| Literature DB >> 20373408 |
Ashvin K George1, J Andrew Derbyshire, Haris Saybasili, Christina E Saikus, Ozgur Kocaturk, Michael A Guttman, Elliot R McVeigh, Robert J Lederman, Anthony Z Faranesh.
Abstract
The accurate visualization of interventional devices is crucial for the safety and effectiveness of MRI-guided interventional procedures. In this paper, we introduce an improvement to the visualization of active devices. The key component is a fast, robust method ("CurveFind") that reconstructs the three-dimensional trajectory of the device from projection images in a fraction of a second. CurveFind is an iterative prediction-correction algorithm that acts on a product of orthogonal projection images. By varying step size and search direction, it is robust to signal inhomogeneities. At the touch of a key, the imaged slice is repositioned to contain the relevant section of the device ("SnapTo"), the curve of the device is plotted in a three-dimensional display, and the point on a target slice, which the device will intersect, is displayed. These features have been incorporated into a real-time MRI system. Experiments in vitro and in vivo (in a pig) have produced successful results using a variety of single- and multichannel devices designed to produce both spatially continuous and discrete signals. CurveFind is typically able to reconstruct the device curve, with an average error of approximately 2 mm, even in the case of complex geometries.Entities:
Mesh:
Year: 2010 PMID: 20373408 PMCID: PMC2852271 DOI: 10.1002/mrm.22307
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668