Liege Franzini Tanamati1, Maria Valéria Schmidt Goffi-Gomez2, Lilian Ferreira Muniz3, Paola Angélica Samuel4, Gislaine Richter Minhoto Wiemes5, Daniele Penna Lima6, Sílvia Badur Curi7, Lucia Cristina Onuki7, Carla Fortunato Queiroz5, Ana Karla Bigois Capistrano6, Adriane Lima Mortari Moret1, Márcia Yuri Tsumura Kimura8, Valeria Oyanguren9, Herbert Mauch9.
Abstract
INTRODUCTION: The conventional evaluation of neural telemetry and impedance requires the use of the computer coupled to an interface, with software that provides visualization of the stimulus and response. Recently, a remote control (CR220®) was launched in the market, that allows the performance of intraoperative tests with minimal instrumentation.Entities:
Keywords: Cochlear implant; Evoked potential; Impedance; Impedâncias; Implante coclear; Potencial evocado; Telemetria; Telemetry
Mesh:
Year: 2018 PMID: 29858159 PMCID: PMC9443058 DOI: 10.1016/j.bjorl.2018.04.003
Source DB: PubMed Journal: Braz J Otorhinolaryngol ISSN: 1808-8686
Demographic data of the studied population with hearing loss etiology and model of the cochlear implant internal unit in the study population.
| Adults | Children | Total | |
|---|---|---|---|
| Idiopathic | 8 (9.5) | 22 (26.2) | 30 (35.7) |
| Progressive idiopathic | 15 (17.8) | 2 (2.4) | 17 (20.2) |
| Meningitis | 3 (3.6) | 3 (3.6) | 6 (7.1) |
| Genetic | 0 | 4 (4.8) | 4 (4.8) |
| Cytomegalovirus | 1 (1.2) | 3 (3.6) | 4 (4.8) |
| Syndromes | 3 (3.6) | 1 (1.2) | 4 (4.8) |
| Congenital | 0 | 3 (3.6) | 3 (3.6) |
| Rubella | 2 (2.4) | 1 (1.2) | 3 (3.6) |
| VIII nerve hypoplasia | 0 | 3 (3.6) | 3 (3.6) |
| Otosclerosis | 2 (2.4) | 0 | 2 (2.4) |
| Hypoxia | 0 | 2 (2.4) | 2 (2.4) |
| Preterm birth | 0 | 2 (2.4) | 2 (2.4) |
| Traumatic | 1 (1.2) | 0 | 1 (1.2) |
| 0 | 1 (1.2) | 1 (1.2) | |
| Toxoplasmosis | 0 | 1 (1.2) | 1 (1.2) |
| Mumps | 1 (1.2) | 0 | 1 (1.2) |
| CI 24RE CA | 17 (20.2) | 22 (26.2) | 39 (46.4) |
| CI 422 | 19 (22.6) | 26 (30.9) | 45 (53.6) |
| Total | 36 (42.8) | 48 (57.1) | 84 (100) |
Figure 1(A) Remote control unit called CR220®. (B) Set constituted by the software and programming interface was carried out by cable.
Figure 2Information available for the remote assistant after evaluating the impedance and information on the electrode status.
Description of the electrode status, represented by the impedance values recorded by Custom Sound® (standard procedure) and by the CR220® (remote assistant).
| Custom Sound® | CR220® | Total | |
|---|---|---|---|
| 1837 (99.40%) | 1837 (99.40%) | 3674 (99.40%) | |
| 7 (0.19%) | |||
| Coincident | 3 (0.16%) | 3 (0.16%) | |
| Discordant | 1 (0.05%) | 0 (0%) | |
| 9 (0.24%) | |||
| Coincident | 4 (0.22%) | 4 (0.22%) | |
| Discordant | 0 (0%) | 1 (0.05%) | |
| 3 (0.16%) | 3 (0.16%) | 6 (0.16%) | |
| 1848 (100%) | 1848 (100%) | 3696 (100%) | |
SC, short circuit.
Comparison of the impedance values (kOhms) recorded by Custom Sound® (standard procedure) and CR220® (remote assistant).
| Custom Sound® | CR220® | ||
|---|---|---|---|
| E1 | 8.70 (3.73) | 8.34 (3.69) | 0.562 |
| E6 | 8.64 (3.88) | 8.32 (3.75) | 0.590 |
| E11 | 8.81 (3.80) | 8.56 (3.54) | 0.695 |
| E16 | 9.36 (3.88) | 9.23 (3.46) | 0.864 |
| E22 | 12.38 (4.76) | 11.67 (4.05) | 0.336 |
SD, standard deviation.
p > 0.05 (paired t-test).
Comparison of the presence of neural response recorded by Custom Sound® (standard procedure) and by CR220® (remote assistant).
| Custom Sound® | CR220® | |
|---|---|---|
| Presence | 361 (88.69) | 351 (86.24) |
| Absence | 46 (11.31) | 56 (13.76) |
Comparison of the neural response thresholds (in current units – cu) recorded by Custom Sound® (standard procedure) and by CR220® (remote assistant).
| Custom Sound® | CR220® | ||
|---|---|---|---|
| E1 | 196.01 (21.88) | 193.30 (19.54) | 0.028 |
| E6 | 196.68 (17.67) | 194.99 (16.88) | 0.016 |
| E11 | 192.79 (14.01) | 193.38 (14.68) | 0.235 |
| E16 | 184.65 (19.57) | 184.34 (18.18) | 0.249 |
| E22 | 182.19 (20.74) | 179.21 (15.39) | 0.191 |
SD, standard deviation.
Figure 3Correlation between ECAP thresholds measured through CR220® and Custom Sound® EP on all collected electrodes.
Mean time (in minutes) to perform the procedures through CR220® and CS EP® for the two electrode models (CI 422 and CI 24RE CA).
| CR220® | CS EP® | |
|---|---|---|
| Mean (SD) (in minutes) | ||
| CI 422 | 5.20 ± 1.39 | 7.75 ± 4.00 |
| CI 24RE CA | 5.68 ± 3.00 | 6.98 ± 2.65 |
| Total | 5.42 ± 2.28 | 7.39 ± 3.44 |
SD, standard deviation.
p < 0.001 – statistically significant (paired t test).