Literature DB >> 26407614

Resurfacing head size and femoral fracture: Are registry conclusions on head size justified?

Adeel Aqil1, Anatole Wiik2, Susannah Clarke2, Milad Masjedi2, Justin Cobb2.   

Abstract

BACKGROUND: Joint registries report that peri-prosthetic fractures are the most common reason for early revision of a hip resurfacing arthroplasty (HRA) and are twice as likely with small implant sizes. However, a national survey found peri-prosthetic fracture to be strongly associated with surgical accuracy. We therefore asked whether the force required to induce a peri-prosthetic fracture: (1) was significantly lower when using smaller implants and (2) correlated to the size of implant used, when surgery was performed accurately.
METHODS: To ensure an adequate power, we calculated our sample size from pilot data. Forty-four femurs were tested in two experiments. The first experiment tested femurs with either a small (48 mm) or a large (54 mm) HRA implant. The second involved testing femurs with a range of implant sizes. A rapid prototyped femur-specific guide ensured accurate implantation. Specimens were then vertically loaded in a servo-hydraulic testing machine till fracture. Displacement (mm) and force (N) required for fracture were recorded.
RESULTS: A median force of 1081 N was required to fracture specimens implanted with small 48-mm heads, while 1134 N was required when a 54-mm head was used (U = 77, z = -0.054, p = 0.957). Implant head size and force required to fracture were not related, r = 0.12, p = 0.63.
CONCLUSIONS: The force required to induce a resurfacing peri-prosthetic fracture was not related to the size of the implant. The increased failure rate seen in all registries is unlikely to be directly the result of this single variable. Correctly performed resurfacing arthroplasty is highly resistant to fracture.

Entities:  

Keywords:  Failure; Femoral fracture; Head size; Resurfacing

Mesh:

Year:  2015        PMID: 26407614     DOI: 10.1007/s00590-015-1704-4

Source DB:  PubMed          Journal:  Eur J Orthop Surg Traumatol        ISSN: 1633-8065


  12 in total

1.  Structural properties of a new design of composite replicate femurs and tibias.

Authors:  A D Heiner; T D Brown
Journal:  J Biomech       Date:  2001-06       Impact factor: 2.712

2.  Reduced articular surface of one-piece cups: a cause of runaway wear and early failure.

Authors:  William L Griffin; Christopher J Nanson; Bryan D Springer; Matthew A Davies; Thomas K Fehring
Journal:  Clin Orthop Relat Res       Date:  2010-09       Impact factor: 4.176

3.  Cancellous bone screw purchase: a comparison of synthetic femurs, human femurs, and finite element analysis.

Authors:  R Zdero; M Olsen; H Bougherara; E H Schemitsch
Journal:  Proc Inst Mech Eng H       Date:  2008-11       Impact factor: 1.617

4.  Femoral neck fracture following hip resurfacing: the effect of alignment of the femoral component.

Authors:  E T Davis; M Olsen; R Zdero; J P Waddell; E H Schemitsch
Journal:  J Bone Joint Surg Br       Date:  2008-11

5.  Role of surgical position on interface stress and initial bone remodeling stimulus around hip resurfacing arthroplasty.

Authors:  Kevin L Ong; Judd S Day; Steven M Kurtz; Richard E Field; Michael T Manley
Journal:  J Arthroplasty       Date:  2008-09-26       Impact factor: 4.757

6.  Femoral neck fractures following Birmingham hip resurfacing: a national review of 50 cases.

Authors:  A J Shimmin; D Back
Journal:  J Bone Joint Surg Br       Date:  2005-04

7.  The gait of patients with one resurfacing and one replacement hip: a single blinded controlled study.

Authors:  Adeel Aqil; Roshan Drabu; Jeroen H Bergmann; Milad Masjedi; Victoria Manning; Barry Andrews; Sarah K Muirhead-Allwood; Justin P Cobb
Journal:  Int Orthop       Date:  2013-02-27       Impact factor: 3.075

8.  Verification of the Australian Orthopaedic Association National Joint Replacement Registry using a surgeon's database.

Authors:  Bob Jang; Nichola A Walsh; Warwick J M Bruce
Journal:  J Orthop Surg (Hong Kong)       Date:  2013-12       Impact factor: 1.118

9.  Mechanical evaluation of large-size fourth-generation composite femur and tibia models.

Authors:  Matthew P Gardner; Alexander C M Chong; Anthony G Pollock; Paul H Wooley
Journal:  Ann Biomed Eng       Date:  2010-01-05       Impact factor: 3.934

10.  Mortality and implant revision rates of hip arthroplasty in patients with osteoarthritis: registry based cohort study.

Authors:  D J W McMinn; K I E Snell; J Daniel; R B C Treacy; P B Pynsent; R D Riley
Journal:  BMJ       Date:  2012-06-14
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.