Depth and quality of electrode insertion: a radiologic and pitch scaling assessment of two cochlear implant systems.

J M Chen1, R Farb, L Hanusaik, D Shipp, J M Nedzelski.   

Abstract

OBJECTIVE: To compare the depth of electrode insertion in two types of cochlear implants, and to assess the ability of the implantees in each group to place-pitch during random electrical stimulation. STUDY
DESIGN: This was a prospective clinical study.
SETTING: This study was performed at an implant program within a university teaching hospital. PATIENTS: Five consecutive patients with the Clarion (Advanced Bionics, Symlar, CA, U.S.A.) device and 5 with the Nucleus-22 (Cochlear Corporation, Sydney, Australia) implants were enrolled. All 10 implantees had fully active and functioning electrodes.
INTERVENTIONS: The depth of insertion was determined using plain anteroposterior skull film and high resolution computed tomography (CT). The quality of electrode insertion was assessed by pitch scaling; electrodes were randomly stimulated to generate subjective pitch responses. OUTCOME MEASURES: The depth of electrode insertion was measured radiographically as degrees of angular rotation within the cochlea. For pitch scaling, the averaged responses to electrical stimulation was plotted against the "place" of the electrodes along the array. Pitch range, plateauing, and reversal of pitches were also noted. Insertion depth was correlated with the result of pitch scaling and open-set speech discrimination at 3 months.
RESULTS: The mean insertion depth was 406 degrees for the Clarion device and 254 degrees for the Nucleus device. CT was used to confirm the intracochlear placement of the electrodes and their relationships to the cochleostomy site. It did not confer more information than the plain films unless kinking had occurred. Pitch perception was consistent with the tonotopic organization of the cochlea. The Nucleus-22 recipients displayed a broader range of pitches with less plateaus and reversals than the Clarion implants. The depth of insertion did not compare well with the outcome of pitch scaling or with open-set speech discrimination scores in either group of implantees.
CONCLUSION: The preformed spiral array of the Clarion device allowed deeper electrode insertion compared to the Nucleus-22 device. However, depth of insertion did not translate into better pitch placement.

Entities:  

Mesh:

Year:  1999        PMID: 10100522

Source DB:  PubMed          Journal:  Am J Otol        ISSN: 0192-9763


  7 in total

1.  Electrode insertion depth in cochlear implantees estimated during surgery, on plain film radiographs and with electrode function testing.

Authors:  Robert H R Bettman; Adriaan F van Olphen; Frans W Zonneveld; Egbert H Huizing
Journal:  Eur Arch Otorhinolaryngol       Date:  2003-05-23       Impact factor: 2.503

2.  Validating a New Tablet-based Tool in the Determination of Cochlear Implant Angular Insertion Depth.

Authors:  Michael W Canfarotta; Margaret T Dillon; Emily Buss; Harold C Pillsbury; Kevin D Brown; Brendan P O'Connell
Journal:  Otol Neurotol       Date:  2019-09       Impact factor: 2.311

3.  Measurement of Cochlear Implant Electrode Position From Intraoperative Post-insertion Skull Radiographs: A Validation Study.

Authors:  Maja Svrakic; David R Friedmann; Phillip M Berman; Adam J Davis; J Thomas Roland; Mario A Svirsky
Journal:  Otol Neurotol       Date:  2015-09       Impact factor: 2.311

4.  Initial Operative Experience and Short-term Hearing Preservation Results With a Mid-scala Cochlear Implant Electrode Array.

Authors:  Maja Svrakic; J Thomas Roland; Sean O McMenomey; Mario A Svirsky
Journal:  Otol Neurotol       Date:  2016-12       Impact factor: 2.311

5.  Evaluation of 4 multisection CT systems in postoperative imaging of a cochlear implant: a human cadaver and phantom study.

Authors:  B M Verbist; R M S Joemai; W M Teeuwisse; W J H Veldkamp; J Geleijns; J H M Frijns
Journal:  AJNR Am J Neuroradiol       Date:  2008-05-15       Impact factor: 3.825

6.  Angular Electrode Insertion Depth and Speech Perception in Adults With a Cochlear Implant: A Systematic Review.

Authors:  Floris Heutink; Simone R de Rijk; Berit M Verbist; Wendy J Huinck; Emmanuel A M Mylanus
Journal:  Otol Neurotol       Date:  2019-08       Impact factor: 2.311

7.  Evaluation of the Relationship between the NRT-Ratio, Cochlear Anatomy, and Insertions Depth of Perimodiolar Cochlear Implant Electrodes.

Authors:  Philipp Mittmann; Grit Rademacher; Sven Mutze; Frederike Hassepass; Arneborg Ernst; Ingo Todt
Journal:  Biomed Res Int       Date:  2015-12-29       Impact factor: 3.411

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.