Gregory L Damhorst1, Carlos Duarte-Guevara2, Weili Chen2, Tanmay Ghonge1, Brian T Cunningham3, Rashid Bashir3. 1. Department of Bioengineering, The University of Illinois at Urbana-Champaign, Urbana, IL 61801, USA ; Micro and Nanotechnology Laboratory, The University of Illinois at Urbana-Champaign, Urbana, IL 61801, USA. 2. Micro and Nanotechnology Laboratory, The University of Illinois at Urbana-Champaign, Urbana, IL 61801, USA ; Department of Electrical and Computer Engineering, The University of Illinois at Urbana-Champaign, Urbana, IL 61801, USA. 3. Department of Bioengineering, The University of Illinois at Urbana-Champaign, Urbana, IL 61801, USA ; Micro and Nanotechnology Laboratory, The University of Illinois at Urbana-Champaign, Urbana, IL 61801, USA ; Department of Electrical and Computer Engineering, The University of Illinois at Urbana-Champaign, Urbana, IL 61801, USA.
Abstract
Viral load measurements are an essential tool for the long-term clinical care of hum an immunodeficiency virus (HIV)-positive individuals. The gold standards in viral load instrumentation, however, are still too limited by their size, cost, and sophisticated operation for these measurements to be ubiquitous in remote settings with poor healthcare infrastructure, including parts of the world that are disproportionately affected by HIV infection. The challenge of developing a point-of-care platform capable of making viral load more accessible has been frequently approached but no solution has yet emerged that meets the practical requirements of low cost, portability, and ease-of-use. In this paper, we perform reverse-transcription loop-mediated isothermal amplification (RT-LAMP) on minimally processed HIV-spiked whole blood samples with a microfluidic and silicon microchip platform, and perform fluorescence measurements with a consumer smartphone. Our integrated assay shows amplification from as few as three viruses in a ~ 60 nL RT-LAMP droplet, corresponding to a whole blood concentration of 670 viruses per µL of whole blood. The technology contains greater power in a digital RT-LAMP approach that could be scaled up for the determination of viral load from a finger prick of blood in the clinical care of HIV-positive individuals. We demonstrate that all aspects of this viral load approach, from a drop of blood to imaging the RT-LAMP reaction, are compatible with lab-on-a-chip components and mobile instrumentation.
Viral load measurements are an essential tool for the long-term clinical care of hum an immunodeficiency virus (HIV)-positive individuals. The gold standards in viral load instrumentation, however, are still too limited by their size, cost, and sophisticated operation for these measurements to be ubiquitous in remote settings with poor healthcare infrastructure, including parts of the world that are disproportionately affected by HIV infection. The challenge of developing a point-of-care platform capable of making viral load more accessible has been frequently approached but no solution has yet emerged that meets the practical requirements of low cost, portability, and ease-of-use. In this paper, we perform reverse-transcription loop-mediated isothermal amplification (RT-LAMP) on minimally processed HIV-spiked whole blood samples with a microfluidic and silicon microchip platform, and perform fluorescence measurements with a consumer smartphone. Our integrated assay shows amplification from as few as three viruses in a ~ 60 nL RT-LAMP droplet, corresponding to a whole blood concentration of 670 viruses per µL of whole blood. The technology contains greater power in a digital RT-LAMP approach that could be scaled up for the determination of viral load from a finger prick of blood in the clinical care of HIV-positive individuals. We demonstrate that all aspects of this viral load approach, from a drop of blood to imaging the RT-LAMP reaction, are compatible with lab-on-a-chip components and mobile instrumentation.
Human immunodeficiency virus (HIV) affects 36.9 million people worldwide
[1]. During the course of nearly four
decades since the emergence of HIV on a pandemic scale, advances in antiretroviral
therapy have transformed HIV infection from a death sentence into a chronic illness
that can have little impact on life expectancy for those in whom the infection is
properly managed [2]. At the population level,
rates of new infections, mother-to-child transmission, and deaths from HIV-related
causes are declining [2]. However, the lack of
availability of the appropriate diagnostic technologies essential to informing
treatment in routine HIV care is still among the chief barriers preventing access to
the standard of care for millions of HIV-positive individuals worldwide,
particularly in resource-limited settings.CD4+ T lymphocyte counts and blood plasma viral load are the two core
diagnostic measurements broadly considered essential to HIV care, as they both guide
the initiation of therapy and indicate the efficacy of each individual’s
treatment regimen [3]. CD4 counts,
traditionally performed by flow cytometry, are increasingly available in remote
settings due to the introduction of new portable platforms [4-7]. Viral load
platforms, on the other hand, are well behind CD4 technologies in penetrating the
developing world. Viral load instruments are traditionally reverse-transcription
polymerase chain reaction (RT-PCR), nucleic acid sequence-based amplification
(NASBA), or branched DNA (bDNA) assays; although these can be capable of detecting
fewer than ten viral RNA copies per mL of blood plasma, these instruments require a
laboratory setting, extensive sample handling, and sophisticated processing [6, 8–10].One promising solution that can help address this issue is loop-mediated
isothermal amplification (LAMP) [11]. LAMP
emerged in the early 2000s as an alternative to PCR for nucleic acid detection
[12]. LAMP is attractive for
point-of-care applications because, unlike PCR, it does not require temperature
cycling (isothermal at 60–65 °C), and because it is typically less
sensitive than PCR to amplification inhibitors. RT-LAMP assays for HIV were first
described soon after the introduction of the initial concept [13, 14], and there have
been several reports since then regarding variations of the assay, including efforts
toward point-of-care applications [15– 21]. Among the novel LAMP
approaches applied to HIV that are presented in these reports are a battery-powered
handheld microfluidic system that was demonstrated with purified DNA [20], a SlipChip device for digital LAMP [21], and an electricity-free heating container
that facilitates a qualitative RT-LAMP assay on minimally processed whole blood
[17, 18]. To date, however, an approach capable of performing a quantitative
RT-LAMP assay from a drop of whole blood on a platform compatible with a fully
automated, portable device has not been presented.Traditionally, it is believed that nucleic acid amplification requires
complete purification of the RNA or DNA target in order to be compatible with the
amplification reaction. The robustness of LAMP, however, has disrupted this
thinking. Whole blood treated only with a cell lysis buffer has been employed by
Curtis et al. for HIV LAMP in a qualitative measurement in an electricity-free
heating device [17]. We present here RT-LAMP
with minimally processed lysed whole blood for a quantitative measurement of HIV
viral load capable of detecting as few as three whole virus particles per ~
60 nL reaction droplet.Our approach to RT-LAMP HIV viral load measurements begins with a drop of
whole blood. The data presented here demonstrate the potential of this approach to
be developed into a fully automated mobile device that does not require manual
processing. First, we compare and contrast the performance of the RT-LAMP reaction
on a standard laboratory thermocycler, both with purified viral RNA in water and
with whole virus particles in whole blood that has only been treated with a cell
lysis buffer. Next, we implement a simple microfluidic mixing module in order to
show that the whole blood lysis step can be performed on a chip without the loss of
analyte or interference with the detection assay. We then move to a microchip
platform and characterize the RT-LAMP reaction with purified RNA and lysed whole
blood spiked with viral RNA, imaging with both a standard fluorescence microscope
and a consumer smartphone without hardware modifications in order to compare and
contrast standard and novel techniques. To demonstrate the robustness of the assay,
we show the compatibility of the on-chip reaction with the presence of viral RNA
from the hepatitis C virus (HCV) and viral DNA from the hepatitis B virus (HBV),
which do not cross-react. We then combine the microfluidic lysis, microchip reaction
platform, and smartphone imaging in order to demonstrate the capability of our
platform to quantitatively determine HIV viral load from a drop of blood. Finally,
we discuss merits and drawbacks, as well as the potential for this approach to
address the need for point-of-care viral load technology.
2 Materials and methods
2.1 Samples
Whole blood
Whole venous blood samples were drawn from HIV-negative donors with a
syringe and transferred to 4 mL BD Va cutainer K2 EDTA collection tubes.
Tubes were stored at room temperature on a sample rotisserie until used for
experiments.
Viruses
HIV-1, strain IIIB propagated in the H9 human T lymphocyte cell line
was purchased from Advanced Biotech nologies, Inc. Virus stock was provided
in purified form at a concentration of 6.7E+10 vp·
mL−1 (vp is short for virus particles) in storage
buffer containing 10 mmol· L−1 Tris, 150
mmol·L−1 NaCl, and 1
mmol·L−1 ethylene diamine tetraacetic acid
(EDTA) at a pH of 7.5. Viruses used for experiments in whole particle form
were diluted from aliquots of the stock, either in additional storage buffer
prepared in-house or in phosphate buffered saline (PBS) from Fisher
Scientific.
Viral nucleic acids
Synthetic HBV DNA (ATCC® VR-3232SD™) and synthetic
HCV RNA (ATCC® VR-3233SD™) were purchased from American Type
Culture Collection (ATCC). HIV-1 RNA was purified from HIV-1 IIIB whole
particles using the PureLink® Viral RNA/DNA Mini Kit from Life
Technologies. Two methods were used to produce dilutions of HIV-1 RNA for
thermocycler characterization of the RT-LAMP reaction with purified RNA. In
method 1, whole virus particles were diluted in PBS and each dilution was
separately purified in the PureLink® kit. In method 2, 10 µL
of 6.7E+10 vp· mL−1 was added to 190 µL
PBS to meet kit specifications, and was then purified and eluted in 150
µL RNase-free water for a final concentration of 4.47E+9
vp·mL−1 (or 8.93E+9 RNA copies per mL). This
purified RNA was then aliquoted and stored at −80 °C until
use.Microchip RT-LAMP experiments were performed with this viral RNA,
with the exception of the integrated experiment. The experiments performed
with purified RNA in water were done because purified RNA is the standard
analyte in reverse-transcription nucleic acid amplification assays and these
experiments serve as a basis for comparison to lysed whole blood. Whole
virus particles are the ideal analyte in whole blood; however, the
preliminary microchip experiments were performed with viral RNA spiked in
whole blood for biosafety reasons while the technique was being developed.
Preliminary “macroscale” amplification experiments with
whole blood in a thermocycler (not on a microchip) did include whole virus
particles because the technique was compatible with biosafety practices. To
perform the final, integrated on-chip experiment, an apparatus was
constructed in a biosafety cabinet in order to accommodate microchip
experiments with whole virus particles.
Blood cell lysis
The whole blood lysis buffer was based on the work by Curtis et al.
[17] and contained 2.5
mmol· L−1 KHCO3, 37.5
mmol·L−1 NH4Cl, and 0.025
mmol·L−1 EDTA. A 1:4 ratio of blood to lysis
buffer was used for all lysed blood experiments. In preliminary experiments,
blood and lysis buffer were metered and mixed with a manual pipettor, while
the final integrated experiment and the preliminary experiment to
characterize microfluidic mixing employed on-chip lysis in a microfluidic
channel. For microfluidic mixing, the ratio of volumes mixed was fixed to
1:4 by setting relative flow rates from two syringe pumps driving each
component.
2.2 RT-LAMP
Reaction components
The RT-LAMP assay was adapted from the work by Curtis et al. [17]. Reaction concentrations of buffers
were 1× isothermal amplification buffer, 1.4m mol·L
−1 deoxyribonucleoside triphosphates (dNTPs), and 10
mmol·L−1 MgSO4 from New England
Biolabs, and 0.4 mol ·L −1 betaine from
Sigma-Aldrich. In some cases, where indicated, 0.8
mol·L−1 betaine was used. These reaction
buffer components were prepared in appropriate ratios in bulk and stored at
−20 °C between experiments. Enzymes and DNA intercalating
dye were added separately to this buffer mix for a complete master mix that
was freshly prepared for each experiment. The RT-LAMP enzymes used in the
reaction were 0.64 U· µL
−1Bst 2.0 DNA polymerase and 0.08
U· µL−1 AMV reverse transcriptase fro m
New Engl and Biolabs. 1× EvaGreen from Biotium, a double-stranded
DNA (dsDNA) intercalating dye, was included in the reaction for the
detection of reaction products.
Primers
Six LAMP primers were based on the study by Curtis et al. [17], including a six-primer set
containing 0.2 µmol· L−1 each of F3
(5’-AGTTCCCTTAGATAAAGACTT-3’) and B3 ( 5’-
CCTACATACAAATCATCCATGT- 3’) primers, 1.6 µmol· L
−1 each of forward inner primer (FIP)
(5’-GTGGAAGCACATT
GTACTGATATCTTTTTGGAAGTATACTGCAT-TTACCAT-3’) and backward inner
primer (BIP)
(5’-GGAAAGGATCACCAGCAATATTCCTCTGGATTTTGTTTTCTAAAAGGC-3’),
and 0.8 µmol·L−1 each of LoopF
(5’-GGTGTCTCATTGTT TATACTA-3’) and LoopB
(5’-GCATGACA-AAAATCTTA GA-3’) primers.
Negative controls
All amplification experiments, whether in the thermocycler or on the
microchip, included negative controls that consisted either of water without
RNA or lysed blood without viruses/viral RNA, according to the nature of the
positive samples being tested. Amplification of the negative control within
the reaction timeframe was considered to be an indication of a contaminated
test. Fluorescence curves are not presented for these negative controls,
although they were included in every experiment.
Reaction platforms
RT-LAMP reactions were performed on two different platforms at
various stages of this study. For the purpose of establishing the RT-LAMP
reaction and comparing and contrasting purified RNA in water with lysed
whole blood, standard 25 µL reactions were performed in 0.2 mL
reaction tubes in an Eppendorf Mastercycler® ep realplex Real-Time
PCR System. The thermocycler was also used for RT-LAMP reactions
characterizing the microfluidic mixing module in order to eliminate possible
noise introduced by the microchip system.To develop the microchip amplification, several microchip
experiments were performed, beginning with RNA in water and RNA-spiked lysed
whole blood (RNA was used for bio-safety reasons in these preliminary
experiments as explained above). Each individual droplet (reaction) on the
microchip contained approximately 60 nL and the entire microchip was placed
in a copper bowl, as described below, and heated on an INSTEC STC200 heating
stage. Imaging was initially performed with a Nikon Eclipse FN1 fluorescence
microscope in order to employ a standard imaging method. Later, a Samsung
Galaxy Note 4 smartphone was introduced. Both the fluorescence microscope
and the smartphone were used in order to compare the imaging capabilities of
the smartphone with those of standard laboratory imaging equipment. The
reactions were initially incubated at 60 °C in the commercial
thermocycler and later at 65 °C for the on-chip experiments.
Fluorescence measurements were performed every 60 s with the thermocycler
and fluorescence microscope, but increased to every 30 s with the smartphone
platform.Data are presented in this paper for microchip reactions with
purified RNA in water imaged with a microscope and RNA-spiked lysed whole
blood imaged with a smartphone. The intermediate experiment, RNA-spiked
lysed whole blood reactions imaged with a microscope, is provided in the
Supplementary
Information.
2.3 Microfluidic lysis module
Fabrication
The microfluidic lysis module is based on an earlier design that was
reported previously [22]. The
polydimethylsiloxane (PDMS)-on-glass microfluidic channel was made from an
SU-8 master mold fabricated using standard clean-room photolithography
techniques. Uncured PDMS was poured over the SU-8 master, degassed in a
desiccator, and cured on a hot plate at 60 °C. Holes for tubing
connections were punched into the PDMS with a 1.5 mm biopsy punch, prior to
solvent degreasing and the oxygen plasma surface activation of both the PDMS
and a glass microscope slide in a Diener PICO plasma system. Activated
surfaces of the PDMS and glass were brought into contact following surface
activation and heated at 60–70 °C on a hot plate, producing
covalent bonds between the two pieces.
Fluidic apparatus
The microfluidic lysis experiments involving whole HIV particles
were performed with a fluidic apparatus that interfaced with the PDMS
microfluidic chips and consisted of syringe pumps and high-performance
liquid chromatography (HPLC) valves built within a biosafety cabinet.
Biosafety level 2+ protocols were followed.
2.4 On-chip RT-LAMP
Chip fabrication
Microchip RT-LAMP experiments employed a microfabricated silicon
substrate [23]. Briefly, a silicon
wafer was thermally oxidized to create a silicon oxide layer of ~
150 nm. The oxide was then patterned with photolithography and a
hydrofluoric acid etch step, exposing the silicon where the wells would be
etched. The wafer was then immersed in a heated tetramethylammonium
hydroxide (TMAH) bath for 18 h in order to anisotropically etch the silicon,
creating inverted square pyramids that would later be used as reaction
wells. An approximation of the dimensions of the inverted square pyramids is
provided in the Supplementary Information.
Chip preparation
Chips for all microchip RT-LAMP experiments were prepared in the
following manner: First, the microchip was cleaned in a piranha solution
containing 1:3 30% hydrogen peroxide and sulfuric acid for 10 min
and then it was rinsed in deionized water. Each chip was then degreased with
acetone, methanol, and isopropanol and dried by blowing with nitrogen gas.
To produce a hydrophobic surface to promote stability of droplets, the chip
was rinsed with Sigmacote® from Sigma-Aldrich by pipetting the
solution repeatedly over the surface of the chip. The chip was then rinsed
briefly with isopropanol, dried by blowing with nitrogen gas, and placed in
a copper bowl.
Microinjection
A Narishige IM-300 Microinjector with Eppendorf VacuTip
microinjection holding capillary (15 µm in inner diameter, 100
µm in outer diameter) was used both to spot primers and to place
reaction droplets. A 20 ms injection pulse was used, resulting in a droplet
of approximately 60 nL. The microinjection procedure was performed after
chip cleaning and preparation as follows: LAMP DNA primers in Tris-EDTA (TE)
buffer were diluted in water to the final reaction concentration. Droplets
were placed in all 36 wells of the microchip array using the microinjection
system and a 3D micromanipulator (MCL-D331) from World Precision Systems.
The process was visualized with a Leica MZFLIII microscope. Droplets
containing primers were allowed to dry completely, leaving dehydrated DNA
LAMP primers in the reaction wells. Following visual confirmation that all
droplets had dehydrated, the chip was submerged in heavy mineral oil
(Fisher) and placed in a desiccator to remove air bubbles. The primary
function of the mineral oil was to protect the reaction droplets from
evaporation during heating at 65 °C.During degassing, the primer-less RT-LAMP reaction was prepared and
transferred to the microinjection capillary. Reaction droplets were then
placed in each well by lightly contacting the bottom of each reaction well,
injecting a droplet of approximately 60 nL, and lifting the capillary out of
the oil. The chip containing all 36 droplets submerged in oil in the copper
bowl was then transferred to the heating stage and imaging apparatus
(fluorescence microscope or smartphone apparatus).Primer spotting and reaction droplet placement for the integrated
experiment were performed in the same manner but within a biosafety cabinet
under a Leica EZ4D microscope with a built-in camera providing a live video
feed to a personal computer (PC).
2.5 Fluorescence microscopy
Fluorescence microscopy images were captured with a Nikon Eclipse FN1
fluorescence microscope with a 2× objective and a Nikon 96311 B-2E/C
FITC fluorescence filter. NIS Elements software was used to capture fluorescence
images for RT-LAMP reactions containing purified RNA in water with 6.3×
gain and 1 s exposure time. Additional measurements (presented in the Supplementary
Information) with lysed whole blood spiked with viral RNA imaged with the
microscope required 8× gain and a 2 s exposure time to compensate for
decreased overall fluorescence intensity.
2.6 Smartphone imaging
Apparatus
A Samsung Galaxy Note 4 smartphone was purchased for the imaging of
the RT-LAMP reaction on the microchip substrate. The smartphone hardware was
not modified from its factory conditions. A Thorlabs 530 nm Longpass Colored
Glass Filter was placed between the camera and the chip to isolate the
fluorophore emission wavelengths. A 3D-printed cradle (Figure 1) was designed to position the smart-phone
horizontally with the camera directly above the microchip. A mounting
cylinder was also 3D printed to hold an Opto Diode Corp high-output blue
light-emitting diode (LED) and a Thorlabs Shortpass Filter with a 500 nm
cut-off wavelength, which fit within the cradle and illuminated the
microchip from an angle. The LED was powered with 3 V from an Agilent E364xA
DC power supply with an automated on/off function controlled with a MATLAB
script. It was also determined that the blue LED could be adequately powered
by a standard 3 V lithium coin battery, but the DC power supply was used for
the purpose of PC control.
Figure 1
The RT-LAMP substrate and smartphone apparatus
(a) Image of 1 cm × 1 cm silicon microchip substrate sitting on
a Samsung smartphone. (b) Scanning electron microscopy cross-section of 160
µm-deep reaction wells. (c) Schematic of microchip and heating stage in
fluorescence microscope apparatus, including: (i) heating stage, (ii) copper
base containing mineral oil, and (iii) fluorescence microscope objective. (d)
Expanded diagram of smartphone LAMP apparatus, including: (i) heating stage,
(ii) copper base containing mineral oil, (iii) wavelength filters placed in
front of the LED and smartphone camera, (iv) blue LED light source, and (v)
smartphone. (d-inset) Image of apparatus assembled in biosafety cabinet.
Software
Due to biosafety considerations, the entire smart-phone imaging
apparatus was placed inside a biosafety cabinet when performing integrated
measurements. For this reason, remote control of the imaging function was
desired, so the Android application IP Webcam [24] was downloaded from the Google Play store and
installed on the smartphone. This application transmits a live image over
the network, which can be viewed in real time in a web browser. The browser
interface allowed for control of the smartphone camera’s focus,
exposure, and gain. Imaging of the RT-LAMP reaction was performed with the
following parameters set in the IP Webcam web browser interface: 8×
zoom, 99% stream quality, exposure compensation of 4, and
“night vision” function with a gain of 10× and
exposure 10.A MATLAB script was written to automate the image-capture process.
The script was initialized simultaneously with the activation of the heating
stage. The MATLAB script imaged the reaction in the following sequence:
Switch on blue LED, delay 3 s; capture image from IP Webcam web browser
interface, delay 2 s; and finally switch off blue LED. This process was
repeated every 30 s while each reaction was imaged.
2.7 Data analysis
Image analysis
Images recorded with fluorescence microscopy or the smart-phone
imaging apparatus were saved as TIFF (microscope) or JPEG (smartphone) files
and fluorescence intensity was analyzed. For this analysis, the physical
location of each droplet was identified manually in a MATLAB script by
importing and displaying the image in a MATLAB figure and adjusting the
position of square boxes outlining each droplet. Grayscale TIFF images were
imported as a matrix of 16-bit unsigned integers (range 0–65 535)
representing each pixel in the image. Grayscale JPEG images were imported as
an array of 8-bit unsigned integers (range 0–255) representing each
pixel in the image.Following manual identification of droplet positions, a MATLAB
script automated the analysis of each droplet by averaging the numerical
value of all pixels within the region defined by the box outlining each
droplet. Absolute numerical values are a function of the range of integer
values (8-bit or 16-bit), as well as the exposure time and gain of the
camera, ambient light in the laboratory, and other factors. For this reason,
the baseline is subtracted from each measurement as described below, and
fluorescence measurements in this paper are presented in arbitrary units
(AU).
Threshold time analysis
Threshold time was determined from raw fluorescence data on all
platforms. First, the baseline was removed by subtracting an early
fluorescence measurement from all subsequent measurements. For all
thermocycler measurements and the microscope measurements with purified RNA
in water, this was the first fluorescence value or image recorded. For
smartphone measurements, the auto fluorescence of whole blood at room
temperature was observed to decrease quickly upon initial heating of the
chip. Thus, the baseline was defined to be 90 s after initialization of the
heating, or the third image recorded by the smartphone.The threshold time for each individual reaction was approximated
from baseline-subtracted fluorescence curves by determining the measurement
n at which the signal exceeded 20% of the
maximum fluorescence value it achieved during the course of the entire
measurement. After determining n, a linear fit was
determined by the fluorescence values I and
I in the form
I(x) = mx +
b and the threshold time Tt
= (0.2 × Imax −
b)/m was determined.
3 Results
3.1 Characterization of RT-LAMP in a benchtop thermocycler
Purified viral RNA in water
The first experiment, presented in Figure 2(a) and (b), was performed in order to characterize the
RT-LAMP reaction with purified analyte in a standard thermocycler apparatus.
HIV-1 IIIB RNA was purified by two different methods as described in Section
2.1. RT-LAMP fluorescence curves for the first method are shown in Figure 2(a), and the calculated threshold
time versus the average number of viruses in each reaction is shown in Figure 2(b). Linear fits to threshold
time versus log of viruses showed a difference in slope of less than
1.3% but a vertical offset of more than 4 min in the
y-intercept.
Figure 2
RT-LAMP performed in a standard benchtop thermocycler
(a) Raw fluorescence data for RT-LAMP of viral RNA diluted and purified
from dilutions of whole HIV-1 IIIB virus particles (RNA purification method 1).
(b) Threshold time curves determined by calculating the time at which
fluorescence curves exceed 20% of their maximum value. Data are included
for both methods of producing purified viral RNA. (c) Fluorescence curves from
six replicates of each condition comparing RT-LAMP in virus-spiked whole blood
versus purified RNA. All reactions contained the equivalent of 9380 virus
particles. (d) Comparisons of threshold time and overall fluorescence intensity
for both conditions. (e) Fluorescence curves and (f) standard curve for RT-LAMP
with HIV-1 IIIB whole viruses spiked in whole blood. The 9.4
vp·RXN−1 sample is not included in (f) due to
inconsistent amplification of all replicates. RXN is short for reaction.
Comparison with lysed whole blood
The next experiment compared the threshold time and fluorescence
intensity of RT-LAMP containing 9380 whole virus particles of HIV-1 IIIB in
lysed blood to the corresponding amount of purified HIV-1 RNA in water as an
initial test of the feasibility of RT-LAMP in lysed whole blood. Figure 2(c) shows fluorescence
measurements and Figure 2(d) provides
bar charts comparing threshold time and maximum fluorescence value. Six
replicates were performed of each of the two conditions, and the average
threshold time in lysed blood varied by less than 2.3% compared to
RNA in water, and gave a P of 0.0755 in a standard
two-sample t test. The maximum overall fluorescence
intensity (determined from raw fluorescence measurements with the baseline
subtracted) showed a decrease in fluorescence signal of 88.93% in
lysed whole blood compared to purified RNA in water.
Standard curve in lysed whole blood
An experiment with whole virus particles at a range of
concentrations was then performed in order to characterize the RT-LAMP
reaction on an ideal platform but with a minimally processed sample. Figure 2(e) (fluorescence intensity
curves) and (f) (threshold time versus virus number) show that even with a
ten-fold reduced overall fluorescence intensity, amplification curves could
still be observed and threshold times analyzed. A linear fit to the
threshold time versus the log of virus number gives a slope comparable to
purified RNA curves (10.3% difference compared to method 1 and
9.1% difference compared to method 2 in Figure 2(b)). Due to inconsistent amplification of all
replicates, the 9.4 vp·RXN−1 sample is not
included in the threshold time curve. Further characterization of the lysed
whole blood RT-LAMP reaction was performed by examining variations on the
ratio of blood sample to lysis buffer. Results are provided in the Supplementary
Information.
3.2 Microfluidic blood lysis module
On-chip lysis in a microfluidic channel, shown in Figure 3(a), was independently characterized in order to
determine the potential for automated sample handling and to characterize any
impact of the microfluidic mixing on the overall method. Whole blood samples
spiked with three different concentrations of HIV-1 IIIB were each mixed with
lysis buffer in the polydimethylsiloxane (PDMS) mixing chip driven by the
fluidics apparatus in a biosafety cabinet, as described in Section 2.3. Output
from the microfluidic chip was collected on three separate instances from each
sample and analyzed separately with RT-LAMP in a thermocycler. Triplicate
RT-LAMP reactions were performed for each of the three collections, resulting in
nine total RT-LAMP reactions for each virus concentration investigated. As a
control, the same spiked blood samples were added to lysis buffer, mixed with a
pipette, and analyzed with RT-LAMP in triplicate (i.e., three control reactions
in total for each virus concentration). Figure
3(b) shows the results. Mean threshold times differed from those of
the manually-pipetted control by 0.85%, 3.88%, and 8.21%
for post-lysis virus concentrations of 1349 vp·
µL−1, 135
vp·µL−1, and 13
vp·µL−1, respectively.
Figure 3
Microfluidic lysis of whole blood samples
(a) Diagram of the microfluidic device with bright-field microscopy at
various points in the channel with their approximate locations indicated. (b) A
comparison of threshold time at three virus concentrations for chip-lysed versus
pipette-lysed samples containing whole virus particles. These data verify that
the microfluidic lysis method does not result in significant differences in
signal compared to the manual method.
3.3 On-chip RT-LAMP
Next, the RT-LAMP reaction was demonstrated on the microchip. This
included experiments with purified RNA in water, shown in Figure 4(a) and (b), and RNA in lysed whole blood (presented
in the Supplementary
Information), imaged with a fluorescence microscope before the
introduction of the smartphone, as shown in Figure
4(c) and (d). This sequence of experiments was performed in order to
establish a basis for comparison and to limit the introduction of new variables
and sources of noise in each step in the progression of the experiments. All
on-chip RT-LAMP measurements were prepared as described in Section 2.4 with DNA
LAMP primers pre-spotted and dehydrated on the chip prior to oil immersion,
degassing, and reaction droplet placement.
Figure 4
On-chip RT-LAMP for HIV-1 IIIB
(a) Baseline-subtracted fluorescence intensity and (b) threshold time
versus virus concentration for purified RNA in water on the micro-well substrate
imaged with a fluorescence microscope. (c) Baseline-subtracted fluorescence
intensity and (d) threshold time versus virus concentration for RNA-spiked lysed
whole blood on the micro-well substrate imaged with the smartphone camera. (e)
Fluorescence images captured by the smartphone showing the amplification of four
RNA concentrations. (f) Endpoint measurements of the same chip in
(c)–(e) showing: 1) fluorescence microscopy, 2) the smartphone image, 3)
color overlay indicating concentrations (the gray bar indicates negative
controls), and 4) a fluorescence intensity colormap created in MATLAB in the
process for quantifying the intensity in images.
Purified RNA in water was first characterized on the chip. Figure 4(a) displays fluorescence curves
measured with a fluorescence microscope, while Figure 4(b) shows the threshold time analysis. In this measurement,
data from two droplets were removed due to outlying behavior believed to be due
to contamination with inhibitors or experimenter error. This includes one of six
reactions in each of the 75 vp· RXN −1 and 7.5E+2
vp·RXN−1 samples. Only two of six 7.5
vp·RXN−1 samples amplified, and as a result,
threshold times for this sample were omitted from the curve in Figure 4(b). In Figure 4(c)–(f), lysed whole blood spiked with viral RNA was
imaged with the smartphone apparatus as described in the Section 2.6. Whole
virus particles were not used in this measurement, as the apparatus had not yet
been converted to be contained within a biosafety cabinet. Figure 4(c) shows fluorescence curves gleaned from the
smartphone images, Figure 4(d) shows the
threshold time analysis, and Figure 4(e)
shows examples of the smartphone fluorescence images every minute for minutes
7–11. The 11 vp·RXN−1 (green) and 1.1 v
p· RXN −1 (blue) samples are omitted from the
threshold time curve because of in-sufficient amplification before 30 min.Figure 4(f) shows an endpoint
measurement obtained following termination of the real-time monitoring at 30
min. Four additional droplets (three of the 11 v p· RXN
−1 samples and one 1.1
vp·RXN−1 sample) in the array had reacted before
the image in Figure 4(f)-1) was captured
with the fluorescence microscope. Figure
4(f)-2) is an additional smartphone image captured shortly after the
microscopy image was obtained. Figure
4(f)-3) is identical to Figure
4(f)-2) with an additional color-coded overlay that is consistent
with the color-concentration convention used throughout this paper. The sixth
column on the far right is a negative control containing a blood sample and
reaction mix without viral RNA.Finally, Figure 4(f)-4) is a color
map rendering of Figure 4(f)-2) that was
produced in MATLAB as an example of the image analysis process.
3.4 Compatibility with other viral nucleic acids
Due to the common occurrence of co-infections of HIV and viral hepatitis
(HBV and/or HCV), we sought to demonstrate the compatibility of this on-chip
RT-LAMP assay with such cases. Because whole virus particles were not available
to us for hepatitis viruses, synthetic viral genomes of DNA (HBV) and RNA (HCV)
were obtained and tested. Since limited quantities of the synthetic viral
genomes were available to us, this experiment was performed in water and not
lysed whole blood in order to avoid the ten-fold dilution of nucleic acids that
would result from spiking lysed blood with viral nucleic acids and to perform
the test at the highest concentration of viral nucleic acids possible.Figure 5(a) shows fluorescence data
obtained with the fluorescence microscope measurement system for the on-chip
RT-LAMP of three samples: HIV both with and without hepatitis virus nucleic
acids, and hepatitis virus nucleic acids without HIV. All 24 of the droplets
containing HIV RNA successfully amplified (twelve with hepatitis virus nucleic
acid, twelve without), regardless of the presence of other nucleic acids. The
twelve HIV-negative droplets, all containing hepatitis virus nucleic acids, did
not amplify.
Figure 5
A demonstration of compatibility with common co-infections
The HIV-1 on-chip RT-LAMP reaction was tested in the presence of HBV and
HCV nucleic acids at concentrations equivalent to 1.6 × 103
of each virus per 60 nL reaction. LAMP primers for HIV-1 detection were
dehydrated in each well of the microchip array and purified nucleic acids in
water were prepared in various combinations with a primer-less RT-LAMP master
mix. The chip was immersed in mineral oil and placed under a fluorescence
microscope on a heating stage at 65 °C. (a) Fluorescence measurements
from the fluorescence microscopy chip of three combinations:
HIV+/HBV−/HCV−, HIV+/HBV+/HCV+, and HIV−/HBV+/HCV+. (b)
A bar chart comparing threshold time for HIV RNA-positive samples with and
without hepatitis virus nucleic acid present.
Figure 5(b) shows a comparison of
threshold time between the two HIV RNA-containing conditions. The average
threshold time for HIV-positive droplets that contained hepatitis nucleic acid
differed from the droplets that did not by 8.90%. A standard
t test comparing the two gives a P of
1.8718E-6.
3.5 Integrated experiments
Integrated experiments were designed to demonstrate the full capacity of
this approach for a sample-to-answer solution to point-of-care HIV viral load
quantification. Figure 6(a) depicts the
complete flow of the process. These experiments differed from the other
measurements presented in this paper in that whole blood samples were spiked
with whole HIV-1 IIIB virus particles (not viral RNA) at a range of
concentrations and each individual sample was analyzed on a separate chip. Of
the 36 wells on the microchip array, 6 were used for negative controls and up to
30 were used to test the sample.
Figure 6
Sample-to-answer RT-LAMP detection of HIV-1 IIIB in lysed whole blood
(a) A schematic of the integrated process: 1) Whole blood spiked with
HIV-1 IIIB was infused into a microfluidic apparatus; 2) 10 µL of sample
was metered based on the volume of the holding coil; 3) the sample was flowed
into a microfluidic mixing module at 10
µL·min−1 with cell lysis buffer at 40
µL·min−1; 4) output from the mixing
module was added to the RT-LAMP master mix without primers; 5) a lysed sample
with master mix was microinjected onto the microwell substrate prepared with
dehydrated primers; 6) the chip was heated to 65 °C in a copper base
with a heating stage and the RT-LAMP reaction was monitored by recording a
fluorescent image every 30 s using a smartphone. (b) Real-time fluorescence
curves as measured by smartphone imaging system. (c) Threshold time values
determined by the time at which baseline-subtracted fluorescence intensity
exceeded 20% of its maximum value. (d) Amplification efficiency, defined
as the fraction of droplets that amplified in the array for each tested
concentration.
Five samples were tested (named A–E), containing approximately
32 000, 3200, 320, 32, and 3.2 whole virus particles per reaction, respectively.
Since each reaction droplet contains approximately 4.8 nL of whole blood, this
corresponds to viremias in the range of 6.7 × 105–6.7
× 109 viruses per mL of blood, or 1.3 ×
106–1.3 × 1010 RNA copies per mL of
blood plasma (assuming 45% haematocrit).Due to biosafety considerations, the entire process was adapted, as
described previously, to be contained within a biosafety cabinet. This
adaptation introduced challenges to the droplet placement, which was performed
with a motorized micromanipulator controlled with a joystick and guided by a
video feed from a small tabletop microscope. Decreased control over droplet
placement led to a decreased success rate in droplet placement. As a result, not
all of the 30 wells designated for a virus-positive reaction droplet were used
in every measurement. The numbers of successfully placed droplets for samples A
through E were as follows: 29, 28, 30, 22, and 22.Figure 6(b) shows the fluorescence
curves, measured with the smartphone system, for all droplets that amplified
within 30 min. Figure 6(c) shows the
threshold time versus the virus number. The slope of the fit to threshold time
versus virus number is −1.9993, which differs in magnitude by
57.6% compared to on-chip RNA in lysed whole blood, by 34.9%
compared to on-chip RNA in water, by 18.5% compared to whole virus
particles in whole blood in the thermocycler, and by 25.9% compared to
purified RNA in water (method 2 in Section 2.1) analyzed in the thermocycler. A
t test was performed to compare the threshold time and
significance was determined by a P < 0.05.In Figure 6(d), we consider a new
metric—amplification efficiency— and observe a trend between
virus number and the fraction of droplets that amplified. A framework for
understanding this phenomenon in the context of digital LAMP measurements is
described in the Supplementary
Information.
4 Discussion
4.1 Characterization of RT-LAMP in a benchtop thermocycler
In the initial thermocycler characterization of this RT-LAMP
reaction, we presented a comparison of two methods (see Section 2.1) of
viral RNA purification in order to ➀ establish a baseline for a
“clean” reaction signal, and ➁ highlight possible
factors that may contribute to variations in RT-LAMP results in subsequent
analyses.While the two independent experiments measuring purified RNA in
water differed in the manner in which dilutions were performed (before RNA
purification in method 1 and after RNA purification in method 2), they also
differed in betaine content (0.8 mol· L−1 in
method 1, 0.4 mol· L−1 in method 2). After the
initial measurements, betaine concentration was decreased in order to allow
for a larger fraction of the reaction volume to consist of sample. One
initial hypothesis was that betaine, which contains a cation and reduces
secondary-structure formation in DNA, may explain the difference in the two
standard curves. However, a careful comparison was performed in which RNA at
various low concentrations was added to a common master mix containing 0.4
mol·L−1 betaine and, in half of the
reactions, additional betaine was added to achieve a concentration of 0.8
mol·L−1. The results of this control
experiment (see the Supplementary Information) indicated no difference in the
threshold time nor in the amplification efficiency at lower concentrations.
Therefore, other factors likely explain the vertical offset in the purified
RNA experiment in the thermocycler.We suspect that these factors may include variations in enzyme
concentration due to inherent variation in the pipetting process when
preparing master mixes, a time-related decrease in enzyme activity (the two
experiments were performed several weeks apart and enzyme activity may have
decreased with freeze-thaw cycles of the reagent), or variations in ambient
temperature and other factors. Additionally, degradation of RNA over time
during storage may have decreased the yield of the samples used in method 2,
and variations due to manual pipetting may contribute to discrepancies
between the actual RNA concentrations gleaned from the two independent
purification procedures.While these factors are important to consider, we chose at this time
to acknowledge their potential effects and the need to minimize variation
and decided to establish rigorous controls in future experimental or
manufacturing processes. Viral RNA stability or yield from the RNA
purification process would not affect the results of experiments using whole
virus particles. Subsequent analysis involving purified viral RNA or viral
RNA spiked in whole blood employed RNA from a third, “fresh”
purification identical in protocol to purification method 2. The purified
RNA from this process was aliquoted and frozen at −80 °C and
thawed only immediately before use in an experiment in order to minimize
degradation of the sample between experiments.To our knowledge, a quantitative RT-LAMP measurement of HIV
concentration in whole blood processed only by mixing with a cell lysis
buffer has not been described previously in literature. The thermocycler
results with whole virus particles spiked in whole blood and mixed with cell
lysis buffer suggest the possibility of quantifying virus concentration
simply based on reaction kinetics.The decrease in overall fluorescence intensity indicates that there
are fluorescence-quenching components in the lysed whole blood sample.
However, because the threshold time is identical when comparing purified RNA
and whole blood, we can conclude that whole blood does not affect the
amplification efficiency. This is a very promising development, given that a
major challenge to most point-of-care diagnostics is the process of
isolating analyte from complex biological samples in the absence of
controlled environments, skilled technicians, or laboratory instruments
[25]. In this case, blood cell
lysis is an extremely simple processing step compared to the more
complicated techniques described in the literature [6].
4.2 Microfluidic blood lysis module
One goal of this paper is to demonstrate the potential for a fully
automated RT-LAMP viral load test from a drop (i.e., a finger prick) of whole
blood. Such a test requires complete on-chip sample processing from that whole
blood drop, which has traditionally been a major barrier for many point-of-care
diagnostics approaches. Despite the popularity of PDMS prototyping in this
field, many microfluidic and lab-on-a-chip techniques suffer from the ability to
translate PDMS devices to commercially viable forms that are compatible with
injection molding and other mass-manufacturing techniques [26]. The converse is also true: That designs compatible
with manufacturing may be difficult to prototype in PDMS when some properties
(e.g., surface-fluid contact angle) are not comparable. Here we demonstrate
valve-assisted sample metering and microfluidic mixing resembling a method that
we are aware is employed in commercial-grade platforms for blood-sample
handling.Volumetric metering begins with a drop of blood of unspecified size from
which 10 µL is precisely metered in a holding coil with an inner
diameter of 203.2 µm, reminiscent of commercial microfluidic cartridges
that employ volumetric metering. Viable methods have also been employed for
fluid mixing in commercial microfluidic designs, making our simple serpentine
channel prototyped in PDMS a reasonable design [27].Data in Figure 3(b) demonstrate
that this simple approach of volumetric metering with the serpentine-channel
mixing of blood with lysis buffer is compatible with the downstream RT-LAMP
analysis. The results from three separate collections show that the method
produces an accurate mixing ratio, and that the on-chip mixing is at least as
consistent as manual pipetting.
4.3 On-chip RT-LAMP
Purified RNA with fluorescence microscope imaging
It is unclear what absolute conclusions can be gleaned from a
comparison of the fit to the threshold time versus the log of virus number
in Figures 2(b) and 4(b). The slope of the fit is smaller in
magnitude for the on-chip measurement, suggesting that there may be some
chip-related factors leading to a decrease in sensitivity for the
quantification of virus number by threshold time analysis. This observation
prompted increasing the frequency of images from 60 s (used in thermo-cycler
and microscope measurements) to 30 s (used in smart-phone measurements).The y-intercept of the fit in Figure 4(b) is also significantly smaller than those in
Figure 2(b) (14.68 min versus 29.1
min and 33.3 min), suggesting that the smaller droplet size may contribute
to a more rapid RT-LAMP reaction, a phenomenon that has been discussed
elsewhere [28]. This result may be
leveraged toward achieving the end goal of a rapid viral load test.
Lysed whole blood with smartphone imaging
Data from lysed blood measurements imaged with the smartphone
demonstrate two significant steps toward the goals of this paper. The
replacement of laboratory hardware (e.g., thermocycler fluorescence
detection or fluorescence microscope) with a common smartphone is a core
aspect of the novelty of this paper. First, this paper demonstrates that the
lysed blood RT-LAMP measurement can be performed with existing mobile
technology that is at least as affordable as a high-end smart-phone.
Furthermore, trends suggest that mobile communications technologies will
continue to improve in capabilities and decrease in cost: an exciting
outlook for fluorescence and other optics-based point-of-care diagnostics.
Second, our platform (consisting of a 3D printed platform, an LED light
source, an emission filter, and a small form-factor heating stage) suggests
that, if adequately robust, an add-on component may be developed as an
attachment to existing smart-phones, shifting the computation and imaging
burden from components integrated with the diagnostic platform to a consumer
item that is becoming ubiquitous, even in resource-limited settings [29]. This shift could significantly
reduce the production and deployment costs of such a technology.
4.4 Compatibility with other viral nucleic acids
Anticipating co-infection with other blood-borne viruses may be
important for practical HIV nucleic acid tests, since the incidence of
co-infection with HIV and one or more other viruses is high, particularly in
populations of intravenous drug users [2,
30]. Most significantly, we have
demonstrated that hepatitis viral nucleic acids at high concentrations
(equivalent to approximately 1.6 × 103 of each virus per 60
nL reaction) in purified form do not amplify in the RT-LAMP assay that is
designed for HIV.A more detailed analysis gives a P of 1.8718E-6 from a
standard t test, indicating a significant decrease in threshold
time for the sample containing nucleic acid from all three viruses versus the
sample containing only RNA from HIV. This result indicates a need for more
rigorous characterization of this phenomenon in future work. One explanation for
a decreased threshold time is the presence of the hepatitis B genome, which is a
circular, partially dsDNA, at a relatively high concentration (approximately
1600 copies·RXN−1). Its presence may produce an
effect that is not seen in whole blood, since mature erythrocytes do not contain
DNA, and our simulated co-infected sample would contain a new signal source from
dsDNA intercalating dye. Leukocytes in the lysed whole blood sample would be
very rare (a few per nL of blood), and genomic DNA from these sources may be
packaged and largely inaccessible to the dsDNA dye. Because no amplification was
observed in the HIV-negative, HBV/HCV-positive samples, we consider it unlikely
that HCV RNA or HBV DNA is acting as a non-specific template for LAMP and
producing incorrect reaction products.
4.5 Integrated experiments
Our integrated experiments demonstrate the capacity of the test for
quantitative viral load measurements based on reaction kinetics or digital
statistical methods. The slope of the fit to the threshold time plot in Figure 6(c) is comparable to the original
characterization shown in Figure 2,
suggesting that the integrated approach has the capacity to be
quantitative— and perhaps can be demonstrated to have greater
sensitivity upon optimization of reaction chemistry, incubation temperature, and
other factors. Bars showing differences that are statistically significant
between concentrations indicate the resolution of the integrated test performed
here at a 95% confidence level.Clearly, several factors need to be improved and addressed in order to
move toward a fully automated platform. The experiment flow described in Figure 6(a) consists of several components
that are compatible with a point-of-care, micro-fluidic cartridge-based
in vitro diagnostic platform: the volumetric metering of 10
µL of whole blood, microfluidic sample processing, nanolitersized
reaction droplets, a silicon chip substrate, and smartphone fluorescence
imaging. The process described here does, however, include the manual step of
transferring lysed blood from the microfluidic module, mixing it with
primer-less RT-LAMP master mix, and placing droplets onto the chip.We believe the issues requiring this handling step can be addressed by
common industry methods not easily demonstrated in the research lab, such as
sample distribution into a nanoliter-scale reaction well [31]. The issue of the addition of reaction components to
the sample could be addressed by lyophilization or by freeze-drying reagents
that are then re-hydrated by the sample. Lyophilized LAMP master mixes have been
described in Refs. [32, 33]. In addition, although air-drying
reaction components other than primers were problematic on our current platform,
we observed that the RT-LAMP master mix used here can be lyophilized in a
commercial freeze-drier in 0.2 mL reaction tubes (see the Supplementary
Information), and that amplification capabilities are retained. We did
not have the equipment available to attempt this process on-chip.
4.6 Quantification by threshold time
The measurements presented here exhibit a trend in threshold time versus
virus number that suggests that a kinetic measurement based on the time takes an
array of droplets to react may be a suitable method to quantify viral load,
though the limits of the resolution are still to be determined in an improved
platform. For this approach, the sensitivity of the reaction, determined by the
slope of the fit to threshold time versus virus number, may need to be increased
by optimizing the reaction, including adjustments to enzyme and buffer
concentrations or to incubation temperature. Additional optimization would need
to be performed in order to improve (i.e., lower) the lower limit of detection
of the reaction.Although these improvements may be made, a lysed whole blood approach
will be inherently limited in its capabilities by the reaction volume and
dilution factor in lysis buffer. For this reason, a variation on the approach
involving digital LAMP may be considered.
4.7 Quantification by digital LAMP
Digital LAMP and PCR approaches have been widely described in Refs.
[21, 34–39]. The primary
advantage of a digital approach is that it relies only on an endpoint
measurement—whether a reaction of a small volume amplifies or
not—from which concentration can be approximated by measuring hundreds,
thousands, or millions of droplets. The approach we describe here could be
scaled up for digital LAMP by constructing large arrays of droplets with an
automated distribution method.The upper and lower limits of viral load detection in a digital approach
are defined by the number of individual reaction droplets and the total volume
of the sample tested. Since the distribution of viruses in reaction droplets is
governed by Poisson statistics, we have briefly reviewed these principles in the
Supplementary
Information for this paper in order to demonstrate the theoretical
utility in clinical HIV management of a scaled-up platform that tests a
finger-prick droplet of whole blood. For example, a digital LAMP approach
requiring just 9 µL of whole blood would be capable of indicating viral
loads lower than 500 mL−1 of whole blood, with much greater
accuracy in the range of 104–106
mL−1 (see the Supplementary Information). While this approach cannot
compete with the technical specifications of state-of-the-art systems (lower
limit < 10 mL−1), it would be of practical value,
such as in showing declines in viremia following a drug regimen change or in
identifying cases of viral rebound in settings where the gold standard of care
is inaccessible [40]. The capacity of
this platform as a digital LAMP test increases with larger sample sizes and
increased number of reaction droplets.
5 Conclusions
In performing these experiments and the preparation of this manuscript, the
challenge of finding a sample-to-answer point-of-care HIV viral load quantification
solution was viewed as two parallel objectives: ➀ sample processing, which
traditionally involves isolating or enriching the analyte from its complex matrix,
and ➁ analyte detection. We broadly considered various approaches to sample
processing that might be integrated with a microchip LAMP approach, many of which
were guided by the notion that the presence of cellular material is not compatible
with nucleic acid amplification methods. Many of these approaches resulted in the
dilution of the analyte by a factor of ten or more, while our approach results in a
dilution by only a factor of five, prior to the addition of LAMP reagents with one
simple and easily implemented processing step that neither purifies nor enriches.
This key merit of our approach can significantly reduce the complexity of and cost
for a point-of-care device.The measurements presented here demonstrate that an RT-LAMP quantification
approach is indeed compatible with minimally processed whole blood. To our
knowledge, RT-LAMP in lysed whole blood has only been employed by one group, which
performed a non-quantitative measurement in a reaction tube on a portable heating
device [17]. We demonstrate quantitative
detection with the ability to resolve 10-fold changes in concentration above 6.7E+4
µL−1 and 100-fold changes in concentrations above 670
µL−1. We observed 60 nL droplets with as few as three
viruses per reaction amplify, which corresponds to a whole blood virus concentration
of 670 µL−1. We also discussed that the true power of
this approach may be in a quantitative digital LAMP format rather than a kinetic
measurement. Our implementation of the lysed whole blood approach in a microchip
format with mobile phone imaging represents a significant stride toward a practical
solution to viral load measurements in resource-limited settings.
Authors: M P de Baar; M W van Dooren; E de Rooij; M Bakker; B van Gemen; J Goudsmit; A de Ronde Journal: J Clin Microbiol Date: 2001-04 Impact factor: 5.948
Authors: Christopher M Hindson; John R Chevillet; Hilary A Briggs; Emily N Gallichotte; Ingrid K Ruf; Benjamin J Hindson; Robert L Vessella; Muneesh Tewari Journal: Nat Methods Date: 2013-09-01 Impact factor: 28.547
Authors: Michael Mauk; Jinzhao Song; Haim H Bau; Robert Gross; Frederic D Bushman; Ronald G Collman; Changchun Liu Journal: Lab Chip Date: 2017-01-31 Impact factor: 6.799
Authors: Sasha Herbst de Cortina; Claire C Bristow; Romney Humphries; Silver Keith Vargas; Kelika A Konda; Carlos F Caceres; Jeffrey D Klausner Journal: Sex Transm Dis Date: 2017-07 Impact factor: 2.830
Authors: Benjamin P Sullivan; Yu-Shan Chou; Andrew T Bender; Coleman D Martin; Zoe G Kaputa; Hugh March; Minyung Song; Jonathan D Posner Journal: Lab Chip Date: 2022-06-14 Impact factor: 7.517
Authors: Aaron M Jankelow; Hankeun Lee; Weijing Wang; Trung-Hieu Hoang; Amanda Bacon; Fu Sun; Seol Chae; Victoria Kindratenko; Katherine Koprowski; Robert A Stavins; Dylann D Ceriani; Zachary W Engelder; William P King; Minh N Do; Rashid Bashir; Enrique Valera; Brian T Cunningham Journal: Analyst Date: 2022-08-22 Impact factor: 5.227
Authors: Jinzhao Song; Vikram Pandian; Michael G Mauk; Haim H Bau; Sara Cherry; Laurence C Tisi; Changchun Liu Journal: Anal Chem Date: 2018-03-22 Impact factor: 6.986
Authors: Esther E Freeman; Aggrey Semeere; Hany Osman; Gary Peterson; Milind Rajadhyaksha; Salvador González; Jeffery N Martin; R Rox Anderson; Guillermo J Tearney; Dongkyun Kang Journal: Biomed Opt Express Date: 2018-03-26 Impact factor: 3.732
Authors: Taylor J Moehling; Dong Hoon Lee; Meghan E Henderson; Mariah K McDonald; Preston H Tsang; Seba Kaakeh; Eugene S Kim; Steven T Wereley; Tamara L Kinzer-Ursem; Katherine N Clayton; Jacqueline C Linnes Journal: Biosens Bioelectron Date: 2020-08-08 Impact factor: 10.618