| Literature DB >> 26839885 |
Philipp Mittmann1, Grit Rademacher2, Sven Mutze2, Frederike Hassepass3, Arneborg Ernst1, Ingo Todt1.
Abstract
The position of the cochlear implant electrode array within the scala tympani is essential for an optimal postoperative hearing benefit. If the electrode array changes in between the scalae intracochlearly (i.e., from scala tympani to scala vestibuli), a reduced auditory performance can be assumed. We established a neural response telemetry-ratio (NRT-ratio) which corresponds with the scalar position of the electrodes but shows within its limits a variability. The aim of this study was to determine if insertion depth angle or cochlea size influences the NRT-ratio. The intraoperative electrophysiological NRT data of 26 patients were evaluated. Using a flat panel tomography system, the position of the electrode array was evaluated radiologically. The insertion depth angle of the electrode, the cochlea size, and the NRT-ratio were calculated postoperatively. The radiological results were compared with the intraoperatively obtained electrophysiological data (NRT-ratio) and statistically evaluated. In all patients the NRT-ratio, the insertion depth angle, and the cochlea size could be determined. A significant correlation between insertional depth, cochlear size, and the NRT-ratio was not found. The NRT-ratio is a reliable electrophysiological tool to determine the scalar position of a perimodiolar electrode array. The NRT-ratio can be applied independent from insertion depth and cochlear size.Entities:
Mesh:
Year: 2015 PMID: 26839885 PMCID: PMC4709608 DOI: 10.1155/2015/706253
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1Exemplary measurement of the insertion depth angle. A reconstructed 2D “cochlear view” from the postoperative rotational tomography of patient 16. Insertion depth angle is 333°.
Figure 2Exemplary measurement of the cochlear size. A reconstructed 2D “cochlear view” from the postoperative rotational tomography of patient 16. Cochlear size is 9 mm.
Overview of the correlation between the NRT-ratio, the insertion depth angle, and the cochlear size.
| Patient | IDA in ° | CS in mm | NRT-ratio | NRT-ratio | NRT-ratio | NRT-ratio | NRT-ratio | NRT-ratio | NRT-ratio | NRT-ratio |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 328 | 7.6 | 0.96 | 0.87 | 0.91 | 0.87 | 0.91 | 0.93 | 0.97 | 1.05 |
| 2 | 333 | 8.2 | 0.84 | 0.81 | 0.84 | 0.77 | 0.76 | 0.79 | 0.82 | 0.83 |
| 3 | 343 | 7.8 | 1.04 | 0.95 | 0.97 | 0.87 | 0.84 | 0.88 | 0.9 | 0.88 |
| 4 | 373 | 8 | 0.91 | 1.08 | 1.11 | 1.02 | 1 | 1.07 | 1.1 | 1.1 |
| 5 | 326 | 8 | 0.93 | 0.91 | 0.92 | 0.89 | 0.88 | 0.89 | 0.9 | 0.91 |
| 6 | 332 | 8.3 | 0.75 | 0.7 | 0.73 | 0.71 | 0.74 | 0.78 | 0.83 | 0.87 |
| 7 | 392 | 7.9 | 1 | 1.06 | 1.03 | 1.09 | 1.08 | 1.09 | 1.06 | 0.99 |
| 8 | 367 | 7.3 | 0.89 | 0.84 | 0.89 | 0.8 | 0.81 | 0.87 | 0.93 | 0.95 |
| 9 | 400 | 8 | 0.95 | 0.94 | 0.96 | 0.88 | 0.86 | 0.91 | 0.93 | 0.93 |
| 10 | 353 | 7.2 | 0.91 | 0.87 | 0.91 | 0.8 | 0.76 | 0.81 | 0.85 | 0.86 |
| 11 | 392 | 7.8 | 0.84 | 0.85 | 0.85 | 0.82 | 0.79 | 0.81 | 0.81 | 0.79 |
| 12 | 373 | 7.2 | 0.77 | 1 | 0.98 | 0.92 | 0.86 | 0.89 | 0.87 | 0.84 |
| 13 | 368 | 7.6 | 0.88 | 0.83 | 0.87 | 0.79 | 0.77 | 0.79 | 0.84 | 0.85 |
| 14 | 332 | 7.6 | 0.98 | 0.99 | 1.01 | 1 | 1 | 1.01 | 1.03 | 1 |
| 15 | 370 | 8.8 | 0.87 | 0.87 | 0.89 | 0.82 | 0.81 | 0.82 | 0.84 | 0.83 |
| 16 | 333 | 9.0 | 0.91 | 0.84 | 0.88 | 0.81 | 0.83 | 0.88 | 0.92 | 0.95 |
| 17 | 381 | 7.4 | 0.86 | 0.88 | 0.88 | 0.82 | 0.81 | 0.84 | 0.84 | 0.82 |
| 18 | 356 | 8.4 | 0.8 | 0.73 | 0.77 | 0.64 | 0.58 | 0.6 | 0.63 | 0.66 |
| 19 | 356 | 8.4 | 1.04 | 1 | 1.04 | 0.97 | 0.95 | 0.97 | 1 | 1.01 |
| 20 | 358 | 8.2 | 0.91 | 0.98 | 0.97 | 0.91 | 0.93 | 0.98 | 0.97 | 0.91 |
| 21 | 332 | 7.8 | 0.74 | 0.7 | 0.74 | 0.68 | 0.73 | 0.78 | 0.82 | 0.83 |
| 22 | 320 | 8.7 | 0.95 | 0.89 | 0.93 | 0.8 | 0.77 | 0.82 | 0.86 | 0.88 |
| 23 | 380 | 7.4 | 0.98 | 0.96 | 0.99 | 0.88 | 0.88 | 0.97 | 1 | 0.99 |
| 24 | 388 | 7.5 | 0.92 | 0.92 | 0.96 | 0.82 | 0.79 | 0.87 | 0.9 | 0.87 |
| 25 | 382 | 7 | 0.79 | 0.7 | 0.75 | 0.67 | 0.71 | 0.75 | 0.8 | 0.85 |
| 26 | 351 | 7.4 | 0.9 | 1 | 0.96 | 1.01 | 1.05 | 1.08 | 1.03 | 0.97 |
|
| ||||||||||
|
| — | — | 0.056 | 0.286 | 0.218 | 0.201 | 0.113 | 0.196 | 0.111 | −0.024 |
|
| — | — | >0.05 | > 0.05 | >0.05 | >0.05 | >0.05 | >0.05 | >0.05 | >0.05 |
|
| ||||||||||
|
| — | — | −0.023 | −0.311 | −0.103 | −0.208 | −0.217 | −0.224 | −0.273 | −0.214 |
|
| — | — | >0.05 | >0.05 | >0.05 | >0.05 | >0.05 | >0.05 | >0.05 | >0.05 |
IDA: insertion depth angle; CS: cochlear size; NRT: neural response telemetry.
p values were calculated using Pearson's product-moment correlation.