OBJECTIVE: To use a Likert scale method to optimize image quality (IQ) for cone beam CT (CBCT) soft-tissue matching for image-guided radiotherapy of the prostate.
METHODS: 23 males with local/locally advanced prostate cancer had the CBCT IQ assessed using a 4-point Likert scale (4 = excellent, no artefacts; 3 = good, few artefacts; 2 = poor, just able to match; 1 = unsatisfactory, not able to match) at three levels of exposure. The lateral separations of the subjects were also measured. The Friedman test and Wilcoxon signed-rank tests were used to determine if the IQ was associated with the exposure level. We used the point-biserial correlation and a χ(2) test to investigate the relationship between the separation and IQ.
RESULTS: The Friedman test showed that the IQ was related to exposure (p = 2 × 10(-7)) and the Wilcoxon signed-rank test demonstrated that the IQ decreased as exposure decreased (all p-values <0.005). We did not find a correlation between the IQ and the separation (correlation coefficient 0.045), but for separations <35 cm, it was possible to use the lowest exposure parameters studied.
CONCLUSION: We can reduce exposure factors to 80% of those supplied with the system without hindering the matching process for all patients. For patients with lateral separations <35 cm, the exposure factors can be reduced further to 64% of the original values. ADVANCES IN KNOWLEDGE: Likert scales are a useful tool for measuring IQ in the optimization of CBCT IQ for soft-tissue matching in radiotherapy image guidance applications.

Entities:  

Mesh:

Year:  2015        PMID: 26689092      PMCID: PMC4986481          DOI: 10.1259/bjr.20150460

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  6 in total

1.  Multislice CT of the pelvis: dose reduction with regard to image quality using 16-row CT.

Authors:  Jessen Gurung; M Fawad Khan; Adel Maataoui; C Herzog; R Bux; H Bratzke; Hanns Ackermann; Thomas J Vogl
Journal:  Eur Radiol       Date:  2005-04-02       Impact factor: 5.315

2.  Limited cone-beam computed tomography imaging of the middle ear: a comparison with multislice helical computed tomography.

Authors:  L I Peltonen; A A Aarnisalo; M K Kortesniemi; A Suomalainen; J Jero; S Robinson
Journal:  Acta Radiol       Date:  2007-03       Impact factor: 1.990

Review 3.  Review of relationships between physical measurements and user evaluation of image quality.

Authors:  M J Tapiovaara
Journal:  Radiat Prot Dosimetry       Date:  2008-02-05       Impact factor: 0.972

4.  Coronary artery stents: influence of adaptive statistical iterative reconstruction on image quality using 64-HDCT.

Authors:  Cathérine Gebhard; Michael Fiechter; Tobias A Fuchs; Julia Stehli; Ennio Müller; Barbara E Stähli; Caroline E Gebhard; Jelena R Ghadri; Bernd Klaeser; Oliver Gaemperli; Philipp A Kaufmann
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2013-02-20       Impact factor: 6.875

5.  Effect of dose reduction on image registration and image quality for cone-beam CT in radiotherapy.

Authors:  B Loutfi-Krauss; J Köhn; N Blümer; K Freundl; T Koch; E Kara; C Scherf; C Rödel; U Ramm; J Licher
Journal:  Strahlenther Onkol       Date:  2014-09-20       Impact factor: 3.621

6.  A comprehensive study on the relationship between the image quality and imaging dose in low-dose cone beam CT.

Authors:  Hao Yan; Laura Cervino; Xun Jia; Steve B Jiang
Journal:  Phys Med Biol       Date:  2012-04-07       Impact factor: 3.609

  6 in total
  4 in total

1.  Patient-based low dose cone beam CT acquisition settings for prostate image-guided radiotherapy treatments on a Varian TrueBeam linear accelerator.

Authors:  Maria Antonietta Piliero; Margherita Casiraghi; Davide Giovanni Bosetti; Simona Cima; Letizia Deantonio; Stefano Leva; Francesco Martucci; Marino Tettamanti; Francesco Pupillo; Luca Bellesi; Antonella Richetti; Stefano Presilla
Journal:  Br J Radiol       Date:  2020-08-27       Impact factor: 3.039

2.  Dose reduction potential of using gold fiducial markers for kilovoltage image-guided radiotherapy.

Authors:  Yoshiki Takei; Hajime Monzen; Mikoto Tamura; Hiroshi Doi; Yasumasa Nishimura
Journal:  J Appl Clin Med Phys       Date:  2020-09-22       Impact factor: 2.102

3.  Registration accuracy with the low dose kilovoltage cone-beam CT: A phantom study.

Authors:  Yoshiki Takei; Hajime Monzen; Kenji Matsumoto; Kohei Hanaoka; Mikoto Tamura; Yasumasa Nishimura
Journal:  BJR Open       Date:  2019-08-29

4.  CBCT imaging: a simple approach for optimising and evaluating concomitant imaging doses, based on patient-specific attenuation, during radiotherapy pelvis treatment.

Authors:  Caroline Ordóñez-Sanz; Mark Cowen; Neda Shiravand; Niall D MacDougall
Journal:  Br J Radiol       Date:  2021-08-01       Impact factor: 3.039

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.