Performance of NiTi endodontic instrument under different temperatures.

Ahmed Jamleh1, Yoshio Yahata2, Arata Ebihara3, Amre R Atmeh4, Turki Bakhsh5, Hideaki Suda6.   

Abstract

The purpose of this study was to test nickel titanium (NiTi) instrument performance under different surrounding temperatures. Twenty-four superelastic NiTi instruments with a conical shape comprising a 0.30-mm-diameter tip and 0.06 taper were equally divided into 3 groups according to the temperature employed. Using a specially designed cyclic fatigue testing apparatus, each instrument was deflected to give a curvature 10 mm in radius and a 30° angle. This position was kept as the instrument was immersed in a continuous flow of water under a temperature of 10, 37, or 50 °C for 20 s to calculate the deflecting load (DL). In the same position, the instrument was then allowed to rotate at 300 rpm to fracture, and the working time was converted to the number of cycles to fracture (NCF). The statistical significance was set at p = 0.05. The mean DL (in N) and NCF (in cycles) of the groups at 10, 37, and 50 °C were 10.16 ± 1.36 and 135.50 ± 31.48, 13.50 ± 0.92 and 89.20 ± 16.44, and 14.70 ± 1.21 and 65.50 ± 15.90, respectively. The group at 10 °C had significantly the lowest DL that favorably resulted in the highest NCF. Within the limitations of this study, the surrounding temperature influences the cyclic fatigue resistance and DL of the superelastic NiTi instruments. Lower temperatures are found to favorably decrease the DL and extend the lifetime of the superelastic NiTi instrument. Further NiTi instrument failure studies should be performed under simulated body temperature.

Entities:  

Keywords:  Deflecting load; Number of cycles to failure; Superelastic NiTi instrument; Temperature

Mesh:

Substances:

Year:  2015        PMID: 26296811     DOI: 10.1007/s10266-015-0214-5

Source DB:  PubMed          Journal:  Odontology        ISSN: 1618-1247            Impact factor:   2.634


  32 in total

1.  Impact of two theoretical cross-sections on torsional and bending stresses of nickel-titanium root canal instrument models.

Authors:  Y L Turpin; F Chagneau; J M Vulcain
Journal:  J Endod       Date:  2000-07       Impact factor: 4.171

2.  A comparison of stainless steel and nickel-titanium H-type instruments of identical design: torsional and bending tests.

Authors:  R B Kazemi; E Stenman; L S Spångberg
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2000-10

3.  Influence of structure on nickel-titanium endodontic instruments failure.

Authors:  G Kuhn; B Tavernier; L Jordan
Journal:  J Endod       Date:  2001-08       Impact factor: 4.171

4.  Predicting in vivo failure of pseudoelastic NiTi devices under low cycle, high amplitude fatigue.

Authors:  Jeremy M Young; Krystyn J Van Vliet
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2005-01-15       Impact factor: 3.368

5.  Factors influencing defects of rotary nickel-titanium endodontic instruments after clinical use.

Authors:  Peter Parashos; Ian Gordon; Harold H Messer
Journal:  J Endod       Date:  2004-10       Impact factor: 4.171

6.  Vickers hardness investigation of work-hardening in used NiTi rotary instruments.

Authors:  Satish B Alapati; William A Brantley; John M Nusstein; Glenn S Daehn; Timothy A Svec; John M Powers; William M Johnston; Wenhua Guo
Journal:  J Endod       Date:  2006-10-13       Impact factor: 4.171

7.  Fatigue resistance of engine-driven rotary nickel-titanium instruments produced by new manufacturing methods.

Authors:  Gianluca Gambarini; Nicola Maria Grande; Gianluca Plotino; Francesco Somma; Manish Garala; Massimo De Luca; Luca Testarelli
Journal:  J Endod       Date:  2008-06-25       Impact factor: 4.171

8.  Cyclic fatigue resistance of ProTaper Next nickel-titanium rotary files.

Authors:  A M Elnaghy
Journal:  Int Endod J       Date:  2014-02-01       Impact factor: 5.264

9.  Phase transformation behavior and mechanical properties of thermomechanically treated K3XF nickel-titanium instruments.

Authors:  Ya Shen; Hui-Min Zhou; Zhejun Wang; Les Campbell; Yu-feng Zheng; Markus Haapasalo
Journal:  J Endod       Date:  2013-05-16       Impact factor: 4.171

10.  The effects of temperature on sodium hypochlorite short-term stability, pulp dissolution capacity, and antimicrobial efficacy.

Authors:  George Sirtes; Tuomas Waltimo; Marc Schaetzle; Matthias Zehnder
Journal:  J Endod       Date:  2005-09       Impact factor: 4.171

View more
  8 in total

1.  Cyclic fatigue resistance of XP-endo Shaper compared with different nickel-titanium alloy instruments.

Authors:  Amr Elnaghy; Shaymaa Elsaka
Journal:  Clin Oral Investig       Date:  2017-10-13       Impact factor: 3.573

2.  Resistance to Cyclic Fatigue of Nickel-Titanium Files Immersed in Sodium Hypochlorite at Body Temperature.

Authors:  Hind F Abuhulaibah; Ammar AbuMostafa
Journal:  Int J Dent       Date:  2020-09-19

3.  Influence of different motions on the cyclic fatigue resistance of Reciproc and Reciproc Blue endodontic instruments.

Authors:  Marco Serafin; Matteo De Biasi; Vittorio Franco; Luigi Generali; Daniele Angerame
Journal:  J Conserv Dent       Date:  2020-08-04

4.  Effect of NaOCl and EDTA irrigating solutions on the cyclic fatigue resistance of EdgeTaper Platinum instruments.

Authors:  Hussam Alfawaz; Abdullah Alqedairi; Maha Alhamdan; Najla Alkhzim; Shatha Alfarraj; Ahmed Jamleh
Journal:  BMC Oral Health       Date:  2022-05-22       Impact factor: 3.747

5.  Cyclic fatigue resistance of HyFlex EDM, Reciproc Blue, WaveOne Gold, and Twisted File Adaptive rotary files under different temperatures and ambient conditions.

Authors:  Mustafa Gündoğar; Taha Özyürek; Koray Yılmaz; Gülşah Uslu
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2019-10-07

6.  Evidence for reduced torsional resistance of rotary files under curved position.

Authors:  Ahmed Jamleh; Raghad Almedlej; Rasha Alomar; Nouf Almayouf; Abdulmohsen Alfadley; Khalid Alfouzan
Journal:  Saudi Dent J       Date:  2020-08-06

7.  Influence of NiTi Wire Diameter on Cyclic and Torsional Fatigue Resistance of Different Heat-Treated Endodontic Instruments.

Authors:  Eugenio Pedullà; Francesco Saverio Canova; Giusy Rita Maria La Rosa; Alfred Naaman; Franck Diemer; Luigi Generali; Walid Nehme
Journal:  Materials (Basel)       Date:  2022-09-22       Impact factor: 3.748

Review 8.  Present status and future directions: Canal shaping.

Authors:  Ana Arias; Ove A Peters
Journal:  Int Endod J       Date:  2022-02-16       Impact factor: 5.165

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.