recently started to be used in radiotherapy clinical practice. Such beams present fundamental differences with respect to the standard filter flattened (FF) beams, making the generally used dosimetric parameters and definitions not always viable. The present study will propose possible definitions and suggestions for some dosimetric parameters for use in quality assurance of FFF beams generated by medical linacs in radiotherapy.
METHODS: The main characteristics of the photon beams have been analyzed using specific data generated by a Varian TrueBeam linac having both FFF and FF beams of 6 and 10 MV energy, respectively.
RESULTS: Definitions for dose profile parameters are suggested starting from the renormalization of the FFF with respect to the corresponding FF beam. From this point the flatness concept has been translated into one of "unflatness" and other definitions have been proposed, maintaining a strict parallelism between FFF and FF parameter concepts.
CONCLUSIONS: Ideas for quality controls used in establishing a quality assurance program when introducing FFF beams into the clinical environment are given here, keeping them similar to those used for standard FF beams. By following the suggestions in this report, the authors foresee that the introduction of FFF beams into a clinical radiotherapy environment will be as safe and well controlled as standard beam modalities using the existing guidelines.

Entities:  

Mesh:

Year:  2012        PMID: 23039680     DOI: 10.1118/1.4754799

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

1.  Evaluation of a Small Animal Irradiation System for Animal Experiments Using EBT3 Model GAFCHROMIC™ Film.

Authors:  Yasuyuki Shimizu; Hiroaki Akasaka; Daisuke Miyawaki; Naritoshi Mukumoto; Masao Nakayama; Tianyuan Wang; Saki Osuga; Sachiko Inubushi; Ryuichi Yada; Yasuo Ejima; Kenji Yoshida; Takeaki Ishihara; Ryohei Sasaki
Journal:  Kobe J Med Sci       Date:  2018-01-17

2.  Automatically gated image-guided breath-hold IMRT is a fast, precise, and dosimetrically robust treatment for lung cancer patients.

Authors:  Anna Simeonova-Chergou; Anika Jahnke; Kerstin Siebenlist; Florian Stieler; Sabine Mai; Judit Boda-Heggemann; Frederik Wenz; Frank Lohr; Lennart Jahnke
Journal:  Strahlenther Onkol       Date:  2016-01-15       Impact factor: 3.621

3.  Determination of an inflection point for a dosimetric analysis of unflattened beam using the first principle of derivatives by python code programming.

Authors:  Ravindra Shende; Gourav Gupta; Subash Macherla
Journal:  Rep Pract Oncol Radiother       Date:  2019-07-29

4.  Determination of inflection points of CyberKnife dose profiles within acceptability criteria of deviations in measurements.

Authors:  Neslihan Sarıgül; Fazlı Yağız Yedekçi; Mete Yeğiner; Fadıl Akyol; Haluk Utku
Journal:  Rep Pract Oncol Radiother       Date:  2019-11-12

5.  Survey on utilization of flattening filter-free photon beams in Japan.

Authors:  Takumi Kodama; Keisuke Yasui; Shie Nishioka; Kazunori Miyaura; Toru Takakura; Tetsurou Katayose; Mitsuhiro Nakamura
Journal:  J Radiat Res       Date:  2021-07-10       Impact factor: 2.724

6.  Evaluation of the Machine Performance Check application for TrueBeam Linac.

Authors:  Alessandro Clivio; Eugenio Vanetti; Steven Rose; Giorgia Nicolini; Maria F Belosi; Luca Cozzi; Christof Baltes; Antonella Fogliata
Journal:  Radiat Oncol       Date:  2015-04-21       Impact factor: 3.481

7.  Acceptance criteria for flattening filter-free photon beam from standard medical electron linear accelerator: AERB task group recommendations.

Authors:  G Sahani; S D Sharma; P K Dash Sharma; D D Deshpande; P S Negi; V K Sathianarayanan; G K Rath
Journal:  J Med Phys       Date:  2014-10

8.  Monte Carlo simulation of beam characteristics from small fields based on TrueBeam flattening-filter-free mode.

Authors:  Zhongsu Feng; Haizhen Yue; Yibao Zhang; Hao Wu; Jinsheng Cheng; Xu Su
Journal:  Radiat Oncol       Date:  2016-02-27       Impact factor: 3.481

9.  EPID-based dosimetry to verify IMRT planar dose distribution for the aS1200 EPID and FFF beams.

Authors:  Narges Miri; Peter Keller; Benjamin J Zwan; Peter Greer
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

10.  Dosimetric properties of equivalent-quality flattening filter-free (FFF) and flattened photon beams of Versa HD linear accelerator.

Authors:  Mukesh N Meshram; Srimanta Pramanik; C P Ranjith; Saravana K Gopal; Rishabh Dobhal
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102