Literature DB >> 15508156

RF-induced temperature elevation along metallic wires in clinical magnetic resonance imaging: influence of diameter and length.

Cristina Armenean1, Emmanuel Perrin, Mircea Armenean, Olivier Beuf, Frank Pilleul, Hervé Saint-Jalmes.   

Abstract

With the development of interventional MRI, heating of biological tissues along the metallic wires in the MRI scanner has become an important issue. To assess thermal response to RF exposure during MRI, we studied the temperature elevation near nonmagnetic metallic wires. All tests were performed on a 1.5 T clinical scanner. Four experiments were conducted to investigate the effects of the wire diameter, the excitation flip angle, the temperature distribution along the wire, and the wire length. Electromagnetic simulations of the experimental setup were made with the use of commercial method of moments (MoM) software and numerical simulations of Hallen's equations. Comparisons between measured and calculated values of the electric field are presented. This study demonstrates that 1) temperature decreases with the diameter of the wire,2) temperature increases quadratically with the excitation flip angle, 3) heating occurs not only at the tip but also along the wire, and 4) the heating peaks are not obtained for the classical resonant length multiple of lambda/4 (where lambda is the RF field wavelength). In addition, significant and rapid heating increases were observed in the close vicinity of the wire. (c) 2004 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2004        PMID: 15508156     DOI: 10.1002/mrm.20246

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  30 in total

1.  RF Heating of MRI-Assisted Catheter Steering Coils for Interventional MRI.

Authors:  Fabio Settecase; Steven W Hetts; Alastair J Martin; Timothy P L Roberts; Anthony F Bernhardt; Lee Evans; Vincent Malba; Maythem Saeed; Ronald L Arenson; Walter Kucharzyk; Mark W Wilson
Journal:  Acad Radiol       Date:  2010-11-13       Impact factor: 3.173

2.  Experience with magnetic resonance imaging of human subjects with passive implants and tattoos at 7 T: a retrospective study.

Authors:  Yacine Noureddine; Andreas K Bitz; Mark E Ladd; Markus Thürling; Susanne C Ladd; Gregor Schaefers; Oliver Kraff
Journal:  MAGMA       Date:  2015-09-26       Impact factor: 2.310

Review 3.  [Physical interactions in MRI: Some rules of thumb for their reduction].

Authors:  M Mühlenweg; G Schaefers; S Trattnig
Journal:  Radiologe       Date:  2015-08       Impact factor: 0.635

4.  Analysis of the role of lead resistivity in specific absorption rate for deep brain stimulator leads at 3T MRI.

Authors:  Leonardo M Angelone; Jyrki Ahveninen; John W Belliveau; Giorgio Bonmassar
Journal:  IEEE Trans Med Imaging       Date:  2010-03-22       Impact factor: 10.048

Review 5.  Real-time, Interactive MRI for cardiovascular interventions.

Authors:  Elliot R McVeigh; Michael A Guttman; Peter Kellman; Amish N Raval; Robert J Lederman
Journal:  Acad Radiol       Date:  2005-09       Impact factor: 3.173

6.  High-resolution MR imaging appearance of colonic tissue in rabbits using an endoluminal coil.

Authors:  Frank Pilleul; Olivier Beuf; Claire Godefroy; Jean-Yves Scoazec; Mircea Armenean; Cristina Armenean; Emmanuel Perrin; Pierre-Jean Valette; Hervé Saint Jalmes
Journal:  MAGMA       Date:  2005-11-18       Impact factor: 2.310

Review 7.  [Safety aspects in high-field magnetic resonance imaging].

Authors:  M Mühlenweg; G Schaefers; S Trattnig
Journal:  Radiologe       Date:  2008-03       Impact factor: 0.635

8.  Heating and safety of a new MR-compatible guidewire prototype versus a standard nitinol guidewire.

Authors:  Malgorzata Wolska-Krawczyk; Martin A Rube; Erwin Immel; Andreas Melzer; Arno Buecker
Journal:  Radiol Phys Technol       Date:  2013-11-08

9.  RF Heating of Gold Cup and Conductive Plastic Electrodes during Simultaneous EEG and MRI.

Authors:  Mukund Balasubramanian; William M Wells; John R Ives; Patrick Britz; Robert V Mulkern; Darren B Orbach
Journal:  Neurodiagn J       Date:  2017

Review 10.  Magnetic Resonance Sequences and Rapid Acquisition for MR-Guided Interventions.

Authors:  Adrienne E Campbell-Washburn; Anthony Z Faranesh; Robert J Lederman; Michael S Hansen
Journal:  Magn Reson Imaging Clin N Am       Date:  2015-08-12       Impact factor: 2.266

View more