Optimizing contrast media injection protocols in state-of-the art computed tomographic angiography.

Michael M Lell1, Gregor Jost, Johannes Georg Korporaal, Andreas Horst Mahnken, Thomas G Flohr, Michael Uder, Hubertus Pietsch.   

Abstract

OBJECTIVE: Very short acquisition times and the increasing use of low-kilovolt protocols in standard computed tomographic (CT) angiography (CTA) examinations demand modifications in the contrast media (CM) injection regimen. The aim of this study was to optimize the use of tube voltage-adjusted CM delivery parameters, especially injection duration and iodine delivery rate (IDR), in thoracoabdominal CTA in a porcine model.
MATERIALS AND METHODS: Eight pigs (53-72 kg) were examined with a third-generation dual-source CT system with a dynamic CTA protocol (4-dimensional spiral, 454-mm scan length, 2.5-second temporal resolution, 70-second total acquisition time). Six CM injection protocols were applied in randomized order and intraindividually compared. The standard CTA protocol was performed at 120 kV, with an injection of 300 mg iodine/kg body weight and a flow of 5 mL/s (IDR, 1.5 g/s). On the basis of phantom measurements for the low-kilovolt CTA protocols, the iodine dosage was adjusted to 150 mg iodine/kg (70 kV) and 210 mg iodine/kg (90 kV). Therefore, either the IDR was kept constant and the injection time was reduced, or the injection time was kept constant and the IDR was reduced by modifying the CM flow or concentration. Time attenuation curves, time to peak, and peak enhancement were calculated for different locations within the aorta, renal arteries, and large veins.
RESULTS: The heart rates were comparable among the different injection protocols (66.9-78.1 beats per minute). The average injection peak pressure depended on the flow rate and CM concentration and ranged from 42.9 to 114.7 psi. The average arterial peak enhancement was comparable for protocols with identical injection times and reduced IDR (362.4 HU [standard 120-kV protocol; 300 mg iodine/kg; IDR, 1.5 g/s], 360.0 HU [70 kV; 150 mg iodine/kg; IDR, 0.75 g/s], 365.4 HU [70 kV; 150 mg iodine/kg; IDR, 0.75 g/s; CM, 150 mg iodine/mL], 344.3 HU [90 kV; 210 mg iodine/kg; IDR, 1.1 g/s]). Higher peak enhancements could be achieved by applying protocols with identical IDR and a reduced injection time (502.5 HU [70 kV; 150 mg iodine/kg; IDR, 1.5 g/s] and 394.6 HU [90 kV; 210 mg iodine/kg; IDR, 1.5 g/s]).
CONCLUSIONS: By adjusting the IDR, low-kilovolt CTA is able to achieve comparable aortic enhancement with a significant reduction in CM dosage. A shorter injection time at constant IDR results in higher enhancement and a narrower scan window and might be preferable for fast CTA acquisition techniques. CLINICAL RELEVANCE/APPLICATION: The optimization of CM injection protocols is mandatory to achieve state-of the art CTA at low kilovolt and can reduce CM doses to patients.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25478743     DOI: 10.1097/RLI.0000000000000119

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  23 in total

1.  Feasibility study of low tube voltage (80 kVp) coronary CT angiography combined with contrast medium reduction using iterative model reconstruction (IMR) on standard BMI patients.

Authors:  Fan Zhang; Li Yang; Xiang Song; Ying-Na Li; Yan Jiang; Xing-Hua Zhang; Hai-Yue Ju; Jian Wu; Rui-Ping Chang
Journal:  Br J Radiol       Date:  2015-11-26       Impact factor: 3.039

Review 2.  Contrast medium administration and image acquisition parameters in renal CT angiography: what radiologists need to know.

Authors:  Charbel Saade; Ibrahim Alsheikh Deeb; Maha Mohamad; Hussain Al-Mohiy; Fadi El-Merhi
Journal:  Diagn Interv Radiol       Date:  2016 Mar-Apr       Impact factor: 2.630

3.  Coronary CT Angiography Using Low Iodine Delivery Rate and Tube Voltage Determined by Body Mass Index: Superiority in Clinical Practice.

Authors:  Wang Yuan; Ting-Ting Qu; Hui-Juan Wang; Mei-Yu Wang; Yuan Qu; Gang Niu; Jian Yang
Journal:  Curr Med Sci       Date:  2019-10-14

4.  Filtered back projection revisited in low-kilovolt computed tomography angiography: sharp filter kernel enhances visualization of the artery of Adamkiewicz.

Authors:  Tatsuya Nishii; Akiyuki Kotoku; Yoshiro Hori; Yuji Matsuzaki; Yoshiaki Watanabe; Yoshiaki Morita; Atsushi K Kono; Hitoshi Matsuda; Tetsuya Fukuda
Journal:  Neuroradiology       Date:  2018-11-21       Impact factor: 2.804

5.  A prospective evaluation of contrast and radiation dose and image quality in cardiac CT in children with complex congenital heart disease using low-concentration iodinated contrast agent and low tube voltage and current.

Authors:  Qiao-Ru Hou; Wei Gao; Ai-Min Sun; Qian Wang; Hai-Sheng Qiu; Fang Wang; Li-Wei Hu; Jian-Ying Li; Yu-Min Zhong
Journal:  Br J Radiol       Date:  2016-12-07       Impact factor: 3.039

6.  Reaching for better image quality and lower radiation dose in head and neck CT: advanced modeled and sinogram-affirmed iterative reconstruction in combination with tube voltage adaptation.

Authors:  Andrea I Schmid; Michael Uder; Michael M Lell
Journal:  Dentomaxillofac Radiol       Date:  2016-09-08       Impact factor: 2.419

7.  Comparison of virtual monoenergetic and polyenergetic images reconstructed from dual-layer detector CT angiography of the head and neck.

Authors:  Victor Neuhaus; Nils Große Hokamp; Nuran Abdullayev; Volker Maus; Christoph Kabbasch; Anastasios Mpotsaris; David Maintz; Jan Borggrefe
Journal:  Eur Radiol       Date:  2017-10-10       Impact factor: 5.315

8.  Advantages of 70-kV CT Angiography for the Visualization of the Adamkiewicz Artery: Comparison with 120-kV Imaging.

Authors:  S Shimoyama; T Nishii; Y Watanabe; A K Kono; K Kagawa; S Takahashi; K Sugimura
Journal:  AJNR Am J Neuroradiol       Date:  2017-09-14       Impact factor: 3.825

Review 9.  Radiation dose management for pediatric cardiac computed tomography: a report from the Image Gently 'Have-A-Heart' campaign.

Authors:  Cynthia K Rigsby; Sarah E McKenney; Kevin D Hill; Anjali Chelliah; Andrew J Einstein; B Kelly Han; Joshua D Robinson; Christina L Sammet; Timothy C Slesnick; Donald P Frush
Journal:  Pediatr Radiol       Date:  2018-01-01

10.  Contrast media injection protocol optimization for dual-energy coronary CT angiography: results from a circulation phantom.

Authors:  Domenico De Santis; Damiano Caruso; U Joseph Schoepf; Marwen Eid; Moritz H Albrecht; Taylor M Duguay; Akos Varga-Szemes; Andrea Laghi; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2018-02-27       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.