Two-dimensional ionization chamber arrays for IMRT plan verification.

Bjoern Poppe1, Arne Blechschmidt, Armand Djouguela, Ralf Kollhoff, Antje Rubach, Kay C Willborn, Dietrich Harder.   

Abstract

In this paper we describe a concept for dosimetric treatment plan verification using two-dimensional ionization chamber arrays. Two different versions of the 2D-ARRAY (PTW-Freiburg, Germany) will be presented, a matrix of 16 x 16 chambers (chamber cross section 8 mm x 8 mm; the distance between chamber centers, 16 mm) and a matrix of 27 x 27 chambers (chamber cross section 5 mm x 5 mm; the distance between chamber centers is 10 mm). The two-dimensional response function of a single chamber is experimentally determined by scanning it with a slit beam. For dosimetric plan verification, the expected two-dimensional distribution of the array signals is calculated via convolution of the planned dose distribution, obtained from the treatment planning system, with the two-dimensional response function of a single chamber. By comparing the measured two-dimensional distribution of the array signals with the expected one, a distribution of deviations is obtained that can be subjected to verification criteria, such as the gamma index criterion. As an example, this verification method is discussed for one sequence of an IMRT plan. The error detection capability is demonstrated in a case study. Both versions of two-dimensional ionization chamber arrays, together with the developed treatment plan verification strategy, have been found to provide a suitable and easy-to-handle quality assurance instrument for IMRT.

Mesh:

Year:  2006        PMID: 16696477     DOI: 10.1118/1.2179167

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  43 in total

1.  Practical guidelines for routine intensity-modulated radiotherapy verification: pre-treatment verification with portal dosimetry and treatment verification with in vivo dosimetry.

Authors:  A J Vinall; A J Williams; V E Currie; A Van Esch; D Huyskens
Journal:  Br J Radiol       Date:  2010-11       Impact factor: 3.039

2.  Quasi-IMAT study with conventional equipment to show high plan quality with a single gantry arc.

Authors:  Judith Alvarez Moret; Oliver Kölbl; Ludwig Bogner
Journal:  Strahlenther Onkol       Date:  2009-02-18       Impact factor: 3.621

3.  Simulation for improvement of system sensitivity of radiochromic film dosimetry with different band-pass filters and scanner light intensities.

Authors:  Takeshi Kamomae; Yuki Miyabe; Akira Sawada; Osamu Matoba; Manabu Nakata; Shinsuke Yano; Toru Takakura; Takashi Mizowaki; Akio Itoh; Masahiro Hiraoka
Journal:  Radiol Phys Technol       Date:  2011-03-17

4.  The effect of MLC speed and acceleration on the plan delivery accuracy of VMAT.

Authors:  J M Park; H-G Wu; J H Kim; J N K Carlson; K Kim
Journal:  Br J Radiol       Date:  2015-03-03       Impact factor: 3.039

5.  Video-rate optical dosimetry and dynamic visualization of IMRT and VMAT treatment plans in water using Cherenkov radiation.

Authors:  Adam K Glaser; Jacqueline M Andreozzi; Scott C Davis; Rongxiao Zhang; Brian W Pogue; Colleen J Fox; David J Gladstone
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

6.  Performance characterization of a 3D liquid scintillation detector for discrete spot scanning proton beam systems.

Authors:  Chinmay D Darne; Fahed Alsanea; Daniel G Robertson; Narayan Sahoo; Sam Beddar
Journal:  Phys Med Biol       Date:  2017-06-23       Impact factor: 3.609

7.  Improvement of I'mRT MatriXX in terms of spatial resolution and large area acquisition for patient-specific intensity-modulated radiotherapy verification.

Authors:  Arun S Oinam; Lakhwant Singh; S C Sharma; Pradeep Goswami
Journal:  J Med Phys       Date:  2009-07

8.  Dosimetric properties and clinical application of an a-Si EPID for dynamic IMRT quality assurance.

Authors:  Kenji Matsumoto; Masahiko Okumura; Yoshiyuki Asai; Kouhei Shimomura; Masaya Tamura; Yasumasa Nishimura
Journal:  Radiol Phys Technol       Date:  2012-12-04

9.  Dosimetric impact of different multileaf collimators on prostate intensity modulated treatment planning.

Authors:  Lucia Clara Orlandini; Margherita Betti; Christian Fulcheri; Marianna Coppola; Luca Cionini
Journal:  Rep Pract Oncol Radiother       Date:  2015-09-29

10.  Sensitivity of the IQM and MatriXX detectors in megavolt photon beams.

Authors:  Oluwaseyi Michael Oderinde; Freek du Plessis
Journal:  Rep Pract Oncol Radiother       Date:  2019-08-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.