Optimizing scintigraphic evaluation of split renal function in living kidney donors using the geometric mean method: a preliminary retrospective study.

Sarah Weinberger1, Michael Baeder2, Christian Scheurig-Muenkler3, Ingo Guenter Steffen2, Ahmed Magheli2, Kurt Miller4, Carsten Kempkensteffen4.   

Abstract

BACKGROUND: Accurate assessment of pre-transplant split renal function in candidates for living kidney donation is indispensable for side-selection and a sufficient long-term residual renal function.
OBJECTIVE: To analyse the need of depth correction in the assessment of split renal function in potential living kidney donors.
METHODS: In 13 consecutive patients screened for living kidney donation split renal function was measured with four different methods including conventional posterior MAG-3-scintigraphy, the geometric mean method in MAG-3-scintigraphy, MAG-3-scintigraphy with CT-based depth correction and CT-volumetry. Correlation and agreement of methods were analyzed using Spearman's rho correlation coefficient and the Bland-Altman method.
RESULTS: Despite good correlation and agreement between the different radioisotopic methods there were clinically relevant differences in split renal function in 2/13 patients (15 %) between conventional posterior MAG-3 scan and the geometric mean method. The best correlation was found between the two scintigraphic methods with depth correction. Comparing radioisotopic methods with CT-volumetry, significant differences were found in up to 6/13 patients (46 %).
CONCLUSIONS: Our results clearly indicate that in the case of living kidney donation further assessment concerning the accuracy and reliability of measuring split renal function is necessary. As there are no differences in duration of examination, costs and radiation exposure between techniques with and without depth correction, but clinically relevant differences in up to 46 % of patients, kidney depth should be incorporated in daily clinical practice of living kidney donor evaluation. The geometric mean method could significantly improve future patient assessment in cases of living kidney donation.

Entities:  

Keywords:  Geometric mean; Kidney function; Kidney transplantation; Living donor; Scintigraphy

Mesh:

Substances:

Year:  2015        PMID: 26286903     DOI: 10.1007/s40620-015-0223-z

Source DB:  PubMed          Journal:  J Nephrol        ISSN: 1121-8428            Impact factor:   3.902


  13 in total

Review 1.  Consensus report on quality control of quantitative measurements of renal function obtained from the renogram: International Consensus Committee from the Scientific Committee of Radionuclides in Nephrourology.

Authors:  A Prigent; P Cosgriff; G F Gates; G Granerus; E J Fine; K Itoh; M Peters; A Piepsz; M Rehling; M Rutland; A Taylor
Journal:  Semin Nucl Med       Date:  1999-04       Impact factor: 4.446

2.  The importance of using anterior and posterior views in the calculation of differential renal function using 99Tcm-DMSA.

Authors:  R Wujanto; R S Lawson; M C Prescott; H J Testa
Journal:  Br J Radiol       Date:  1987-09       Impact factor: 3.039

3.  Volume matters: CT-based renal cortex volume measurement in the evaluation of living kidney donors.

Authors:  Fabian Halleck; Gerd Diederichs; Torsten Koehlitz; Torsten Slowinski; Florian Engelken; Lutz Liefeldt; Frank Friedersdorff; T Florian Fuller; Ahmed Magheli; Hans-H Neumayer; Klemens Budde; Johannes Waiser
Journal:  Transpl Int       Date:  2013-10-19       Impact factor: 3.782

4.  Influence of kidney depth on the renographic estimation of relative renal function.

Authors:  P Cosgriff; H Brown
Journal:  J Nucl Med       Date:  1990-09       Impact factor: 10.057

Review 5.  Living kidney donation: outcomes, ethics, and uncertainty.

Authors:  Peter P Reese; Neil Boudville; Amit X Garg
Journal:  Lancet       Date:  2015-05-16       Impact factor: 79.321

6.  A technique for analysis of geometric mean renography.

Authors:  John S Fleming
Journal:  Nucl Med Commun       Date:  2006-09       Impact factor: 1.690

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  The conditions for which the geometric mean method revealed a more accurate calculation of relative renal function in 99mTc-DMSA scintigraphy.

Authors:  A Fuat Yapar; Mehmet Aydin; Mehmet Reyhan; Zeynep Yapar; Aysun Sukan
Journal:  Nucl Med Commun       Date:  2005-02       Impact factor: 1.690

9.  Kidney depth measurement and its influence on quantitation of function from gamma camera renography.

Authors:  S M Gruenewald; L T Collins; R M Fawdry
Journal:  Clin Nucl Med       Date:  1985-06       Impact factor: 7.794

10.  Optimizing evaluation of split renal function in a living kidney donor using scintigraphy and calculation of the geometric mean: a case report.

Authors:  S Weinberger; M Bäder; C Scheurig-Münkler; S Hinz; J Neymeyer; K Miller; C Kempkensteffen
Journal:  Case Rep Nephrol Urol       Date:  2014-01-10
View more
  4 in total

1.  Influence of CT-based depth correction of renal scintigraphy in evaluation of living kidney donors on side selection and postoperative renal function: is it necessary to know the relative renal function?

Authors:  Sarah Weinberger; Carola Klarholz-Pevere; Lutz Liefeldt; Michael Baeder; Nico Steckhan; Frank Friedersdorff
Journal:  World J Urol       Date:  2018-03-22       Impact factor: 4.226

2.  Can Split Renal Volume Assessment by Computed Tomography Replace Nuclear Split Renal Function in Living Kidney Donor Evaluations? A Systematic Review and Meta-Analysis.

Authors:  Steven Habbous; Carlos Garcia-Ochoa; Gary Brahm; Chris Nguan; Amit X Garg
Journal:  Can J Kidney Health Dis       Date:  2019-09-19

3.  The next era of renal radionuclide imaging: novel PET radiotracers.

Authors:  Rudolf A Werner; Xinyu Chen; Constantin Lapa; Kazuhiro Koshino; Steven P Rowe; Martin G Pomper; Mehrbod S Javadi; Takahiro Higuchi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2019-05-30       Impact factor: 9.236

4.  11C-PABA as a PET Radiotracer for Functional Renal Imaging: Preclinical and First-in-Human Study.

Authors:  Camilo A Ruiz-Bedoya; Alvaro A Ordonez; Rudolf A Werner; Donika Plyku; Mariah H Klunk; Jeff Leal; Wojciech G Lesniak; Daniel P Holt; Robert F Dannals; Takahiro Higuchi; Steven P Rowe; Sanjay K Jain
Journal:  J Nucl Med       Date:  2020-03-20       Impact factor: 11.082

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.