Feasibility study of low tube voltage (80 kVp) coronary CT angiography combined with contrast medium reduction using iterative model reconstruction (IMR) on standard BMI patients.

Fan Zhang1, Li Yang2, Xiang Song2, Ying-Na Li2, Yan Jiang3, Xing-Hua Zhang2, Hai-Yue Ju2, Jian Wu2, Rui-Ping Chang2.   

Abstract

OBJECTIVE: To investigate the feasibility of low-tube-voltage (80 kVp) coronary CT angiography (CCTA) combined with contrast medium (CM) reduction and iterative model reconstruction (IMR) on patients with standard body mass index compared with clinical routine protocol.
METHODS: Retrospectively gated helical CCTA scans were acquired using a 256-slice multi-slice CT (Brilliance iCT; Philips Healthcare, Cleveland, OH) on 94 patients with standard body mass index (20-25 kg m(-2)) who were randomly assigned into 2 groups. The scan protocol for Group 1 was 100 kVp and 600 mAs with 70 ml CM at an injection rate of 4.5-5.5 ml s(-1); images were reconstructed by a hybrid iterative reconstruction technique (iDose(4); Philips Healthcare). Group 2 was scanned at 80 kVp and 600 mAs with 35 ml CM at an injection rate of 3.5-4.5 ml s(-1); images were reconstructed with IMR. Objective measurements such as the mean image noise and contrast-to-noise ratio of the two groups were measured on CT images and compared using the paired t-test. In addition, a subjective image quality evaluation was performed by two radiologists who were blinded to the scan protocol, using a 5-point scale [1 (poor) to 5 (excellent)]. The results of the two groups were compared using Mann-Whitney U test.
RESULTS: The iodine delivery rate of Group 2 was 1.0 ± 0.5 gI s(-1) compared with 2.1 ± 0.5 gI s(-1) in Group 1 resulting in a reduction of 52.4%. In addition, an effective radiation dose reduction of 56.4% was achieved in Group 2 (2.4 ± 1.2 mSv) compared with Group 1 (5.5 ± 1.4 mSv). The mean CT attenuation, contrast-to-noise ratio and image quality of all segments in Group 2 were significantly improved compared with those in Group 1 (all, p < 0.01).
CONCLUSION: The use of IMR along with a low tube voltage (80 kVp) combined with a low CM protocol for CCTA can reduce both radiation and CM dose with improved image quality. ADVANCES IN KNOWLEDGE: In this study, we used a novel knowledge-based IMR which remarkably reduced the image noise. We compared the quality of the images obtained when the tube voltage was reduced to 80 kVp and that of those obtained according to the clinical routine protocols to determine whether ultra-low-dose imaging plus IMR is feasible in CCTA scans. We found that a low dose protocol combined with 80 kVp and reduced CM for CCTA can reduce both radiation dose and CM dose with improved image quality by the use of IMR in non-obese patients.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26607646      PMCID: PMC4985220          DOI: 10.1259/bjr.20150766

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  35 in total

Review 1.  Intravenous contrast medium administration and scan timing at CT: considerations and approaches.

Authors:  Kyongtae T Bae
Journal:  Radiology       Date:  2010-07       Impact factor: 11.105

2.  Image quality and radiation dose of low dose coronary CT angiography in obese patients: sinogram affirmed iterative reconstruction versus filtered back projection.

Authors:  Rui Wang; U Joseph Schoepf; Runze Wu; Ryan P Reddy; Chuanchen Zhang; Wei Yu; Yi Liu; Zhaoqi Zhang
Journal:  Eur J Radiol       Date:  2012-05-10       Impact factor: 3.528

3.  Six iterative reconstruction algorithms in brain CT: a phantom study on image quality at different radiation dose levels.

Authors:  A Löve; M-L Olsson; R Siemund; F Stålhammar; I M Björkman-Burtscher; M Söderberg
Journal:  Br J Radiol       Date:  2013-09-18       Impact factor: 3.039

4.  Quantification of coronary artery calcium using ultrafast computed tomography.

Authors:  A S Agatston; W R Janowitz; F J Hildner; N R Zusmer; M Viamonte; R Detrano
Journal:  J Am Coll Cardiol       Date:  1990-03-15       Impact factor: 24.094

Review 5.  Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm.

Authors:  Alvin C Silva; Holly J Lawder; Amy Hara; Jennifer Kujak; William Pavlicek
Journal:  AJR Am J Roentgenol       Date:  2010-01       Impact factor: 3.959

6.  Computed Tomography Angiography With High Flow Rates: An In Vitro and In Vivo Feasibility Study.

Authors:  Casper Mihl; Madeleine Kok; Joachim E Wildberger; Jakub Turek; Georg Muehlenbruch; Marco Das
Journal:  Invest Radiol       Date:  2015-07       Impact factor: 6.016

7.  Diagnostic accuracy of low-dose 256-slice multi-detector coronary CT angiography using iterative reconstruction in patients with suspected coronary artery disease.

Authors:  Yang Hou; Yue Ma; Weipeng Fan; Yuke Wang; Mei Yu; Mani Vembar; Qiyong Guo
Journal:  Eur Radiol       Date:  2013-07-26       Impact factor: 5.315

Review 8.  Systematic review of the clinical effectiveness and cost-effectiveness of 64-slice or higher computed tomography angiography as an alternative to invasive coronary angiography in the investigation of coronary artery disease.

Authors:  G Mowatt; E Cummins; N Waugh; S Walker; J Cook; X Jia; G S Hillis; C Fraser
Journal:  Health Technol Assess       Date:  2008-05       Impact factor: 4.014

9.  Optimizing contrast media injection protocols in state-of-the art computed tomographic angiography.

Authors:  Michael M Lell; Gregor Jost; Johannes Georg Korporaal; Andreas Horst Mahnken; Thomas G Flohr; Michael Uder; Hubertus Pietsch
Journal:  Invest Radiol       Date:  2015-03       Impact factor: 6.016

10.  Contrast agent and radiation dose reduction in abdominal CT by a combination of low tube voltage and advanced image reconstruction algorithms.

Authors:  Nico Buls; Gert Van Gompel; Toon Van Cauteren; Koenraad Nieboer; Inneke Willekens; Guy Verfaillie; Paul Evans; Sven Macholl; Ben Newton; Johan de Mey
Journal:  Eur Radiol       Date:  2014-11-29       Impact factor: 5.315

View more
  12 in total

1.  A low-dose and an ultra-low-dose contrast agent protocol for coronary CT angiography in a clinical setting: quantitative and qualitative comparison to a standard dose protocol.

Authors:  Dominik C Benz; Christoph Gräni; Beatrice Hirt Moch; Fran Mikulicic; Jan Vontobel; Tobias A Fuchs; Julia Stehli; Olivier F Clerc; Mathias Possner; Aju P Pazhenkottil; Oliver Gaemperli; Ronny R Buechel; Philipp A Kaufmann
Journal:  Br J Radiol       Date:  2017-05-25       Impact factor: 3.039

2.  Nothing endures but change.

Authors:  Simon A Jackson; Kevin M Prise
Journal:  Br J Radiol       Date:  2017-01       Impact factor: 3.039

Review 3.  Computed tomography coronary angiography - past, present and future.

Authors:  Pei Ing Ngam; Ching Ching Ong; Ping Chai; Siong Sung Wong; Chong Ri Liang; Lynette Li San Teo
Journal:  Singapore Med J       Date:  2020-03       Impact factor: 1.858

Review 4.  The role of advanced reconstruction algorithms in cardiac CT.

Authors:  Sandra S Halliburton; Yuki Tanabe; Sasan Partovi; Prabhakar Rajiah
Journal:  Cardiovasc Diagn Ther       Date:  2017-10

5.  Contrast dose reduction with shortened injection durations in coronary CT angiography on 16-cm Wide-detector CT scanner.

Authors:  Ting-Ting Qu; Jian-Ying Li; Xi-Jun Jiao; Xiang-Li Zhang; Zhe-Fan Song; Yin-Xia Guo; Jian-Xin Guo
Journal:  Br J Radiol       Date:  2018-09-04       Impact factor: 3.039

6.  Improved image quality of low-dose CT combining with iterative model reconstruction algorithm for response assessment in patients after treatment of malignant tumor.

Authors:  Xiaoyan Xin; Jingtao Shen; Shangwen Yang; Song Liu; Anning Hu; Bin Zhu; Yan Jiang; Baoxin Li; Bing Zhang
Journal:  Quant Imaging Med Surg       Date:  2018-08

7.  Impact of iterative model reconstruction combined with dose reduction on the image quality of head and neck CTA in children.

Authors:  Bochao Cheng; Haoyang Xing; Du Lei; Yingkun Guo; Gang Ning; Qiyong Gong; Wu Cai
Journal:  Sci Rep       Date:  2018-08-22       Impact factor: 4.379

8.  CT Pulmonary Angiography at Reduced Radiation Exposure and Contrast Material Volume Using Iterative Model Reconstruction and iDose4 Technique in Comparison to FBP.

Authors:  Azien Laqmani; Maximillian Kurfürst; Sebastian Butscheidt; Susanne Sehner; Jakob Schmidt-Holtz; Cyrus Behzadi; Hans Dieter Nagel; Gerhard Adam; Marc Regier
Journal:  PLoS One       Date:  2016-09-09       Impact factor: 3.240

9.  Chest CT with iterative reconstruction algorithms for airway stent evaluation in patients with malignant obstructive tracheobronchial diseases.

Authors:  Tingting Li; Yonggao Zhang; Yadong Wang; Jianbo Gao; Yan Jiang
Journal:  Medicine (Baltimore)       Date:  2016-09       Impact factor: 1.889

10.  Application of multiple injections of contrast agent in head and neck CT arteriovenous angiography in children: Study protocol clinical trial (SPIRIT Compliant).

Authors:  Shilong Tang; Guanping Zhang; Zhuo Chen; Xianfan Liu; Xiao Fan; Daisong Liu; Ling He
Journal:  Medicine (Baltimore)       Date:  2020-04       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.