Lalit Kumar1, Girigesh Yadav1, Kothanda Raman1, Manindra Bhushan1, Manoj Pal1.
Abstract
The main purpose of this study is to know the effect of three different photon energies viz., 6, 10, and 15 mega voltage (MV) on RapidArc (RA) planning for deep-seated cervix tumor and to develop clinically acceptable RA plans with suitable photon energy. RA plans were generated for 6, 10, and 15 MV photon energies for twenty patients reported with cervix carcinoma. RA plans were evaluated in terms of planning target volume (PTV) coverage, dose to organs at risk (OARs), conformity index (CI), homogeneity index (HI), gradient measure, external volume index of dose distribution produced, total number of monitor units (MUs), nontumor integral dose (ID), and low dose volume of normal tissue. A two-sample paired t-test was performed to compare the dosimetric parameters of RA plans. Irrespective of photon energy used for RA planning, plans were dosimetrically similar in terms of PTV coverage, OARs sparing, CI and HI. The numbers of MUs were 13.4 ± 1.4% and 18.2 ± 1.5% higher and IDs were 2.7 ± 0.8% and 3.7 ± 0.9% higher in 6 MV plans in comparison to that in the 10 and 15 MV plans, respectively. V1Gy, V2Gy, V3Gy, and V4Gy were higher in 6 MV plans in comparison to that in 10 and 15 MV plans. Based on this study, 6 MV photon beam is a good choice for RA planning in case of cervix carcinoma, as it does not deliver additional exposure to patients caused by photoneutrons produced in high energy beams.Entities:
Keywords: Dose distribution; RapidArc; neutron dose; photon energy
Year: 2015 PMID: 26865756 PMCID: PMC4728891 DOI: 10.4103/0971-6203.170787
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
DVH dosimetric data of PTV for RA using 6, 10, and 15 MV photon energies
Figure 1The isodose distribution generated from RapidArc planning in case of Ca-Cervix for same patient in axial, coronal and sagittal planes with (a) 6 MV, (b) 10 MV, and (c) 15 MV photon beam energies
Plan comparison parameters for RA plans using 6, 10, and 15 MV energies
Dose-volume parameter for different OARs for RA plans using 6, 10, and 15 MV energies
DVH parameters of normal tissues (body-PTV) for NTID and low doses from RA plans using 6, 10, and 15 MV