Temperature and SAR measurement errors in the evaluation of metallic linear structures heating during MRI using fluoroptic probes.

E Mattei1, M Triventi, G Calcagnini, F Censi, W Kainz, H I Bassen, P Bartolini.   

Abstract

The purpose of this work is to evaluate the error associated with temperature and SAR measurements using fluoroptic temperature probes on pacemaker (PM) leads during magnetic resonance imaging (MRI). We performed temperature measurements on pacemaker leads, excited with a 25, 64, and 128 MHz current. The PM lead tip heating was measured with a fluoroptic thermometer (Luxtron, Model 3100, USA). Different contact configurations between the pigmented portion of the temperature probe and the PM lead tip were investigated to find the contact position minimizing the temperature and SAR underestimation. A computer model was used to estimate the error made by fluoroptic probes in temperature and SAR measurement. The transversal contact of the pigmented portion of the temperature probe and the PM lead tip minimizes the underestimation for temperature and SAR. This contact position also has the lowest temperature and SAR error. For other contact positions, the maximum temperature error can be as high as -45%, whereas the maximum SAR error can be as high as -54%. MRI heating evaluations with temperature probes should use a contact position minimizing the maximum error, need to be accompanied by a thorough uncertainty budget and the temperature and SAR errors should be specified.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17327653     DOI: 10.1088/0031-9155/52/6/006

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  16 in total

1.  Feasibility of real-time magnetic resonance imaging-guided endomyocardial biopsies: An in-vitro study.

Authors:  Dirk Lossnitzer; Sebastian A Seitz; Birgit Krautz; Bernhard Schnackenburg; Florian André; Grigorios Korosoglou; Hugo A Katus; Henning Steen
Journal:  World J Cardiol       Date:  2015-07-26

2.  It's the little things: On the complexity of planar electrode heating in MRI.

Authors:  Johannes B Erhardt; Thomas Lottner; Jessica Martinez; Ali C Özen; Martin Schuettler; Thomas Stieglitz; Daniel B Ennis; Michael Bock
Journal:  Neuroimage       Date:  2019-03-29       Impact factor: 6.556

3.  Assessment of MRI issues at 7 T.

Authors:  Terry O Woods; Leonardo M Angelone; Jana G Delfino
Journal:  AJR Am J Roentgenol       Date:  2014-11       Impact factor: 3.959

4.  Polymer thick film technology for improved simultaneous dEEG/MRI recording: Safety and MRI data quality.

Authors:  Catherine Poulsen; Daniel G Wakeman; Seyed Reza Atefi; Phan Luu; Amy Konyn; Giorgio Bonmassar
Journal:  Magn Reson Med       Date:  2016-02-15       Impact factor: 4.668

5.  Direct cooling of the catheter tip increases safety for CMR-guided electrophysiological procedures.

Authors:  Theresa Reiter; Daniel Gensler; Oliver Ritter; Ingo Weiss; Wolfgang Geistert; Ralf Kaufmann; Sabine Hoffmeister; Michael T Friedrich; Stefan Wintzheimer; Markus Düring; Peter Nordbeck; Peter M Jakob; Mark E Ladd; Harald H Quick; Wolfgang R Bauer
Journal:  J Cardiovasc Magn Reson       Date:  2012-02-01       Impact factor: 5.364

6.  Reconfigurable MRI technology for low-SAR imaging of deep brain stimulation at 3T: Application in bilateral leads, fully-implanted systems, and surgically modified lead trajectories.

Authors:  Ehsan Kazemivalipour; Boris Keil; Alireza Vali; Sunder Rajan; Behzad Elahi; Ergin Atalar; Lawrence L Wald; Joshua Rosenow; Julie Pilitsis; Laleh Golestanirad
Journal:  Neuroimage       Date:  2019-05-13       Impact factor: 6.556

7.  Measurements of RF heating during 3.0-T MRI of a pig implanted with deep brain stimulator.

Authors:  Krzysztof R Gorny; Michael F Presti; Stephan J Goerss; Sun C Hwang; Dong-Pyo Jang; Inyong Kim; Hoon-Ki Min; Yunhong Shu; Christopher P Favazza; Kendall H Lee; Matt A Bernstein
Journal:  Magn Reson Imaging       Date:  2012-12-07       Impact factor: 2.546

8.  In-vitro mapping of E-fields induced near pacemaker leads by simulated MR gradient fields.

Authors:  Howard I Bassen; Gonzalo G Mendoza
Journal:  Biomed Eng Online       Date:  2009-12-15       Impact factor: 2.819

9.  RF Safety Evaluation of a Breast Tissue Expander Device for MRI: Numerical Simulation and Experiment.

Authors:  Bu S Park; Amir Razjouyan; Leonardo M Angelone; Brent McCright; Sunder S Rajan
Journal:  IEEE Trans Electromagn Compat       Date:  2017-10       Impact factor: 2.006

10.  Safety of localizing epilepsy monitoring intracranial electroencephalograph electrodes using MRI: radiofrequency-induced heating.

Authors:  David W Carmichael; John S Thornton; Roman Rodionov; Rachel Thornton; Andrew McEvoy; Philip J Allen; Louis Lemieux
Journal:  J Magn Reson Imaging       Date:  2008-11       Impact factor: 4.813

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.