Abstract
Data on the performance of rapid molecular point-of-care use platforms for diagnosis of influenza are lacking. We validated nasopharyngeal (NP) flocked specimens in universal transport medium (UTM) and evaluated the clinical sensitivity and specificity of the Alere i influenza A&B test compared to those of the Xpert flu A/B assay. The Alere i influenza A&B test had an overall sensitivity and specificity of 93.8% and 62.5% for influenza A, respectively, and of 91.8% and 53.6% for influenza B, respectively. The poor specificity was due to influenza virus samples determined positive for both type A and B.Entities:
Mesh:
Year: 2014 PMID: 25502527 PMCID: PMC4298524 DOI: 10.1128/JCM.02783-14