Accuracy of templating the acetabular cup size in Total Hip Replacement using conventional acetate templates on digital radiographs.

Vignesh P Krishnamoorthy1, Rajamani Perumal2, Alfred J Daniel3, Pradeep M Poonnoose3.   

Abstract

OBJECTIVE: Templating of the acetabular cup size in Total Hip Replacement (THR) is normally done using conventional radiographs. As these are being replaced by digital radiographs, it has become essential to create a technique of templating using digital films. We describe a technique that involves templating the digital films using the universally available acetate templates for THR without the use of special software.
MATERIALS AND METHODS: Preoperative digital radiographs of the pelvis were taken with a 30 mm diameter spherical metal ball strapped over the greater trochanter. Using standard acetate templates provided by the implant company on magnified digital radiographs, the size of the metal ball (X mm) and acetabular cup (Y mm) were determined. The size of the acetabular cup to be implanted was estimated using the formula 30*Y/X. The estimated size was compared with the actual size of the cup used at surgery.
RESULTS: Using this technique, it was possible to accurately predict the acetabular cup size in 28/40 (70%) of the hips. When the accuracy to within one size was considered, templating was correct in 90% (36/40). When assessed by two independent observers, there was good intra-observer and inter-observer reliability with intra-class correlation coefficient values greater than 0.8.
CONCLUSION: It was possible to accurately and reliably predict the size of the acetabular cup, using acetate templates on digital films, without any digital templates.

Entities:  

Keywords:  Acetabular cup; Acetate templates; Radiographs; Templating; Total Hip Replacement

Year:  2015        PMID: 26566332      PMCID: PMC4600879          DOI: 10.1016/j.jcot.2015.04.001

Source DB:  PubMed          Journal:  J Clin Orthop Trauma        ISSN: 0976-5662


  8 in total

1.  Accuracy of digital templating in total hip arthroplasty.

Authors:  David R Whiddon; James V Bono; Jason E Lang; Eric L Smith; Aaron K Salyapongse
Journal:  Am J Orthop (Belle Mead NJ)       Date:  2011-08

Review 2.  Preoperative planning for primary total hip arthroplasty.

Authors:  Alejandro González Della Valle; Douglas E Padgett; Eduardo A Salvati
Journal:  J Am Acad Orthop Surg       Date:  2005-11       Impact factor: 3.020

3.  Digital templating in total hip arthroplasty.

Authors:  James V Bono
Journal:  J Bone Joint Surg Am       Date:  2004       Impact factor: 5.284

4.  The value of preoperative planning for total hip arthroplasty.

Authors:  S Eggli; M Pisan; M E Müller
Journal:  J Bone Joint Surg Br       Date:  1998-05

5.  Accuracy of digital preoperative templating in 100 consecutive uncemented total hip arthroplasties: a single surgeon series.

Authors:  Shahril R Shaarani; Gavin McHugh; Denis A Collins
Journal:  J Arthroplasty       Date:  2012-07-31       Impact factor: 4.757

6.  The accuracy of digital templating in uncemented total hip arthroplasty.

Authors:  Patrick Gamble; Justin de Beer; Danielle Petruccelli; Mitchell Winemaker
Journal:  J Arthroplasty       Date:  2009-06-02       Impact factor: 4.757

7.  Determination of accuracy of preoperative templating of noncemented femoral prostheses.

Authors:  L W Carter; D O Stovall; T R Young
Journal:  J Arthroplasty       Date:  1995-08       Impact factor: 4.757

8.  The accuracy of preoperative templating in cementless total hip arthroplasty.

Authors:  Aasis Unnanuntana; Dominic Wagner; Stuart B Goodman
Journal:  J Arthroplasty       Date:  2008-04-08       Impact factor: 4.757

  8 in total
  3 in total

1.  Accuracy and Reliability of Preoperative On-screen Templating Using Digital Radiographs for Total Hip Arthroplasty.

Authors:  Jong Ki Shin; Seung Min Son; Tae Woo Kim; Won Chul Shin; Jung Sub Lee; Kuen Tak Suh
Journal:  Hip Pelvis       Date:  2016-12-28

2.  A Novel Method of Determining Acetabular Component Size to Guide Explant in Revision Hip Arthroplasty.

Authors:  Fiachra O Deasmhunaigh; Jeremy Abetz; Shaun English; David Mitchell; Biswadev Mitra
Journal:  Arthroplast Today       Date:  2020-04-24

3.  Use of Acetate Templates Over Digital Radiographs for Templating in Total Hip Arthroplasty: Technique and Its Validation.

Authors:  Tarun Goyal; Siddharth S Sethy; Sajid Ansari; Lakshmana Das; Souvik Paul
Journal:  Indian J Orthop       Date:  2020-11-03       Impact factor: 1.251

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.