Literature DB >> 17174681

Vickers hardness investigation of work-hardening in used NiTi rotary instruments.

Satish B Alapati1, William A Brantley, John M Nusstein, Glenn S Daehn, Timothy A Svec, John M Powers, William M Johnston, Wenhua Guo.   

Abstract

A contributing mechanism for clinical failure of NiTi rotary instruments might be excessive work hardening. Vickers hardness was measured with 300 gm load near the flutes in three regions (D2-D4, D6-D10, and D14 towards the shank) for nine representative clinically used ProFile GT instruments that had been axially sectioned (10 measurements in each region). Consistent values could not be obtained at D1. Minimum hardness occurred at D2 to D4, with means from 313 to 324 (SD from 7 to 16). Maximum hardness with means ranging widely from 330 to 481 was found beyond D14. Mean hardness at D2 to D4 was 320 for an as-received ProFile instrument. Because a mean Vickers hardness of 326 has been reported for a shape-memory NiTi orthodontic wire product, it can be concluded that the NiTi instruments did not experience substantial work hardening at D2 to D4 during clinical use, in agreement with previous differential scanning calorimetric analyses.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17174681     DOI: 10.1016/j.joen.2006.07.007

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  4 in total

1.  Performance of NiTi endodontic instrument under different temperatures.

Authors:  Ahmed Jamleh; Yoshio Yahata; Arata Ebihara; Amre R Atmeh; Turki Bakhsh; Hideaki Suda
Journal:  Odontology       Date:  2015-08-22       Impact factor: 2.634

2.  In vitro simulation of stent fracture mechanisms in ureteric nitinol wire stents.

Authors:  Tine Hajdinjak; Mayur Patel; Athanasios Papatsoris; Junaid Masood; Noor Buchholz; Malcolm Birch
Journal:  Urol Res       Date:  2008-08-28

3.  Comparison of Apical Transportation with the Use of Rotary System and Reciprocating Handpiece with Precurved Hand Files: An In Vitro Study.

Authors:  Fatemeh Mohammadian; Atefeh Sadeghi; Fatemeh Dibaji; Mona Sadegh; Zahra Ghoncheh; Mohammad Javad Kharrazifard
Journal:  Iran Endod J       Date:  2017

4.  Advancing Nitinol: From heat treatment to surface functionalization for nickel-titanium (NiTi) instruments in endodontics.

Authors:  Wai-Sze Chan; Karan Gulati; Ove A Peters
Journal:  Bioact Mater       Date:  2022-09-27
  4 in total