Phase transformation behavior and mechanical properties of thermomechanically treated K3XF nickel-titanium instruments.

Ya Shen1, Hui-Min Zhou, Zhejun Wang, Les Campbell, Yu-feng Zheng, Markus Haapasalo.   

Abstract

INTRODUCTION: The bending and torsional properties of thermomechanically treated K3XF (SybronEndo, Orange, CA) nickel-titanium instruments in relation to their phase transformation behavior were evaluated.
METHODS: NiTi instruments K3 (SybronEndo) and K3XF, both in sizes 25/.04 and 40/.04, were examined by differential scanning calorimetry and X-ray diffraction. The metal composition was determined by scanning electron microscopy with X-ray energy-dispersive spectrometric analyses. The bending property of K3 and K3XF instruments was measured in a cantilever-bending test with a maximum deflection of 4.00 mm. A torsional test of the instruments was evaluated according to the American National Standards Institute/American Dental Association Specification No. 28.
RESULTS: K3 and K3XF instruments had approximately the same chemical composition with a nickel content of 48-49 atomic %. The differential scanning calorimetry analyses showed that each segment of the K3XF instruments (24.89°C ± 1.98°C) had a higher austenite finish temperature than the K3 instruments (17.63°C ± 1.76°C) (P < .05). The bending load values were significantly lower for K3XF than for K3 in the superelastic ranges (P < .05). There was no statistically significant difference between K3 and K3XF in the maximum torque or maximum angular deflection before failure. The torque at fracture values of K3 and K3XF increased significantly with the diameter (P < .05).
CONCLUSIONS: K3XF exhibited different phase transformation behavior and flexibility when compared with K3, which may be attributed to the special heat treatment history of K3XF instruments.
Copyright © 2013 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23791264     DOI: 10.1016/j.joen.2013.04.004

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  14 in total

1.  Performance of NiTi endodontic instrument under different temperatures.

Authors:  Ahmed Jamleh; Yoshio Yahata; Arata Ebihara; Amre R Atmeh; Turki Bakhsh; Hideaki Suda
Journal:  Odontology       Date:  2015-08-22       Impact factor: 2.634

2.  Effect of autoclave sterilization on cyclic fatigue and torsional fracture resistance of NiTi rotary instruments.

Authors:  Wooyoung Kim; Soram Oh; Gil-Joo Ryu; Tae-Hwan Kim; Sung-Jae Kim; Dong-Hyung Kim; Bin-Na Lee; Kee-Yeon Kum; Seok Woo Chang; Ji-Hyun Jang
Journal:  Odontology       Date:  2019-09-11       Impact factor: 2.634

3.  Evaluation of apical root defects during canal instrumentation with two different nickel-titanium (NiTi) systems by optical coherence tomography (OCT) scan.

Authors:  Chen Chen; Wenxin Zhang; Yuhong Liang
Journal:  J Dent Sci       Date:  2021-10-23       Impact factor: 3.719

4.  To evaluate the influence of smear layer with different instruments and obturation methods on microleakage of root canal filled teeth: In vitro study.

Authors:  Manoj S Likhitkar; Shantaram V Kulkarni; Aravind Burande; Vishal Solanke; C Sushil Kumar; Babasaheb Kamble
Journal:  J Int Soc Prev Community Dent       Date:  2016-05-30

5.  Clinical and Radiographic Evaluation of Procedural Errors during Preparation of Curved Root Canals with Hand and Rotary Instruments: A Randomized Clinical Study.

Authors:  Rajesh Khanna; Aashish Handa; Rupam Kaur Virk; Deepika Ghai; Rajni Sharma Handa; Asim Goel
Journal:  Contemp Clin Dent       Date:  2017 Jul-Sep

Review 6.  Nickel-Titanium Rotary Instruments: Development of the Single-File Systems.

Authors:  Maryam Kuzekanani
Journal:  J Int Soc Prev Community Dent       Date:  2018-08-06

7.  Cyclic fatigue resistance of HyFlex EDM, Reciproc Blue, WaveOne Gold, and Twisted File Adaptive rotary files under different temperatures and ambient conditions.

Authors:  Mustafa Gündoğar; Taha Özyürek; Koray Yılmaz; Gülşah Uslu
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2019-10-07

8.  Assessment of various endodontic instrumentation systems on the amount of apically extruded bacteria - An in vitro study.

Authors:  Garima Sahni; Nisha Garg; Ajay Chhabra; Shabnam Negi; Sandeep Gupta; Vandana Chhabra
Journal:  J Conserv Dent       Date:  2020-12-04

9.  Cyclic Fatigue Resistance and Force Generated by OneShape Instruments during Curved Canal Preparation.

Authors:  Zhuyu Wang; Wen Zhang; Xiaolei Zhang
Journal:  PLoS One       Date:  2016-08-11       Impact factor: 3.240

10.  Influence of autoclave sterilization procedures on the cyclic fatigue resistance of heat-treated nickel-titanium instruments: a systematic review.

Authors:  Emmanuel João Nogueira Leal Silva; Mayara Zanon; Fernanda Hecksher; Felipe Gonçalves Belladonna; Rafaela Andrade de Vasconcelos; Tatiana Kelly da Silva Fidalgo
Journal:  Restor Dent Endod       Date:  2020-03-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.