Computed Tomography Angiography With High Flow Rates: An In Vitro and In Vivo Feasibility Study.

Casper Mihl1, Madeleine Kok, Joachim E Wildberger, Jakub Turek, Georg Muehlenbruch, Marco Das.   

Abstract

OBJECTIVE: The aims of this study were to test high-flow application of contrast media (CM) using novel high-flow needles and to assess injection- and flow-related parameters in a circulation phantom and in an in vivo population.
MATERIALS AND METHODS: A circulation phantom simulating physiological parameters was used. Preheated CM (300 mg/mL) was injected at flow rates varying between 5 and 15 mL/s through a novel 18-gauge high-flow intravenous injection needle. In addition, feasibility of these high-flow needles was tested with administration of flow rates of 9 mL/s in 20 patients referred for pre-transcatheter aortic valve implantation assessment. Injection parameters (eg, peak pressures, peak flow rates) in both phantom and in vivo setup were continuously monitored by a data acquisition program. Attenuation at predefined levels of the aorta (eg, aortic root to common femoral arteries) was measured in all patients to determine clinical applicability.
RESULTS: In the phantom setup, injection rates up to 15 mL/s were feasible. An enhancement plateau was reached at 11 mL/s (464 [20] HU). In patients, no pressure- or flow-related complications (eg, extravasation) were recorded (mean [SD] peak pressure, 154 [8] psi; mean [SD] peak flow rate, 9.2 [0.1 mL/s; range, 9.1-9.6]). Diagnostic attenuation values were reached at all predefined levels of the aorta (330.8 [113.1] HU to 622.9 [81.5] HU).
CONCLUSIONS: These results indicate that injections with 9 mL/s using high-flow injection needles are safe. The pressure limit of 325 psi was not reached, and the injections resulted in diagnostic attenuation values. Using this dedicated needle, high flow rates should not be considered a drawback for CM application in routine CT angiography examinations.

Entities:  

Mesh:

Year:  2015        PMID: 25816214     DOI: 10.1097/RLI.0000000000000153

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  7 in total

1.  Feasibility study of low tube voltage (80 kVp) coronary CT angiography combined with contrast medium reduction using iterative model reconstruction (IMR) on standard BMI patients.

Authors:  Fan Zhang; Li Yang; Xiang Song; Ying-Na Li; Yan Jiang; Xing-Hua Zhang; Hai-Yue Ju; Jian Wu; Rui-Ping Chang
Journal:  Br J Radiol       Date:  2015-11-26       Impact factor: 3.039

2.  Contrast media extravasations in patients undergoing computerized tomography scanning: a systematic review and meta-analysis of risk factors and interventions.

Authors:  Sandrine Ding; Nicole Richli Meystre; Cosmin Campeanu; Giuseppe Gullo
Journal:  JBI Database System Rev Implement Rep       Date:  2018-01

3.  Relationship between low tube voltage (70 kV) and the iodine delivery rate (IDR) in CT angiography: An experimental in-vivo study.

Authors:  Michael M Lell; Ulrike Fleischmann; Hubertus Pietsch; Johannes G Korporaal; Ulrike Haberland; Andreas H Mahnken; Thomas G Flohr; Michael Uder; Gregor Jost
Journal:  PLoS One       Date:  2017-03-20       Impact factor: 3.240

4.  Impact of CT Injector Technology and Contrast Media Viscosity on Vascular Enhancement: Evaluation in a Circulation Phantom.

Authors:  Michael McDermott; Corey Kemper; William Barone; Gregor Jost; Jan Endrikat
Journal:  Br J Radiol       Date:  2020-02-20       Impact factor: 3.039

5.  Impact of iterative model reconstruction combined with dose reduction on the image quality of head and neck CTA in children.

Authors:  Bochao Cheng; Haoyang Xing; Du Lei; Yingkun Guo; Gang Ning; Qiyong Gong; Wu Cai
Journal:  Sci Rep       Date:  2018-08-22       Impact factor: 4.379

6.  Personalization of CM Injection Protocols in Coronary Computed Tomographic Angiography (People CT Trial).

Authors:  N G Eijsvoogel; B M F Hendriks; P Nelemans; C Mihl; J Willigers; B Martens; J E Wildberger; M Das
Journal:  Contrast Media Mol Imaging       Date:  2020-01-17       Impact factor: 3.161

7.  Personalized application of three different concentrations of iodinated contrast media in coronary computed tomography angiography.

Authors:  Meng Zhang; Panpan Hao; Chenyu Jiang; Guoxiang Hao; Bin Li; Peixin Hu; Qingjie Chen; Yuguo Chen; Aifeng Zhang; Yun Zhang; Yanping Liu
Journal:  J Cell Mol Med       Date:  2020-03-30       Impact factor: 5.310

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.