Sundaram Thangavelu1, S Jayakumar, K N Govindarajan, Sanjay S Supe, V Nagarajan, M Nagarajan.
Abstract
The goal of the present study was to study the effects of low- and high-energy intensity-modulated photon beams on the planning of target volume and the critical organs in cases of localized prostate tumors in a cohort of 8 patients. To ensure that the difference between the plans is due to energy alone, all other parameters were kept constant. A mean dose volume histogram (DVH) for each value of energy and for each contoured structure was created and was considered as completely representative for all patients. To facilitate comparison between 6-MV and 15-MV beams, the DVH-s were normalized. The different parameters that were compared for 6-MV and 15-MV beams included mean DVH, different homogeneity indices, conformity index, etc. Analysis of several indices depicts more homogeneous dose for 15-MV beam and more conformity for 6-MV beam. Comparison of all these parameters showed that there was little difference between the 6-MV and 15-MV beams. For rectum, 2 to 4 % more volume received high dose with the 6-MV beam in comparison with the 15-MV beam, which was not clinically significant, since in practice much tighter constraints are maintained, such that Normal Tissue Complication Probability (NTCP) is kept within 5 %. Such tighter constraints might increase the dose to other regions and other critical organs but are unlikely to increase their complication probabilities. Hence the slight advantages of 15-MV beam in providing benefits of better normal-tissue sparing and better coverage cannot be considered to outweigh its well-known risk of non-negligible neutron production.Entities:
Keywords: Energy; H-index; HI-index; Sigma-index; normal tissue complication probability; prostate intensity-modulated radiation therapy; tumor control probability
Year: 2011 PMID: 21430856 PMCID: PMC3048951 DOI: 10.4103/0971-6203.75469
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Constraints used for segmental inverse optimizer
Figure 1a: Mean dose volume histogram of Planning Target Volume
b: Mean dose volume histogram of rectum
c: Mean dose volume histogram of bladder
d: Mean dose volume histogram of body
e: Mean dose volume histogram of left femoral head
f: Mean dose volume histogram of right femoral head
H and Mod H-Indices for eight patient intensity-modulated radiation therapy plans with 6 MV and 15 MV X-rays
HI and Mod HI-indices for eight patient intensity-modulated radiation therapy plans with 6 MV and 15 MV X-rays
Sigma indices for eight patient intensitymodulated radiation therapy plans with 6 MV and 15 MV X-rays
Conformity indices for eight patient intensity-modulated radiation therapy plans with 6 MV and 15 MV X-rays
Figure 2Cumulative dose volume histogram of Patient 1(red) and Patient 3 (green) - 6 MV