Literature DB >> 26835503

Accounting for anatomical noise in search-capable model observers for planar nuclear imaging.

Anando Sen1, Howard C Gifford1.   

Abstract

Model observers intended to predict the diagnostic performance of human observers should account for the effects of both quantum and anatomical noise. We compared the abilities of several visual-search (VS) and scanning Hotelling-type models to account for anatomical noise in a localization receiver operating characteristic (LROC) study involving simulated nuclear medicine images. Our VS observer invoked a two-stage process of search and analysis. The images featured lesions in the prostate and pelvic lymph nodes. Lesion contrast and the geometric resolution and sensitivity of the imaging collimator were the study variables. A set of anthropomorphic mathematical phantoms was imaged with an analytic projector based on eight parallel-hole collimators with different sensitivity and resolution properties. The LROC study was conducted with human observers and the channelized nonprewhitening, channelized Hotelling (CH) and VS model observers. The CH observer was applied in a "background-known-statistically" protocol while the VS observer performed a quasi-background-known-exactly task. Both of these models were applied with and without internal noise in the decision variables. A perceptual search threshold was also tested with the VS observer. The model observers without inefficiencies failed to mimic the average performance trend for the humans. The CH and VS observers with internal noise matched the humans primarily at low collimator sensitivities. With both internal noise and the search threshold, the VS observer attained quantitative agreement with the human observers. Computational efficiency is an important advantage of the VS observer.

Entities:  

Keywords:  collimators; image quality; internal noise; model observers; nuclear medicine; planar imaging; scanning observers; visual search

Year:  2016        PMID: 26835503      PMCID: PMC4726820          DOI: 10.1117/1.JMI.3.1.015502

Source DB:  PubMed          Journal:  J Med Imaging (Bellingham)        ISSN: 2329-4302


  17 in total

1.  Effect of filtering on the detection and localization of small Ga-67 lesions in thoracic single photon emission computed tomography images.

Authors:  R G Wells; P H Simkin; P F Judy; M A King; H Pretorius; H C Gifford
Journal:  Med Phys       Date:  1999-07       Impact factor: 4.071

2.  Characterizing human perceptual inefficiencies with equivalent internal noise.

Authors:  Z L Lu; B A Dosher
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  1999-03       Impact factor: 2.129

3.  Recent advances in SPECT imaging.

Authors:  Mark T Madsen
Journal:  J Nucl Med       Date:  2007-04       Impact factor: 10.057

4.  Evaluation of internal noise methods for Hotelling observer models.

Authors:  Yani Zhang; Binh T Pham; Miguel P Eckstein
Journal:  Med Phys       Date:  2007-08       Impact factor: 4.071

5.  Assessment of cardiac single-photon emission computed tomography performance using a scanning linear observer.

Authors:  Chih-Jie Lee; Matthew A Kupinski; Lana Volokh
Journal:  Med Phys       Date:  2013-01       Impact factor: 4.071

6.  The meaning and use of the area under a receiver operating characteristic (ROC) curve.

Authors:  J A Hanley; B J McNeil
Journal:  Radiology       Date:  1982-04       Impact factor: 11.105

7.  Human- and model-observer performance in ramp-spectrum noise: effects of regularization and object variability.

Authors:  C K Abbey; H H Barrett
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2001-03       Impact factor: 2.129

8.  A random-sum Wilcoxon statistic and its application to analysis of ROC and LROC data.

Authors:  Liansheng Larry Tang; N Balakrishnan
Journal:  J Stat Plan Inference       Date:  2010-06-12       Impact factor: 1.111

9.  Toward realistic and practical ideal observer (IO) estimation for the optimization of medical imaging systems.

Authors:  Xin He; Brian S Caffo; Eric C Frey
Journal:  IEEE Trans Med Imaging       Date:  2008-10       Impact factor: 10.048

10.  Efficient visual-search model observers for PET.

Authors:  H C Gifford
Journal:  Br J Radiol       Date:  2014-05-16       Impact factor: 3.039

View more
  7 in total

1.  Foveated Model Observers to predict human performance in 3D images.

Authors:  Miguel A Lago; Craig K Abbey; Miguel P Eckstein
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2017-03-10

2.  Investigation of Exposure Factors for Various Breast Composition and Thicknesses in Digital Screening Mammography Related to Breast Dose.

Authors:  Khaled Alkhalifah; Ajit Brindhaban
Journal:  Med Princ Pract       Date:  2018-03-07       Impact factor: 1.927

3.  Assessing CT acquisition parameters with visual-search model observers.

Authors:  Zohreh Karbaschi; Howard C Gifford
Journal:  J Med Imaging (Bellingham)       Date:  2018-04-05

4.  Medical image quality metrics for foveated model observers.

Authors:  Miguel A Lago; Craig K Abbey; Miguel P Eckstein
Journal:  J Med Imaging (Bellingham)       Date:  2021-08-16

5.  Visual-search observers for assessing tomographic x-ray image quality.

Authors:  Howard C Gifford; Zhihua Liang; Mini Das
Journal:  Med Phys       Date:  2016-03       Impact factor: 4.071

6.  Foveated Model Observers for Visual Search in 3D Medical Images.

Authors:  Miguel A Lago; Craig K Abbey; Miguel P Eckstein
Journal:  IEEE Trans Med Imaging       Date:  2021-03-02       Impact factor: 10.048

Review 7.  Virtual clinical trials in medical imaging: a review.

Authors:  Ehsan Abadi; William P Segars; Benjamin M W Tsui; Paul E Kinahan; Nick Bottenus; Alejandro F Frangi; Andrew Maidment; Joseph Lo; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2020-04-11
  7 in total