We compared an FDA-cleared rapid (<20 min) PCR assay (Cobas Liat; Roche Diagnostics) to our routine influenza A and B real-time PCR assay (Simplexa Flu A/B & RSV Direct; Focus Diagnostics) using respiratory swabs (n = 197). The Cobas Liat influenza A and B assays demonstrated sensitivities of 99.2% (123/124) and 100% (23/23), respectively, while showing a specificity of 100% for each target.Mesh:
Substances:
Year: 2015 PMID: 25926484 PMCID: PMC4473187 DOI: 10.1128/JCM.00791-15