| Literature DB >> 26695490 |
Adrienne E Campbell-Washburn1, Toby Rogers2, Burcu Basar3,4, Merdim Sonmez5, Ozgur Kocaturk6,7, Robert J Lederman8, Michael S Hansen9, Anthony Z Faranesh10.
Abstract
BACKGROUND: CMR-guidance has the potential to improve tissue visualization during cardiovascular catheterization procedures and to reduce ionizing radiation exposure, but a lack of commercially available CMR guidewires limits widespread adoption. Standard metallic guidewires are considered to be unsafe in CMR due to risks of RF-induced heating. Here, we propose the use of RF-efficient gradient echo (GRE) spiral imaging for reduced guidewire heating (low flip angle, long readout), in combination with positive contrast for guidewire visualization.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26695490 PMCID: PMC4688983 DOI: 10.1186/s12968-015-0219-9
Source DB: PubMed Journal: J Cardiovasc Magn Reson ISSN: 1097-6647 Impact factor: 5.364
Fig. 1a Pulse sequence diagram for spiral gradient echo imaging demonstrating change in slice refocusing gradient to transition from standard anatomical contrast (1) to positive contrast (2). b Resulting in vivo anatomical (1) and positive contrast (2) images with red arrowheads indicating nitinol guidewire in the aorta of a Yorkshire swine
Fig. 2Image processing algorithm used to isolate the guidewire signal from the background in the positive contrast image. Thresholding (Steps 1, 2 and 3) and selection of elongated connected structures (Step 4 and 5) are displayed. The device signal is overlaid in green on the anatomical image and displayed to the interventionist for procedural guidance
Fig. 3Example left heart catheterization data set showing anatomical images (a), positive contrast images (b) and resulting color overlay (c) for three different insertion lengths: guidewire tip in descending aorta (left), guidewire tip at aortic arch (middle) and guidewire in left ventricle (right)
Fig. 4Temperature increase observed in an acrylic gel phantom during 2 minutes of continuous scanning (scanning period indicated by vertical black lines). Temperature is measured at the tip of the nitinol guidewire for the guidewire and catheter co-located (a), guidewire extended 10 cm distal to catheter (b) and guidewire alone (c). Right-hand plots show reduced temperature scale (−0.2 °C to 1.2 °C ) to demonstrate subtle heating during GRE imaging
Nitinol guidewire heating in ASTM gel phantom
| Wire + Catheter | Wire 10 cm distal to catheter | Wire only | ||||
|---|---|---|---|---|---|---|
| Measured | Theoretical | Measured | Theoretical | Measured | Theoretical | |
| Cartesian bSSFP (flip = 60°, TR = 2.54 ms) | 50.39 °C | -- | 5.41 °C | -- | 5.90 °C | -- |
| Cartesian bSSFP (flip = 45°, TR = 2.54 ms) | 27.37 °C | 28.34 °C | 3.10 °C | 3.05 °C | 3.05 °C | 3.32 °C |
| Cartesian GRE (flip = 20°, TR = 2.86 ms) | 4.31 °C | 4.97 °C | 0.56 °C | 0.53 °C | 0.65 °C | 0.58 °C |
| Cartesian GRE (flip = 10°, TR = 2.86 ms) | 1.07 °C | 1.24 °C | 0.14 °C | 0.13 °C | 0.16 °C | 0.15 °C |
| Spiral GRE (16 interleaves, flip = 10°, TR = 5.16 ms) | 0.61 °C | 0.70 °C | 0.09 °C | 0.08 °C | 0.08 °C | 0.08 °C |
| Spiral GRE (8 interleaves, flip = 10°, TR = 8.60 ms) | 0.37 °C | 0.44 °C | 0.03 °C | 0.05 °C | 0.07 °C | 0.05 °C |
Temperature increase from baseline observed during 2 minutes of continuous imaging in an acrylic gel phantom. Temperature was measured at the tip of the nitinol guidewire (Nitrex, Covidien, Plymouth, MN) with the guidewire collocated with the catheter, with the guidewire extended 10 cm distal to catheter and with the wire alone. Theoretical predications, based on flip angle and TR (Eqn2), of the relative heating compared to Cartesian bSSFP (flip angle = 60°) are also displayed
Fig. 5In vivo temperature measurements at the tip of the nitinol guidewire and catheter at three insertion lengths: guidewire tip within the introducer sheath (a), in the descending aorta (b) and at the aortic arch (c)
Nitinol guidewire heating in vivo
| Wire + Catheter | Wire 10 cm distal to catheter | |||||
|---|---|---|---|---|---|---|
| Sheath | Aorta | Arch | Sheath | Aorta | Arch | |
| Cartesian bSSFP (flip = 60°, TR = 2.86 ms) | 4.34 °C | 0.18 °C | 0.14 °C | 0.22 °C | 0.13 °C | 0.05 °C |
| Cartesian bSSFP (flip = 45°, TR = 2.86 ms) | 2.50 °C | 0.08 °C | 0.10 °C | 0.17 °C | 0.09 °C | 0.04 °C |
| Spiral GRE (16 interleaves, flip = 10°, TR = 5.16 ms) | 0.09 °C | −0.04 °C | 0.01 °C | −0.02 °C | −0.02 °C | 0.00 °C |
| Spiral GRE (8 interleaves, flip = 10°, TR = 8.60 ms) | 0.06 °C | 0.04 °C | 0.06 °C | 0.00 °C | 0.03 °C | 0.03 °C |
In vivo temperature increase from baseline observed during 2 minutes of continuous imaging measured at the tip of the nitinol guidewire. Temperature was measured with the guidewire tip in the introducer sheath, in the aorta and at the aortic arch in a 60 kg swine