Rapid and accurate diagnosis of influenza is important for patient management and infection control. We determined the performance of the cobas(®) Influenza A/B assay, a rapid automated nucleic acid assay performed on the cobas(®) Liat System for qualitative detection of influenza A and influenza B from nasopharyngeal (NP) swab specimens. Retrospective frozen and prospectively collected NP swabs from patients with signs and symptoms of influenza collected in universal transport medium (UTM) were tested at multiple sites including CLIA-waived sites using the cobas® Influenza A/B assay. Results were compared to the Prodesse Pro-Flu+ assay and to viral culture. Compared to the Prodesse ProFlu+ Assay, sensitivities of the cobas(®) Influenza A/B assay for influenza A and B were 97.7 and 98.6%, respectively; specificity was 99.2 and 99.4%. Compared to viral culture, the cobas(®) Influenza A/B assay showed sensitivities of 97.5 and 96.9% for influenza virus A and B, respectively; specificities were 97.9% for both viruses. Polymerase chain reaction (PCR)/sequencing showed that the majority of viral culture negative but cobas(®) Influenza A/B positive results were true positive results, indicating that the cobas(®) Influenza A/B assay has higher sensitivity compared to viral culture. In conclusion, the excellent accuracy, rapid time to result, and remarkable ease of use make the cobas(®) Influenza A/B nucleic acid assay for use on the cobas(®) Liat System a highly suitable point-of-care solution for the management of patients with suspected influenza A and B infection.Entities:
Keywords: PCR; diagnosis; influenza; influenza A; influenza B; point of care; sensitivity; viral culture
Year: 2015 PMID: 26716012 PMCID: PMC4681351 DOI: 10.1556/1886.2015.00046
Source DB: PubMed Journal: Eur J Microbiol Immunol (Bp) ISSN: 2062-509X
Clinical performance of the cobas® Influenza A/B assay compared to the Prodesse ProFlu+ assay in retrospective samples for the detection of influenza A
| Influenza A | ProFlu+ | ||||||
|---|---|---|---|---|---|---|---|
| Positive | Negative | Total | Agreement (%) | 95% CI | |||
| cobas Liat | Positive | 79 | 0 | 79 | Positive | 98.8 | 93.3–99.8 |
| Negative | 1 | 220 | 221 | Negative | 100.0 | 98.3–100.0 | |
| Total | 80 | 220 | 300 | ||||
* One Liat negative, ProFlu+ positive specimen was negative by PCR/sequencing
Clinical performance of the cobas® Influenza A/B assay compared to the Prodesse ProFlu+ assay in retrospective samples for the detection of influenza B
| Influenza B | ProFlu+ | ||||||
|---|---|---|---|---|---|---|---|
| Positive | Negative | Total | Agreement (%) | 95% CI | |||
| cobas Liat | Positive | 100 | 0 | 100 | Positive | 100.0 | 96.3–100.0 |
| Negative | 0 | 200 | 200 | Negative | 100.0 | 98.1–100.0 | |
| Total | 100 | 200 | 300 | ||||
Comparison of the cobas® Influenza A/B assay with Prodesse ProFlu+ Assay on prospective samples for the detection of influenza A
| Influenza A | ProFlu+ | ||||||
|---|---|---|---|---|---|---|---|
| Positive | Negative | Total | Accuracy (%) | 95% CI | |||
| cobas Liat | Positive | 94 | 8 | 102 | Sensitivity | 96.9 | 91.3–98.9 |
| Negative | 3[ | 736 | 739 | Specificity | 98.9 | 97.9–99.5 | |
| Total | 97 | 744 | 841[ | ||||
* Of eight cobas Liat positive, ProFlu+ negative specimen, three were positive and five were negative by PCR/sequencing. One was positive, and seven were negative by culture
† Of three cobas Liat negative, ProFlu+ positive specimens, three were positive by PCR/sequencing. One was positive, one was negative, and one was not tested by culture
‡ One specimen was indeterminate for Inf A by ProFlu+ due to late Ct. This sample was positive for influenza A by cobas Liat and PCR/sequencing
Comparison of the cobas® Influenza A/B assay with Prodesse ProFlu+ Assay on prospective samples for the detection of influenza B
| Influenza B | ProFlu+ | ||||||
|---|---|---|---|---|---|---|---|
| Positive | Negative | Total | Accuracy (%) | 95% CI | |||
| cobas Liat | Positive | 42 | 6 | 48 | Sensitivity | 95.5 | 84.9–98.7 |
| Negative | 72[ | 792 | 794 | Specificity | 99.2 | 98.4–99.7 | |
| Total | 44 | 798 | 842 | ||||
* Of six cobas Liat positive, ProFlu+ negative specimens, four were positive and two were negative by PCR/sequencing. All six were negative by culture
† Of two cobas Liat negative, ProFlu+ positive specimens, one was positive and one was negative by PCR/sequencing. One was negative, and one was not tested by culture
Comparison of the cobas® Influenza A/B assay with Prodesse ProFlu+ Assay on retrospective and prospective samples for the detection of influenza A
| Influenza A | Viral culture | ||||||
|---|---|---|---|---|---|---|---|
| Positive | Negative | Total | Accuracy (%) | 95% CI | |||
| cobas Liat | Positive | 173 | 8 | 181 | Sensitivity | 97.7 | 94.3–99.1 |
| Negative | 4 | 956 | 960 | Specificity | 99.2 | 98.4–99.6 | |
| Total | 177 | 964 | 1141 | ||||
* Of eight cobas Liat positive, ProFlu+ negative specimens, three were positive and five were negative by PCR/sequencing. One was positive, and seven were negative by viral culture
† Of four cobas Liat negative, ProFlu+ positive specimens, three were positive and one was negative by PCR/sequencing. One was positive, one was negative, and two were not tested by viral culture
‡ One specimen was indeterminate for Inf A by ProFlu+ due to late Ct. This sample was positive for Influenza A by cobas Liat and PCR/sequencing
Comparison of the cobas® Influenza A/B assay with Prodesse ProFlu+ Assay on retrospective and prospective samples for the detection of influenza B
| Influenza B | Viral culture | ||||||
|---|---|---|---|---|---|---|---|
| Positive | Negative | Total | Accuracy (%) | 95% CI | |||
| cobas Liat | Positive | 142 | 6 | 148 | Sensitivity | 98.6 | 95.1–99.6 |
| Negative | 2 | 992 | 994 | Specificity | 99.4 | 98.7–99.7 | |
| Total | 144 | 998 | 1142 | ||||
* Of six cobas positive, ProFlu+ negative specimens, four were positive and two were negative by PCR/sequencing. All six were negative by viral culture
† Of two cobas negative, ProFlu+ positive specimens, one was positive and one was negative by PCR/sequencing. One was negative, and one was not tested by viral culture
Clinical performance of the cobas® Influenza A/B assay compared to viral culture on prospective samples for the detection of influenza A
| Influenza A | Viral culture | ||||||
|---|---|---|---|---|---|---|---|
| Positive | Negative | Total | Accuracy (%) | 95% CI | |||
| cobas Liat | Positive | 77 | 15 | 92 | Sensitivity | 97.5 | 91.2–99.3 |
| Negative | 2 | 690 | 692 | Specificity | 97.9 | 96.5–98.7 | |
| Total | 79 | 705 | 784 | ||||
* Of 15 cobas Liat positive, culture negative specimens, nine were positive and six were negative by PCR/sequencing. Seven were positive, one was indeterminate due to high Ct, and seven were negative by PoFlu+
† Of two cobas Liat negative, culture positive specimens, two were positive by PCR/sequencing. One was positive, and one was negative by ProFlu+
Clinical performance of the cobas® Influenza A/B assay compared to viral culture on prospective samples for the detection of influenza B
| Influenza B | Viral culture | ||||||
|---|---|---|---|---|---|---|---|
| Positive | Negative | Total | Accuracy (%) | 95% CI | |||
| cobas Liat | Positive | 31 | 16 | 47 | Sensitivity | 96.9 | 84.3–99.4 |
| Negative | 1[ | 736 | 737 | Specificity | 97.9 | 96.6–98.7 | |
| Total | 32 | 752 | 784 | ||||
* Of 16 cobas Liat positive, culture negative specimens, 14 were positive and two were negative by PCR/sequencing. Ten were positive, and six were negative by PoFlu+
† One cobas Liat negative, culture positive specimen was positive by PCR/sequencing and was negative by ProFlu+