Literature DB >> 17452000

Performance of six influenza rapid tests in detecting human influenza in clinical specimens.

Aeron C Hurt1, Robert Alexander, Jan Hibbert, Nicola Deed, Ian G Barr.   

Abstract

BACKGROUND: The rapid diagnosis of influenza can alter the management of a patient's illness, resulting in reduced antibiotic usage, correct use of influenza antivirals and reduced length of stay in hospital emergency departments. The rapid tests have also been used to detect outbreaks in institutions and may play a role in pandemic influenza control.
OBJECTIVES: To test six different rapid influenza tests, in a head-to-head comparison for the detection of seasonal influenza types A and B, compared to laboratory-based tests. STUDY
DESIGN: One hundred and seventy-seven clinical specimens taken from mostly paediatric patients between June and October 2006 were tested using six influenza diagnostic tests and three laboratory-based techniques (immunofluorescence, cell culture and real-time RT-PCR). RESULTS AND
CONCLUSION: Compared with cell culture, five of the rapid tests (Binax Now Influenza A&B, Directigen EZ Flu A+B, Denka Seiken Quick Ex-Flu, Fujirebio Espline Influenza A&B-N, and Quidel QuickVue Influenza A+B Test) demonstrated a similar influenza A sensitivity of between 67-71% and a specificity of 99-100%, however one rapid test (Rockeby Influenza A Antigen Test) had a significantly lower influenza A sensitivity of only 10% (specificity was 100%). For the five kits that detected influenza B antigen, sensitivity was considerably lower than that seen for influenza A (sensitivity for all the kits was 30%), although the number of specimens containing influenza B viruses was low.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17452000     DOI: 10.1016/j.jcv.2007.03.002

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


  68 in total

1.  Perspective on diagnostics for global health.

Authors:  Elain Fu; Paul Yager; Pierre N Floriano; Nicolaos Christodoulides; John T McDevitt
Journal:  IEEE Pulse       Date:  2011-11       Impact factor: 0.924

2.  Enhanced sensitivity of lateral flow tests using a two-dimensional paper network format.

Authors:  Elain Fu; Tinny Liang; Jared Houghtaling; Sujatha Ramachandran; Stephen A Ramsey; Barry Lutz; Paul Yager
Journal:  Anal Chem       Date:  2011-09-21       Impact factor: 6.986

3.  Evaluation of the ESPLINE® Influenza A & B-N assay for the detection of influenza A and B in nasopharyngeal aspirates.

Authors:  E De Witte; H Goossens; M Ieven
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2011-09-28       Impact factor: 3.267

Review 4.  The role of rapid antigen testing for influenza in the era of molecular diagnostics.

Authors:  Suzanne E Dale
Journal:  Mol Diagn Ther       Date:  2010-08-01       Impact factor: 4.074

Review 5.  Influenza.

Authors:  John H Beigel
Journal:  Crit Care Med       Date:  2008-09       Impact factor: 7.598

6.  Comparison of the BD Veritor System for Flu A+B with the Alere BinaxNOW influenza A&B card for detection of influenza A and B viruses in respiratory specimens from pediatric patients.

Authors:  Ferdaus Hassan; Ashley Nguyen; Ashley Formanek; James J Bell; Rangaraj Selvarangan
Journal:  J Clin Microbiol       Date:  2014-01-03       Impact factor: 5.948

7.  Rapid Molecular Detection and Differentiation of Influenza Viruses A and B.

Authors:  Caitlin C Otto; Samuel E Kaplan; Jeffrey Stiles; Albina Mikhlina; Cindy Lee; N Esther Babady; Yi-Wei Tang
Journal:  J Vis Exp       Date:  2017-01-30       Impact factor: 1.355

8.  Comparison of machine learning classifiers for influenza detection from emergency department free-text reports.

Authors:  Arturo López Pineda; Ye Ye; Shyam Visweswaran; Gregory F Cooper; Michael M Wagner; Fuchiang Rich Tsui
Journal:  J Biomed Inform       Date:  2015-09-16       Impact factor: 6.317

9.  Comparison of the Simplexa FluA/B & RSV direct assay and laboratory-developed real-time PCR assays for detection of respiratory virus.

Authors:  Mitchell W Woodberry; Rohit Shankar; Anne Cent; Keith R Jerome; Jane Kuypers
Journal:  J Clin Microbiol       Date:  2013-09-18       Impact factor: 5.948

10.  Comparison of Xpert Flu rapid nucleic acid testing with rapid antigen testing for the diagnosis of influenza A and B.

Authors:  Michael A DiMaio; Malaya K Sahoo; Jesse Waggoner; Benjamin A Pinsky
Journal:  J Virol Methods       Date:  2012-07-25       Impact factor: 2.014

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.