| Literature DB >> 27582513 |
Frederick S Nolte1, Lori Gauld2, Susan B Barrett2.
Abstract
We compared two rapid, point-of care nucleic acid amplification tests for detection of influenza A and B viruses (Alere i [Alere] and cobas Liat [Roche Diagnostics]) with the influenza A and B virus test components of the FilmArray respiratory panel (BioFire Diagnostics) using 129 respiratory specimens collected in universal viral transport medium (80 influenza A virus and 16 influenza B virus positive) from both adult and pediatric patients. The sensitivities of the Alere test were 71.3% for influenza A virus and 93.3% for influenza B virus, with specificities of 100% for both viruses. The sensitivities and specificities of the Liat test were 100% for both influenza A and B viruses. The poor sensitivity of the Alere test for detection of influenza A virus was likely due to a study set that included many low-positive samples that were below its limit of detection.Entities:
Mesh:
Year: 2016 PMID: 27582513 PMCID: PMC5078555 DOI: 10.1128/JCM.01586-16
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948