Polymer thick film technology for improved simultaneous dEEG/MRI recording: Safety and MRI data quality.

Catherine Poulsen1, Daniel G Wakeman2, Seyed Reza Atefi2, Phan Luu1, Amy Konyn1, Giorgio Bonmassar2.   

Abstract

PURPOSE: To develop a 256-channel dense-array electroencephalography (dEEG) sensor net (the Ink-Net) using high-resistance polymer thick film (PTF) technology to improve safety and data quality during simultaneous dEEG/MRI.
METHODS: Heating safety was assessed with temperature measurements in an anthropomorphic head phantom during a 30-min, induced-heating scan at 7T. MRI quality assessment used B1 field mapping and functional MRI (fMRI) retinotopic scans in three humans at 3T. Performance of the 256-channel PTF Ink-Net was compared with a 256-channel MR-conditional copper-wired electroencephalography (EEG) net and to scans with no sensor net. A visual evoked potential paradigm assessed EEG quality within and outside the 3T scanner.
RESULTS: Phantom temperature measurements revealed nonsignificant heating (ISO 10974) in the presence of either EEG net. In human B1 field and fMRI scans, the Ink-Net showed greatly reduced cross-modal artifact and less signal degradation than the copper-wired net, and comparable quality to MRI without sensor net. Cross-modal ballistocardiogram artifact in the EEG was comparable for both nets.
CONCLUSION: High-resistance PTF technology can be effectively implemented in a 256-channel dEEG sensor net for MR conditional use at 7T and with significantly improved structural and fMRI data quality as assessed at 3T. Magn Reson Med 77:895-903, 2017.
© 2016 International Society for Magnetic Resonance in Medicine. © 2016 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  B1 maps; ISO10974; anthropomorphic head phantom; eccentricity maps; fMRI/BOLD; temperature measurements

Mesh:

Substances:

Year:  2016        PMID: 26876960      PMCID: PMC4985508          DOI: 10.1002/mrm.26116

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  45 in total

1.  Empirical validation of SAR values predicted by FDTD modeling.

Authors:  P Gajsek; T J Walters; W D Hurt; J M Ziriax; D A Nelson; P A Mason
Journal:  Bioelectromagnetics       Date:  2002-01       Impact factor: 2.010

Review 2.  Safety of strong, static magnetic fields.

Authors:  J F Schenck
Journal:  J Magn Reson Imaging       Date:  2000-07       Impact factor: 4.813

3.  Influence of dense-array EEG cap on fMRI signal.

Authors:  Qingfei Luo; Gary H Glover
Journal:  Magn Reson Med       Date:  2011-12-09       Impact factor: 4.668

4.  Simultaneous 3-T fMRI and high-density recording of human auditory evoked potentials.

Authors:  Carrie J Scarff; Angela Reynolds; Bradley G Goodyear; Curtis W Ponton; Joseph C Dort; Jos J Eggermont
Journal:  Neuroimage       Date:  2004-11       Impact factor: 6.556

5.  Removal of FMRI environment artifacts from EEG data using optimal basis sets.

Authors:  R K Niazy; C F Beckmann; G D Iannetti; J M Brady; S M Smith
Journal:  Neuroimage       Date:  2005-09-16       Impact factor: 6.556

6.  Temperature and SAR measurement errors in the evaluation of metallic linear structures heating during MRI using fluoroptic probes.

Authors:  E Mattei; M Triventi; G Calcagnini; F Censi; W Kainz; H I Bassen; P Bartolini
Journal:  Phys Med Biol       Date:  2007-02-27       Impact factor: 3.609

7.  Ballistocardiogram artifact removal with a reference layer and standard EEG cap.

Authors:  Qingfei Luo; Xiaoshan Huang; Gary H Glover
Journal:  J Neurosci Methods       Date:  2014-06-22       Impact factor: 2.390

8.  An open-source hardware and software system for acquisition and real-time processing of electrophysiology during high field MRI.

Authors:  Patrick L Purdon; Hernan Millan; Peter L Fuller; Giorgio Bonmassar
Journal:  J Neurosci Methods       Date:  2008-08-05       Impact factor: 2.390

9.  Effects of simultaneous EEG recording on MRI data quality at 1.5, 3 and 7 tesla.

Authors:  Karen Mullinger; Stefan Debener; Ronald Coxon; Richard Bowtell
Journal:  Int J Psychophysiol       Date:  2007-08-09       Impact factor: 2.997

10.  A novel brain stimulation technology provides compatibility with MRI.

Authors:  Peter Serano; Leonardo M Angelone; Husam Katnani; Emad Eskandar; Giorgio Bonmassar
Journal:  Sci Rep       Date:  2015-04-29       Impact factor: 4.379

View more
  7 in total

1.  Numerical and Experimental Analysis of Radiofrequency-Induced Heating Versus Lead Conductivity During EEG-MRI at 3 T.

Authors:  Seyed Reza Atefi; Peter Serano; Catherine Poulsen; Leonardo M Angelone; Giorgio Bonmassar
Journal:  IEEE Trans Electromagn Compat       Date:  2018-06-25       Impact factor: 2.006

2.  Numerical simulation of the radiofrequency safety of 128-channel hd-EEG nets on a 29-month-old whole-body model in a 3 Tesla MRI.

Authors:  Hongbae Jeong; Georgios Ntolkeras; P Ellen Grant; Giorgio Bonmassar
Journal:  IEEE Trans Electromagn Compat       Date:  2021-08-16       Impact factor: 2.036

3.  Multimodal EEG-fMRI: advancing insight into large-scale human brain dynamics.

Authors:  Catie Chang; Jingyuan E Chen
Journal:  Curr Opin Biomed Eng       Date:  2021-03-15

Review 4.  EEG-Informed fMRI: A Review of Data Analysis Methods.

Authors:  Rodolfo Abreu; Alberto Leal; Patrícia Figueiredo
Journal:  Front Hum Neurosci       Date:  2018-02-06       Impact factor: 3.169

5.  A temporal sequence of thalamic activity unfolds at transitions in behavioral arousal state.

Authors:  Beverly Setzer; Nina E Fultz; Daniel E P Gomez; Stephanie D Williams; Giorgio Bonmassar; Jonathan R Polimeni; Laura D Lewis
Journal:  Nat Commun       Date:  2022-09-16       Impact factor: 17.694

6.  Segmenting electroencephalography wires reduces radiofrequency shielding artifacts in simultaneous electroencephalography and functional magnetic resonance imaging at 7 T.

Authors:  Thanh Phong Lê; Rolf Gruetter; João Jorge; Özlem Ipek
Journal:  Magn Reson Med       Date:  2022-05-16       Impact factor: 3.737

7.  MRI-Induced Heating of Coils for Microscopic Magnetic Stimulation at 1.5 Tesla: An Initial Study.

Authors:  Giorgio Bonmassar; Peter Serano
Journal:  Front Hum Neurosci       Date:  2020-03-13       Impact factor: 3.169

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.