Hemoximetry as the "gold standard"? Error assessment based on differences among identical blood gas analyzer devices of five manufacturers.

Hartmut Gehring1, Lutz Duembgen, Mareike Peterlein, Söhnke Hagelberg, Leif Dibbelt.   

Abstract

BACKGROUND: The calibration and testing procedures of a pulse oximeter with arterial blood samples from healthy subjects are based on reference values from the hemoximeter. There are no tests to identify the accuracy of the reference devices. Because of this limitation and since the true values of oxygen saturation (sO2 in %) in blood samples were not known, we used the differences between two identical devices, A and B, for error assessment.
METHODS: Two identical devices, A and B, from five leading manufacturers were investigated. Seventy-two arterial blood samples from 12 healthy volunteers at three different levels of saturation between 100% and 70% sO2 were randomly evaluated by the test systems.
RESULTS: The observed differences (Delta) between Devices A and B, as a measure for the error of the hemoximeters, increased significantly with all manufacturers from level 97 (Deltamin, -0.9%; Deltamax, 2.6%) to 85 (Deltamin, -2.4%; Deltamax, 4.3), this effect was even stronger between levels 97 and 75 (Deltamin, -4.6%; Deltamax, 4.3%). A variance proportion analysis revealed the concentration of the reduced hemoglobin as the main error source for sO2 measurements. Independent from the sO2 levels there were also significant differences for the carboxy hemoglobin concentration in the range of 0%-4% and for the methemoglobin concentration in the range of 0%-1%.
CONCLUSIONS: The variance of sO2 measurements between identical devices increased significantly when saturation decreased from the normal level of 97% to the hypoxemic levels of 85% and 75%.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18048894     DOI: 10.1213/01.ane.0000268713.58174.cc

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  11 in total

1.  Quantitative mapping of cerebral metabolic rate of oxygen (CMRO2 ) using quantitative susceptibility mapping (QSM).

Authors:  Jingwei Zhang; Tian Liu; Ajay Gupta; Pascal Spincemaille; Thanh D Nguyen; Yi Wang
Journal:  Magn Reson Med       Date:  2014-09-26       Impact factor: 4.668

Review 2.  Factors affecting hemoglobin measurement.

Authors:  Lauren Berkow
Journal:  J Clin Monit Comput       Date:  2013-03-26       Impact factor: 2.502

3.  Non-invasive estimation of jugular venous oxygen saturation: a comparison between near infrared spectroscopy and transcutaneous venous oximetry.

Authors:  Douglas A Colquhoun; Jason M Tucker-Schwartz; Marcel E Durieux; Robert H Thiele
Journal:  J Clin Monit Comput       Date:  2012-04       Impact factor: 2.502

4.  Accuracy of non-invasive continuous total hemoglobin measurement by Pulse CO-Oximetry in severe traumatized and surgical bleeding patients.

Authors:  Werner Baulig; Burkhardt Seifert; Donat R Spahn; Oliver M Theusinger
Journal:  J Clin Monit Comput       Date:  2015-12-19       Impact factor: 2.502

5.  Does a digital regional nerve block improve the accuracy of noninvasive hemoglobin monitoring?

Authors:  Ronald D Miller; Theresa A Ward; Charles E McCulloch; Neal H Cohen
Journal:  J Anesth       Date:  2012-08-01       Impact factor: 2.078

6.  Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer.

Authors:  N Shah; E A Osea; G J Martinez
Journal:  Int J Lab Hematol       Date:  2013-06-27       Impact factor: 2.877

Review 7.  Continuous Noninvasive Hemoglobin Monitoring: A Measured Response to a Critical Review.

Authors:  Steven J Barker; Aryeh Shander; Michael A Ramsay
Journal:  Anesth Analg       Date:  2016-02       Impact factor: 5.108

8.  Continuous and noninvasive hemoglobin monitoring reduces red blood cell transfusion during neurosurgery: a prospective cohort study.

Authors:  Wael N Awada; Maher F Mohmoued; Tarek M Radwan; Gomaa Z Hussien; Hany W Elkady
Journal:  J Clin Monit Comput       Date:  2015-02-04       Impact factor: 2.502

9.  Utility of non-invasive haemoglobin monitoring in oncosurgery patients.

Authors:  Namrata Gupta; Anita Kulkarni; A K Bhargava; Aditya Prakash; Nitika Gupta
Journal:  Indian J Anaesth       Date:  2017-07

Review 10.  Rationale and limitations of the SpO2/FiO2 as a possible substitute for PaO2/FiO2 in different preclinical and clinical scenarios.

Authors:  Eduardo Butturini de Carvalho; Thiago Ravache Sobreira Leite; Raquel Ferreira de Magalhães Sacramento; Paulo Roberto Loureiro do Nascimento; Cynthia Dos Santos Samary; Patrícia Rieken Macedo Rocco; Pedro Leme Silva
Journal:  Rev Bras Ter Intensiva       Date:  2022 Jan-Mar
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.