Impact of acuros XB algorithm in deep-inspiration breath-hold (DIBH) respiratory techniques used for the treatment of left breast cancer.

Lalit Kumar1,2, Vimal Kishore3, Manindra Bhushan2, Abhinav Dewan2, Girigesh Yadav2, Kothanda Raman2, Gourav Kumar2, Irfan Ahmad2, Kundan S Chufal2, Munish Gairola2.   

Abstract

AIM: To investigate the impact of Acuros XB (AXB) algorithm in the deep-inspiration breath-hold (DIBH) technique used for treatment of left sided breast cancer.
BACKGROUND: AXB may estimate better lung toxicities and treatment outcome in DIBH.
MATERIALS AND METHODS: Treatment plans were computed using the field-in-field technique for a 6 MV beam in two respiratory phases - free breathing (FB) and DIBH. The AXB-calculations were performed under identical beam setup and the same numbers of monitor units as used for AAA-calculation.
RESULTS: Mean Hounsfield units (HU), mass density (g/cc) and relative electron density were -782.1 ± 24.8 and -883.5 ± 24.9; 0.196 ± 0.025 and 0.083 ± 0.032; 0.218 ± 0.025 and 0.117 ± 0.025 for the lung in the FB and DIBH respiratory phase, respectively. For a similar target coverage (p > 0.05) in the DIBH respiratory phase between the AXB and AAA algorithm, there was a slight increase in organ at risk (OAR) dose for AXB in comparison to AAA, except for mean dose to the ipsilateral lung. AAA predicts higher mean dose to the ipsilateral lung and lesser V20Gy for the ipsilateral and common lung in comparison to AXB. The differences in mean dose to the ipsilateral lung were 0.87 ± 2.66 % (p > 0.05) in FB, and 1.01 ± 1.07% (p < 0.05) in DIBH, in V20Gy the differences were 1.76 ± 0.83% and 1.71 ± 0.82% in FB (p < 0.05), 3.34 ± 1.15 % and 3.24 ± 1.17 % in DIBH (p < 0.05), for the ipsilateral and common lung, respectively.
CONCLUSION: For a similar target volume coverage, there were important differences between the AXB and AAA algorithm for low-density inhomogeneity medium present in the DIBH respiratory phase for left sided breast cancer patients. DIBH treatment in conjunction with AXB may result in better estimation of lung toxicities and treatment outcome.
© 2020 Greater Poland Cancer Centre. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Acuros XB (AXB) algorithm; Deep inspiratory breathhold, low density medium, AAA

Year:  2020        PMID: 32494224      PMCID: PMC7252204          DOI: 10.1016/j.rpor.2020.04.011

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  26 in total

1.  Measurement of lung density by means of quantitative CT scanning. A study of correlations with pulmonary function tests.

Authors:  A Heremans; J A Verschakelen; L Van fraeyenhoven; M Demedts
Journal:  Chest       Date:  1992-09       Impact factor: 9.410

Review 2.  Radiation dose-volume effects in the lung.

Authors:  Lawrence B Marks; Soren M Bentzen; Joseph O Deasy; Feng-Ming Spring Kong; Jeffrey D Bradley; Ivan S Vogelius; Issam El Naqa; Jessica L Hubbs; Joos V Lebesque; Robert D Timmerman; Mary K Martel; Andrew Jackson
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-03-01       Impact factor: 7.038

3.  Comparison of a finite-element multigroup discrete-ordinates code with Monte Carlo for radiotherapy calculations.

Authors:  Kent A Gifford; John L Horton; Todd A Wareing; Gregory Failla; Firas Mourtada
Journal:  Phys Med Biol       Date:  2006-04-19       Impact factor: 3.609

4.  The accuracy of the pencil beam convolution and anisotropic analytical algorithms in predicting the dose effects due to attenuation from immobilization devices and large air gaps.

Authors:  A Gray; L D Oliver; P N Johnston
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

5.  ESTRO consensus guideline on target volume delineation for elective radiation therapy of early stage breast cancer.

Authors:  Birgitte V Offersen; Liesbeth J Boersma; Carine Kirkove; Sandra Hol; Marianne C Aznar; Albert Biete Sola; Youlia M Kirova; Jean-Philippe Pignol; Vincent Remouchamps; Karolien Verhoeven; Caroline Weltens; Meritxell Arenas; Dorota Gabrys; Neil Kopek; Mechthild Krause; Dan Lundstedt; Tanja Marinko; Angel Montero; John Yarnold; Philip Poortmans
Journal:  Radiother Oncol       Date:  2015-01-24       Impact factor: 6.280

6.  On the dosimetric impact of inhomogeneity management in the Acuros XB algorithm for breast treatment.

Authors:  Antonella Fogliata; Giorgia Nicolini; Alessandro Clivio; Eugenio Vanetti; Luca Cozzi
Journal:  Radiat Oncol       Date:  2011-08-26       Impact factor: 3.481

7.  Dosimetric comparison of Acuros XB deterministic radiation transport method with Monte Carlo and model-based convolution methods in heterogeneous media.

Authors:  Tao Han; Justin K Mikell; Mohammad Salehpour; Firas Mourtada
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

8.  The dosimetric impact of different photon beam energy on RapidArc radiotherapy planning for cervix carcinoma.

Authors:  Lalit Kumar; Girigesh Yadav; Kothanda Raman; Manindra Bhushan; Manoj Pal
Journal:  J Med Phys       Date:  2015 Oct-Dec

9.  Critical appraisal of the accuracy of Acuros-XB and Anisotropic Analytical Algorithm compared to measurement and calculations with the compass system in the delivery of RapidArc clinical plans.

Authors:  Murugesan Kathirvel; Shanmuga Subramanian; Alessandro Clivio; Gandhi Arun; Antonella Fogliata; Giorgia Nicolini; Vellaiyan Subramani; Shanmugam Thirumalai Swamy; Eugenio Vanetti; Luca Cozzi
Journal:  Radiat Oncol       Date:  2013-06-11       Impact factor: 3.481

10.  Evaluation of the analytical anisotropic algorithm in an extreme water-lung interface phantom using Monte Carlo dose calculations.

Authors:  Isabelle M Gagné; Sergei Zavgorodni
Journal:  J Appl Clin Med Phys       Date:  2006-06-16       Impact factor: 2.102

View more
  5 in total

1.  Dosimetric impact of Acuros XB on cervix radiotherapy using RapidArc technique: a dosimetric study.

Authors:  Lalit Kumar; Vimal Kishore; Manindra Bhushan; Pawan Kumar; Rahul Lal Chaudhary
Journal:  Rep Pract Oncol Radiother       Date:  2021-08-12

2.  Three-dimensional conformal radiotherapy (3D-CRT) vs. volumetric modulated arc therapy (VMAT) in deep inspiration breath-hold (DIBH) technique in left-sided breast cancer patients-comparative analysis of dose distribution and estimation of projected secondary cancer risk.

Authors:  Iga Racka; Karolina Majewska; Janusz Winiecki
Journal:  Strahlenther Onkol       Date:  2022-08-09       Impact factor: 4.033

3.  Dosimetric Influence of Acuros XB Dose-to-Medium and Dose-to-Water Reporting Modes on Carcinoma Cervix Using Intensity-Modulated Radiation Therapy and Volumetric RapidArc Technique.

Authors:  Lalit Kumar; Manindra Bhushan; Vimal Kishore; Rahul Lal Chowdhary; Soumitra Barik; Anurag Sharma; Munish Gairola
Journal:  J Med Phys       Date:  2022-02-18

4.  Dosimetric Evaluation of the Treatment Plan on Indigenous Heterogeneous Phantoms using Analytical Anisotropic Algorithm and Acuros-XB Algorithm for Different Photon Energies.

Authors:  Vinod Kumar Gangwar; Om Prakash Gurjar; Lalit Kumar; Avinash Agarwal; Vineet Kumar Mishra; Surendra Prasad Mishra; Saket Pandey
Journal:  J Biomed Phys Eng       Date:  2022-06-01

5.  A comparative analysis of Acuros XB and the analytical anisotropic algorithm for volumetric modulation arc therapy.

Authors:  Raju P Srivastava; K Basta; Werner De Gersem; Carlos De Wagter
Journal:  Rep Pract Oncol Radiother       Date:  2021-06-09
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.