| Literature DB >> 26445499 |
Abimbola A Akintola1, Raymond Noordam1, Steffy W Jansen1, Anton J de Craen1, Bart E Ballieux2, Christa M Cobbaert2, Simon P Mooijaart1, Hanno Pijl3, Rudi G Westendorp4, Diana van Heemst1.
Abstract
BACKGROUND: The validity of continuous glucose monitoring (CGM) is well established in diabetic patients. CGM is also increasingly used for research purposes in normo-glycemic individuals, but the CGM validity in such individuals is unknown. We studied the accuracy of CGM measurements in normo-glycemic individuals by comparing CGM-derived versus venous blood-derived glucose levels and measures of glycemia and glycemic variability.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26445499 PMCID: PMC4596806 DOI: 10.1371/journal.pone.0139973
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Characteristics of the study population.
| Demographics | N = 34 |
|---|---|
| Female, n (%) | 15 (44.1) |
| Age, years | 65.7 (4.8) |
| Weight, kg | 75.5 (13.3) |
| BMI, kg/m2 | 25.2 (3.9) |
| Waist circumference, cm | 93.1 (12.2) |
| Waist: hip ratio | 0.90 (0.09) |
| Fat mass, percentage | 30.8 (8.04) |
| Total lean mass, kg | 52.3 (11.6) |
| Fasted (venous) glucose, mmol/L | 4.9 (0.6) |
Data represent mean with standard deviation unless stated otherwise.
Comparison of estimates of glycemia and glycemic variability.
| Mean CGM | Mean venous | Difference (SE) | P-value | |
|---|---|---|---|---|
|
| ||||
| 24h glucose | 5.18 (0.46) | 5.10 (0.35) | 0.08 (0.08) | 0.35 |
| Daytime glucose (09.00h–23.00h) | 5.61 (0.52) | 5.39 (0.43) | 0.22 (0.09) | 0.02 |
| Nighttime glucose (23.00h–09.00h) | 4.57 (0.62) | 4.68 (0.47) | -0.11 (0.12) | 0.33 |
|
| ||||
| 24h standard deviation | 1.07 (0.29) | 1.26 (0.36) | -0.19 (0.06) | 0.004 |
| SD | 0.47 (0.09) | 0.68 (0.15) | -0.21 (0.02) | <0.001 |
| SD | 0.72 (0.18) | 0.95 (0.26) | -0.23 (0.04) | <0.001 |
| Range | 4.45 (1.19) | 6.24 (1.72) | -1.79 (0.25) | <0.001 |
| IQR (median (IQR)) | 1.24 (1.05–1.42) | 1.27 (1.00–1.46) | 0.354 | |
| % Coefficient of Variation | 20.9 (6.04) | 24.3 (6.67) | -3.39 (1.20) | 0.008 |
| MAGE | 3.02 (1.02) | 3.52 (1.36) | -0.50 (0.22) | 0.034 |
| CONGA1 | 1.50 (0.42) | 1.97 (0.61) | -0.47 (0.07) | <0.001 |
| CONGA4 | 1.57 (0.51) | 2.07 (0.66) | -0.50 (0.11) | <0.001 |
|
| ||||
| J- index | 0.039 (0.007) | 0.041 (0.007) | 0.002 (0.007) | 0.341 |
| Hypoglycemia index (median (IQR)) | 0.46 (0.07–2.0) | 1.33 (0.35–2.15) | 0.17 | |
| Hyperglycemia index (median (IQR)) | 0.003 (0.0–0.04) | 0.031 (0.01–0.08) | <0.001 |
Values are mean (SD) unless otherwise indicated. Abbreviations: CGM, continuous glucose monitoring; SD, standard deviation; SE, standard error; SDws1and SDws4, standard deviation within time series of respectively 1 and 4 hours; CONGA1 and CONGA4, continuous overlapping net glycemic action over respectively 1 and 4 hours; and MAGE, mean amplitude of glycemic excursions.